液相色谱-串联质谱法测定尿液中的可替宁

上一篇

下一篇

刘彤, 侯宏卫, 张小涛, 陈欢, 唐纲岭, 胡清源. 2012: 液相色谱-串联质谱法测定尿液中的可替宁, 质谱学报, 33(4): 252-256.
引用本文: 刘彤, 侯宏卫, 张小涛, 陈欢, 唐纲岭, 胡清源. 2012: 液相色谱-串联质谱法测定尿液中的可替宁, 质谱学报, 33(4): 252-256.
LIU Tong, HOU Hong-wei, ZHANG Xiao-tao, CHEN Huan, TANG Gang-ling, HU Qing-yuan. 2012: Determination of Cotinine in Urine Using High Performance Liquid Chromatography-Electrospray Ionization Tandem Mass Spectrometry, Journal of Chinese Mass Spectrometry Society, 33(4): 252-256.
Citation: LIU Tong, HOU Hong-wei, ZHANG Xiao-tao, CHEN Huan, TANG Gang-ling, HU Qing-yuan. 2012: Determination of Cotinine in Urine Using High Performance Liquid Chromatography-Electrospray Ionization Tandem Mass Spectrometry, Journal of Chinese Mass Spectrometry Society, 33(4): 252-256.

液相色谱-串联质谱法测定尿液中的可替宁

Determination of Cotinine in Urine Using High Performance Liquid Chromatography-Electrospray Ionization Tandem Mass Spectrometry

  • 摘要: 为了准确测定吸烟者与非吸烟者尿液中可替宁含量水平,建立了液相色谱-串联质谱(LC-MS/MS)测定方法.尿液经稀释、离心和过滤,引入LC-MS/MS,经Agilent Zorbax Eclipse XDB-C18色谱柱(150 mm×2.1 mm,3.5 μm)分离,以0.1%甲酸水溶液-0.1%乙酸甲醇溶液为流动相,采用梯度洗脱方式,以电喷雾离子源正离子多反应监测模式进行分析.方法检出限为0.08 μg/L,线性范围为1~500 μg/L,空白样品中加标回收率范围在94.3%~99.9%,精密度(以相对标准偏差计)为0%~3.2%.利用此方法检测了58个非吸烟者和246个吸烟者尿液中可替宁的含量水平,结果表明:吸烟者24 h尿液中的可替宁含量是非吸烟者的180多倍,但是由于个体之间以及吸烟行为之间的差异,尿液中的可替宁含量并不随卷烟主流烟气中烟碱含量的升高而升高.
  • 加载中
  • 加载中
计量
  • 文章访问数:  300
  • HTML全文浏览数:  34
  • PDF下载数:  56
  • 施引文献:  0
出版历程
  • 刊出日期:  2012-08-28

液相色谱-串联质谱法测定尿液中的可替宁

  • 国家烟草质量监督检验中心,河南郑州,450001

摘要: 为了准确测定吸烟者与非吸烟者尿液中可替宁含量水平,建立了液相色谱-串联质谱(LC-MS/MS)测定方法.尿液经稀释、离心和过滤,引入LC-MS/MS,经Agilent Zorbax Eclipse XDB-C18色谱柱(150 mm×2.1 mm,3.5 μm)分离,以0.1%甲酸水溶液-0.1%乙酸甲醇溶液为流动相,采用梯度洗脱方式,以电喷雾离子源正离子多反应监测模式进行分析.方法检出限为0.08 μg/L,线性范围为1~500 μg/L,空白样品中加标回收率范围在94.3%~99.9%,精密度(以相对标准偏差计)为0%~3.2%.利用此方法检测了58个非吸烟者和246个吸烟者尿液中可替宁的含量水平,结果表明:吸烟者24 h尿液中的可替宁含量是非吸烟者的180多倍,但是由于个体之间以及吸烟行为之间的差异,尿液中的可替宁含量并不随卷烟主流烟气中烟碱含量的升高而升高.

English Abstract

参考文献 (0)

目录

/

返回文章
返回