王晟, 刘晶儒, 胡志云, 张振荣, 张立荣, 叶景峰, 赵新艳, 黄梅生. 2008: 分子过滤瑞利散射技术测量火焰温度和密度, 强激光与粒子束, 20(12): 2001-2005.
引用本文: 王晟, 刘晶儒, 胡志云, 张振荣, 张立荣, 叶景峰, 赵新艳, 黄梅生. 2008: 分子过滤瑞利散射技术测量火焰温度和密度, 强激光与粒子束, 20(12): 2001-2005.
WANG Sheng, LIU Jing-ru, HU Zhi-yun, ZHANG Zhen-rong, ZHANG Li-rong, YE Jing-feng, ZHAO Xin-yan, HUANG Mei-sheng. 2008: Molecular filtered Rayleigh scattering diagnostic for measurement of temperature and density in flame, High Power Lase and Particle Beams, 20(12): 2001-2005.
Citation: WANG Sheng, LIU Jing-ru, HU Zhi-yun, ZHANG Zhen-rong, ZHANG Li-rong, YE Jing-feng, ZHAO Xin-yan, HUANG Mei-sheng. 2008: Molecular filtered Rayleigh scattering diagnostic for measurement of temperature and density in flame, High Power Lase and Particle Beams, 20(12): 2001-2005.

分子过滤瑞利散射技术测量火焰温度和密度

Molecular filtered Rayleigh scattering diagnostic for measurement of temperature and density in flame

  • 摘要: 为了解决瑞利散射光易受米散射和背景杂散光干扰的问题,发展了结合窄线宽激光器、分子过滤器以及像增强器等技术的分子过滤瑞利散射技术.在图像诊断的基础上,依据测量的碘蒸气吸收光谱曲线,对CH4/air预混火焰进行了诊断,获得了密度场和温度场分布.距炉面15 mm火焰中心区域处,分子过滤瑞利散射(FRS)技术测量的温度为1827 K±84 K,密度为0.19 kg/m3,其测温结果与CARS法的测温结果基本吻合.最后分析了FRS技术测温不确定度.实验表明FRS技术具有较高的信噪比,可以定量测量温度和密度信息,有望应用于超音速燃烧流场、紊流场等复杂流场的诊断.
  • 加载中
  • 加载中
计量
  • 文章访问数:  491
  • HTML全文浏览数:  43
  • PDF下载数:  92
  • 施引文献:  0
出版历程
  • 刊出日期:  2008-12-30

分子过滤瑞利散射技术测量火焰温度和密度

  • 西北核技术研究所,西安,710024

摘要: 为了解决瑞利散射光易受米散射和背景杂散光干扰的问题,发展了结合窄线宽激光器、分子过滤器以及像增强器等技术的分子过滤瑞利散射技术.在图像诊断的基础上,依据测量的碘蒸气吸收光谱曲线,对CH4/air预混火焰进行了诊断,获得了密度场和温度场分布.距炉面15 mm火焰中心区域处,分子过滤瑞利散射(FRS)技术测量的温度为1827 K±84 K,密度为0.19 kg/m3,其测温结果与CARS法的测温结果基本吻合.最后分析了FRS技术测温不确定度.实验表明FRS技术具有较高的信噪比,可以定量测量温度和密度信息,有望应用于超音速燃烧流场、紊流场等复杂流场的诊断.