电热蒸发-电感耦合等离子体质谱法测量铀、钚同位素丰度
Determination of Isotope Ratio of 235U/238U and 239Pu/240Pu by Electrothermal Vaporization Inductively Coupled Plasma Mass Spectrometry
-
摘要: A method on determination of isotope ratio of 235U/238U and 239Pu/240Pu by electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICPMS) is developed. The operation parameters of ETV and ICPMS are optimized respectively. The absolute detected limit is 0.3 pg (3k), and the RSDs of 235U/238U and 239Pu/240Pu are better than 5%.
-
关键词:
- ETV /
- ICPMS /
- chemical modifier /
- U /
- Pu
Abstract: A method on determination of isotope ratio of 235U/238U and 239Pu/240Pu by electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICPMS) is developed. The operation parameters of ETV and ICPMS are optimized respectively. The absolute detected limit is 0.3 pg (3k), and the RSDs of 235U/238U and 239Pu/240Pu are better than 5%. -
-