皮克量级钚同位素丰度比的TIMS分析
Analysis on the Isotopic Ratio of Picogram Pu by TIMS
-
摘要: The isotopic abundance ratios for Pu at 6.7 pg level were measured by thermal surface ionization mass spectrometry (TIMS). During the preparation of Pu samples carbon power was used as an emitting and stabilizing reagent, which increase the collection efficiency, the ion current intense and the ion emission stability for Pu. The measurment results indicate that the relative standard deviation of 2.7% for the isotopic abundance ratios of 240Pu to 239Pu is achieved when the 240Pu ion current is 8-20 cps.
-
关键词:
- picogram Pu /
- TIMS /
- isotopic abundance ratio
Abstract: The isotopic abundance ratios for Pu at 6.7 pg level were measured by thermal surface ionization mass spectrometry (TIMS). During the preparation of Pu samples carbon power was used as an emitting and stabilizing reagent, which increase the collection efficiency, the ion current intense and the ion emission stability for Pu. The measurment results indicate that the relative standard deviation of 2.7% for the isotopic abundance ratios of 240Pu to 239Pu is achieved when the 240Pu ion current is 8-20 cps. -
-