2025 Volume 38 Issue 1
Article Contents

Shaowen Feng, Zhen Li, Wenli Liu, Qiang Zhang, Yang Chen, Dongfeng Zhao. A Plasma Injection Ion Source for Photoelectron Imaging of Metal-Containing Anions[J]. Chinese Journal of Chemical Physics, 2025, 38(1): 1-7. doi: 10.1063/1674-0068/cjcp2310106
Citation: Shaowen Feng, Zhen Li, Wenli Liu, Qiang Zhang, Yang Chen, Dongfeng Zhao. A Plasma Injection Ion Source for Photoelectron Imaging of Metal-Containing Anions[J]. Chinese Journal of Chemical Physics, 2025, 38(1): 1-7. doi: 10.1063/1674-0068/cjcp2310106

A Plasma Injection Ion Source for Photoelectron Imaging of Metal-Containing Anions

通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(6)

Article Metrics

Article views(158) PDF downloads(0) Cited by(0)

Access History

A Plasma Injection Ion Source for Photoelectron Imaging of Metal-Containing Anions

Abstract: A plasma injection ion source has been developed for the photoelectron velocity imaging studies of metal-containing anions. The source employs a pulse discharge nozzle for generating a plasma beam that perpendicularly crosses the master supersonic jet beam from a home-made pulsed piezo valve. The discharge nozzle is designed for high voltage gas discharge with efficient metal sputtering of the cathode, and thus plays a role in metal atom and ion source. Supersonically jet-cooled anions can be produced in the master gas jet via reactions of the plasma products. The source is integrated into a photoelectron velocity imaging spectrometer. Test mass spectrometry experiments show that the ion source can efficiently produce transition metal containing anions, such as FeOm–, CuOm–, CuCn–, CuCnOm–. The photoelectron imaging results by photodetachment of O– show that the photoelectron energy resolution of the whole instrument is ΔE/E ≈ 2.3%, and the results of FeO− indicate that internal temperatures of anions from the source could be efficiently cooled down.

    • Transition metal-containing anions and their corresponding neutral compounds have been identified as significant contributors in numerous research fields [1–4]. Their characteristic physical, chemical and optical properties have drawn a lot of attention in fundamental studies by techniques such as collision-induced dissociation [5], Raman [6] and infrared spectroscopy [7], and photoelectron spectroscopy [8, 9]. However, in negative ion photoelectron spectroscopic studies, investigation of metal-containing anions significantly depends on the establishment of a source that can efficiently generate these reactive particles. The primary experimental challenges include the capability to generate a high-density ion sample, the development of an effective method to produce a mass-selected beam of anionic particles, and the availability of an inexpensive and easy way for the cooling of anions.

      In recent years, the combination of supersonic molecular beams and mass spectrometry has been widely used for the study of positive and negative ions. For the generation of metal-containing anions, a variety of plasma-based sources have been developed, e.g., pulsed-arc discharge [10, 11], gas aggregation [12], laser ablation [13, 14], sputtering [15], electron impact ion source [16], and electrospray ionization [17, 18]. Furthermore, the ion trap technique [18, 19], despite its considerable cost, is commonly employed to reduce the internal temperature of anions, thereby enhancing the resolution and reliability of experimental results.

      In this work, we demonstrate a plasma injection ion source that is coupled to a newly constructed photoelectron imaging spectrometer, aiming for spectroscopic characterization of metal-containing anions and other gas-phase anionic clusters. The general design of the ion source is based on the idea of a combination of two supersonic molecular beams, where a supersonic plasma beam is injected into the collision zone of a master supersonic jet expansion. The plasma assembly is designed based on high power gas discharge, with efficient metal sputtering of its cathode, and thus produce transition metal-containing species for subsequent photoelectron imaging studies.

