考虑体积效应的液闪测氚计数条件优化研究

上一篇

下一篇

陈志林, 孟丹, 李君利, 邱睿, 彭述明, PENG Shuming. 2014: 考虑体积效应的液闪测氚计数条件优化研究, 原子核物理评论, null(1): 58-62. doi: 10.11804/NuclPhysRev.31.01.058
引用本文: 陈志林, 孟丹, 李君利, 邱睿, 彭述明, PENG Shuming. 2014: 考虑体积效应的液闪测氚计数条件优化研究, 原子核物理评论, null(1): 58-62. doi: 10.11804/NuclPhysRev.31.01.058
CHEN Zhilin, MENG Dan, LI Junli, QIU Rui, WANG Heyi, PENG Shuming. 2014: Optimization of Counting Conditions on Tritium Measurement in LSC Considering Volume Effect, Nuclear Physics Review, null(1): 58-62. doi: 10.11804/NuclPhysRev.31.01.058
Citation: CHEN Zhilin, MENG Dan, LI Junli, QIU Rui, WANG Heyi, PENG Shuming. 2014: Optimization of Counting Conditions on Tritium Measurement in LSC Considering Volume Effect, Nuclear Physics Review, null(1): 58-62. doi: 10.11804/NuclPhysRev.31.01.058

考虑体积效应的液闪测氚计数条件优化研究

Optimization of Counting Conditions on Tritium Measurement in LSC Considering Volume Effect

  • 摘要: 采用外标转换谱指数(tSIE)作为淬灭校正因子,通过研究闪烁液及样品总体积、闪烁液用量的影响获得液闪测氚时的最佳计数条件。结果表明,对于22 mL标准玻璃计数瓶,最佳计数体积为7~17 mL,体积过少或过多可导致高到10%以上的测量误差,这主要是由液闪谱仪光电倍增管的面响应特性引起的。在获得的最佳计数体积基础上,对闪烁液的用量进行了研究。为了获得好的测量结果,闪烁液与样品的比例应高于2.4。另外,还通过实验研究了光照对测量结果的影响。在实验室条件下,实验室灯光及自然光所引起的本底辐射在样品静置2h后将不会对测量结果构成影响。
  • 加载中
  • 加载中
计量
  • 文章访问数:  313
  • HTML全文浏览数:  112
  • PDF下载数:  0
  • 施引文献:  0
出版历程
  • 刊出日期:  2014-03-20

考虑体积效应的液闪测氚计数条件优化研究

  • 中国工程物理研究院核物理与化学研究所,四川绵阳,621900
  • 清华大学工程物理系,北京,100084
  • Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, Sichuan, China

摘要: 采用外标转换谱指数(tSIE)作为淬灭校正因子,通过研究闪烁液及样品总体积、闪烁液用量的影响获得液闪测氚时的最佳计数条件。结果表明,对于22 mL标准玻璃计数瓶,最佳计数体积为7~17 mL,体积过少或过多可导致高到10%以上的测量误差,这主要是由液闪谱仪光电倍增管的面响应特性引起的。在获得的最佳计数体积基础上,对闪烁液的用量进行了研究。为了获得好的测量结果,闪烁液与样品的比例应高于2.4。另外,还通过实验研究了光照对测量结果的影响。在实验室条件下,实验室灯光及自然光所引起的本底辐射在样品静置2h后将不会对测量结果构成影响。