利用MICROMEGAS探测器测量第一汤逊系数
Measurement of First Townsend Coefficient Using MICROMEGAS Detector
-
摘要: 利用新型的MICROMEGAS探测器测量了6种不同比例的Ar+CO2混合气体中第一汤逊系数α的两个实验参数A和B。实验结果分析表明,电子在栅极的透过率在75%时,α的测量误差小于13%,而探测器增益的涨落造成的α的测量误差小于8%。
-
关键词:
- 第一汤逊系数 /
- MICROMEGAS探测器 /
- 气体探测器
Abstract: In this paper,the MICROMEGAS detector was employed to measure the experimental parameters A and B of the first Townsend coefficient α in 6 different ratio of Ar+CO2 gas mixtures.The results indicate that the error of α is less than 13% in the condition of electron transmission rate of the mesh up to 75% and is less than 8% due to gain fluctuation of the detector. -
-
计量
- 文章访问数: 361
- HTML全文浏览数: 56
- PDF下载数: 0
- 施引文献: 0