    II.   EXPERIMENTAL DETAILS
    • A schematic view of the newly constructed photoelectron velocity imaging spectrometer, which has recently applied to the experimental study on the photoelectron spectroscopy and gas-phase stability of the C3O2– anion [20], is shown in FIG. 1. The whole instrument consists of four chambers for ion generation, ion deflection and acceleration, ion collimation, and photoelectron imaging, respectively. Interconnections between the ion source and ion deflection chambers are facilitated by a gate valve, while others by small apertures for differential pumping. By using a set of turbo molecular pumps, the vacuum pressures of the whole system can be maintained up to 10−9 mbar when the ion source is not running. During the experiment running, the four chambers can be maintained with vacuum pressures of ~1×10−4 mbar, 5×10−6 mbar, 5×10−8 mbar, and 1×10−8 mbar, respectively.

    • A major development in the constructed photoelectron imaging spectrometer is the plasma injection ion source (FIG. 2). The source consists of three parts: (i) a homemade piezo valve that generates a free-running supersonic jet expansion from a high pressure (>15 bar) gas mixture; (ii) a pulsed discharge nozzle assembly that generates a pulsed plasma beam perpendicularly crossing the master jet near the piezo valve exit; and (iii) a double skimmer system for sequential collimation of the master supersonic beam. To improve the anion production and jet-cooling efficiency, we have also developed a home-made pulsed piezo valve for the master jet beam in the ion source.

      The design of the pulsed piezo valve, similar to that in Ref.[21], is illustrated in FIG. 3. It is constructed based on a high-force piezoelectric stack actuator (PSA, PST1000, CoreMorrow) that can rapidly switch the nozzle exit throat open and closed. The custom-ordered PSA can achieve an 80 μm linear motion with pushing and pulling forces up to 4000 and 700 N, respectively. The PSA is assembled on the valve body via a cylindrical hollow bracket. The valve orifice is machined with a throat diameter of 0.5 mm and an exit cone angle of 40°. To prevent any torsion that could damage the actuator, a bolt-on geometry of the valve body is chosen, allowing for a fine adjustment of the distance between the plunger and the nozzle orifice via the internal threads. The stagnation gas chamber in the valve body, with a volume of 5 cm3, is sealed via a set of thin O-rings. To drive the valve, we have also designed a high voltage (HV) pulse generator based on a fast HV push-pull switch (Behlke, HTS 31-03-HB-C), which offers HV pulses with tunable pulse duration (20–100 μs) and amplitude (0–500 V). With these configurations, the valve can be operated with a stagnation gas pressure higher than 15 bar and a maximum repetition rate up to 20 kHz. Additional test shows that the valve can be operated with a pressure up to 20 bar.

      The pulsed discharge nozzle (PDN) [22], assembled on a pulsed general valve (Parker-Hannifin, Series 9), is comprised of three electrodes. The inner electrode is fixed in the center of the stainless connector, and the middle electrode are connected to the ground potential, while the outer one for discharge is pulsed to a high negative voltage. Electrical isolation in the whole discharge nozzle is achieved by machined ceramic plates. The polarity of the outer and middle discharge electrodes can be switched, which has been shown to be an effective way to further enhance the intensity of gas-phase radicals [23]. The negative high voltage pulses that discharge the supersonic gas from the valve are supplied by a combination of a high voltage supply power (DongWen, DM-N302-200ACB1) and another fast HV push-pull switch (Behlke, HTS 31-03-HB-C). The outer electrode is designed as a 2 mm thick round plate with a 3 mm apperture, so that it can work as a hollow cathode during the gas discharge. We found that, the larger aperture of the outer electrodes will require a higher discharge voltage for a stable plasma, which could be due to lower gas concentration in the discharge zone. On the other side, higher discharge voltage can yield high efficiency of cathode sputtering, acting as a source of metal atoms and ions. It has been known that hollow cathode discharge is an efficient way to evaporate almost all metals, and such hollow cathode lamps have been commercially available. In the present design, for different metal materials, many experimental parameters, such as the gas density, the discharge voltage and the central hole diameter of the electrodes, will need to be carefully optimized to obtain a reasonable metal evaporation efficiency. Here, we have used an aperture of 3 mm in the outer electrode. Typical discharge conditions are obtained by voltages between −1000 V and −2000 V, with a discharge current of ~10 mA, and a pulse duration of 50 μs.

