大气气溶胶中超痕量钚的电感耦合等离子体质谱(ICP-MS)分析

上一篇

下一篇

周国庆, 朱凤蓉, 张子斌, 白峻峰, 金玉仁, 张佳媚, 董宏波, 李梅, 翟利华. 2002: 大气气溶胶中超痕量钚的电感耦合等离子体质谱(ICP-MS)分析, 质谱学报, 23(3): 151-155. doi: 10.3969/j.issn.1004-2997.2002.03.005
引用本文: 周国庆, 朱凤蓉, 张子斌, 白峻峰, 金玉仁, 张佳媚, 董宏波, 李梅, 翟利华. 2002: 大气气溶胶中超痕量钚的电感耦合等离子体质谱(ICP-MS)分析, 质谱学报, 23(3): 151-155. doi: 10.3969/j.issn.1004-2997.2002.03.005
2002: Determination of Ultra-trace Plutonium in Atmospheric Aerosols by Inductively Coupled Plasma Mass Spectrometry (ICP-MS), Journal of Chinese Mass Spectrometry Society, 23(3): 151-155. doi: 10.3969/j.issn.1004-2997.2002.03.005
Citation: 2002: Determination of Ultra-trace Plutonium in Atmospheric Aerosols by Inductively Coupled Plasma Mass Spectrometry (ICP-MS), Journal of Chinese Mass Spectrometry Society, 23(3): 151-155. doi: 10.3969/j.issn.1004-2997.2002.03.005

大气气溶胶中超痕量钚的电感耦合等离子体质谱(ICP-MS)分析

Determination of Ultra-trace Plutonium in Atmospheric Aerosols by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

  • 摘要: 本文通过采用膜去溶进样系统与ICP-MS联用技术和同位素稀释质谱法定量分析了大气气溶胶样品中fg量级的钚.方法对大气中239Pu的定量限达到了5.7×10-8 Bq/m3.膜去溶系统的使用提高了样品的利用效率,并有效降低了复合离子238U1H+对239Pu测定的影响,使仪器的分析灵敏度(相对)提高了近一个数量级,对239Pu的检出限达到了0.4 fg/mL.
  • 加载中
  • 加载中
计量
  • 文章访问数:  354
  • HTML全文浏览数:  51
  • PDF下载数:  12
  • 施引文献:  0
出版历程
  • 刊出日期:  2002-08-28

大气气溶胶中超痕量钚的电感耦合等离子体质谱(ICP-MS)分析

  • 西北核技术研究所,陕西,西安,710024

摘要: 本文通过采用膜去溶进样系统与ICP-MS联用技术和同位素稀释质谱法定量分析了大气气溶胶样品中fg量级的钚.方法对大气中239Pu的定量限达到了5.7×10-8 Bq/m3.膜去溶系统的使用提高了样品的利用效率,并有效降低了复合离子238U1H+对239Pu测定的影响,使仪器的分析灵敏度(相对)提高了近一个数量级,对239Pu的检出限达到了0.4 fg/mL.

English Abstract

参考文献 (0)