      The plasma expansion that contains metal-containing ions, low energy electrons, and other reactive species, is arranged to cross the master jet expansion a few mm downstream from the exit of the piezo valve. The cross position is still within the collision zone of the master jet expansion, where complicated chemical processes, including multibody collisions, electron attachment, and ion-molecule reactions, are expected, yielding the production of a variety of anions of interest. In addition, the master jet expansion also plays a role of supersonic jet cooling the internal temperatures of produced anions [14, 24].

    • The master jet beam containing the produced anions is skimmed by a double skimmer system, and then freely flies into a three-plate-electrode assembly for ion deflection and acceleration. Typical repeller and extractor voltages on the ion extraction electrodes are pulsed to coincide with the produced anions in the master jet, and are set to −970 V and −848 V, respectively. After anion extraction, the ion beam is further collimated by two sets of collimating electrodes, each of which consists of two perpendicularly arranged deflecting plate electrodes and one Einzel lens system. Before entering the photoelectron imaging assembly, a mass gate, driven by a short high voltage pulse (−800 V, 50 μs), is employed for mass gating and selection of the target anions.

      Prior to the photoelectron imaging (PI) system, a series of electrodes featuring central apertures of decreasing size are designed to collimate anions at the entrance of the PI lens system. The PI lens assembly, shielded from the earth’s magnetic field and other trace magnetic fields in the lab by a two-layer μ-metal shielding, is similar to the design in Refs.[25–27]. The PI lens includes four crucial electrodes (repeller, extractor, lens, and ground) and three slave electrodes. The selected anions interact with a nanosecond pulsed laser beam in the photodetachment zone between the repeller and extractor electrodes. Through the PI lens assembly, photoelectrons from photodetachment of anions are focused and projected onto a position-sensitive imaging detector consisting of two 75 mm diameter microchannel plates (MCPs) coupled to a P43 phosphor screen (BIS, BOS-75). The detector is mounted 0.35 m away from the repeller electrodes of the PI lens assembly. A complementary metal oxide semiconductor (CMOS) camera (Tucsen, Dhyana 400BSI V2, 2048×2048 pixels, 2×2 binding) captures the 2D photoelectron images on the phosphor screen and transmits the data to a computer for further analysis. The raw photoelectron imaging data are reconstructed through inverse Abel transformation to get an electron kinetic energy (eKE) spectrum [28]. The photoelectron spectra are presented in terms of the electron-binding energy (eBE = $h\nu$–eKE), which is independent of the laser photon energy $h\nu$. In addition, the detector can also be operated as a single-point ion detector for TOF mass spectrometry, which is enabled from the deflection and acceleration electrodes of the instruments. This will allow us to characterize the types of anion species produced in our experiment.

      Voltages applied to the PI lens assembly are set based on ion trajectory simulations using the SIMION software [29], and further optimized based on the test experiments on photodetachment of selected O− and FeO−. Typical settings we have used are −300 V, −252 V, and −184 V, for the repeller, extractor, and lens electrodes, respectively. In addition, for different ranges of electron kinetic energy, these voltages can be proportionally decreased to get a better-resolved photoelectron imaging.

    III.   EXPERIMENTAL RESULTS AND DISCUSSION
    • FIG. 4 shows a set of experimentally recorded TOF mass spectra under different experimental conditions to demonstrate the production capability of metal-containing anions by the plasma injection anion source. In the experiment of iron compounds, we have used discharge electrodes made by iron, a gas mixture of 4%O2 in He at a backing pressure of 5 bar in the plasma source, and a 10%O2 and Ar gas mixture with a backing pressure of 15 bar in the master jet expansion. We have chosen heavier Ar carrier gas for the master jet, and light He carrier gas for the plasma jet to ensure that the side plasma jet would not significantly affect the properties of the master jet expansion, such as the expansion profile and cooling efficiency. As depicted in FIG. 4(a), the mass spectrum of one-iron oxide anions (FeOm–) spans from m=1 to 18 oxygen atoms. Contrary to the observed pattern for cationic complexes in [30], for $ m $>2, FeOm– species with an odd $ m $ exhibit higher abundance than those with the nearby even $ m $. Interestingly, FeO9– anion appears to be the inflection point of anionic intensity, while FeO10+ cation consistently emerges as the most intense peak among the cationic complexes [30–32].

      We have also tested the production of copper containing anions by using high-purity copper electrodes in the plasma assembly. Copper-containing compounds are straightforwardly identified in the TOF mass spectra in FIG. 4(b), due to the distinct natural abundance ratio of 63Cu and 65Cu isotopes, which is approximately 0.69:0.31. We found that, the copper compound can be efficiently produced with a discharge voltage as low as −800 V. This is not surprising, as copper has a higher cathode sputtering efficiency than iron due to its lower melting point. On the other hand, the lower discharge voltage limits the production yields of larger anionic species. In addition, the results in FIG. 4(b) also show that, neutral and/or anionic copper atoms seems to be very reactive, yielding a variety of anionic complexes, such as CuOm–, CuCn–, CuCnOm–.

    • By using the plasma injection anion source, we have recorded photoelectron images by photodetachment of O− and FeO−. FIG. 5 shows the results of the O− with a detachment laser at 620.5 nm. Since the photoelectron spectroscopy of O− has been well studied in Refs.[33–36], it is used to calibrate the pixel velocity that is required to convert the photoelectron image to electron kinetic energy. The final calibrated spectrum is also shown in FIG. 5, where intense and partially resolved features can be assigned as the spin-orbit resolved transitions of O(3P)$ \leftarrow $O− (2P$_{3/2, 1/2}$), with spin-orbit excited O−(2P$ _{1/2} $) features much weaker. From this result, we estimate an phtoelectron kinetic energy resolution of $ \Delta $E/E=2.3% (at E=0.54 eV) in our experiment.

      FIG. 6 shows photoelectron images of FeO− recorded with the detachment laser at 620.5 nm and 532.1 nm, respectively. The experiment is performed by selecting anions with a mass-to-charge ratio of 72, which signal disappears when using copper electrodes. The eBE spectra are very similar to that in Ref.[8]. Assignments of observed features to the spin-orbit splitting components and vibrational numbers of neutral product are made following Ref.[8]. It should be noted that, previous experiments have shown that, two low-lying states of FeO−, X$ ^4\Delta $ and a$ ^6\Sigma^+ $, can be produced in laboratory experiments [37–39]. Even in Ref.[8], where a 10 K ion trap is used to efficiently cool down the anions, photoelectrons from FeO− (a$ ^6\Sigma^+ $) are still significantly observed. Based on the assignment of our eBE spectra, we found that all features are due to photodetachment from the X$ ^4\Delta_{7/2} $ (v=0) and a$ ^6\Sigma^+ $ (v=0) states. In other words, the spin-orbit and vibrationally excited states in both X$ ^4\Delta $ and a$ ^6\Sigma^+ $ have been efficiently cooled down in our anion source. Moreover, by comparing the most intense eBE peak signal (X$^5\Delta_4 \leftarrow$X$ ^4\Delta_{7/2} $) and the weaker eBE peak signal (a$ ^7\Sigma^+ \leftarrow $a$ ^6\Sigma^+ $), with the experimental spectra presented in Ref.[8], we roughly estimate that the rotational temperature of the FeO− anion in our experiment could be very close to that in a 10 K ion trap, with an upper limit of 30 K. These results indicate that, internal temperatures of the anions produced in the plasma injection source could be efficiently cooled down for subsequent photodetachment studies.

    IV.   CONCLUSION
    • In summary, the plasma injection ion source demonstrated here has proven its capability in the production of metal-containing anions, where the metal source is due to cathode sputtering during high voltage gas discharge. By applying the ion source to a newly constructed photoelectron imaging spectrometer, we have obtained an energy resolution of 2.3$\%$ in the obtained photoelectron kinetic energy spectrum. Test experiments on the photoelectron spectra of FeO– further show that internal temperatures of anions from the source can be efficiently cooled down and could be close to that in a 10 K ion trap. In commonly used single beam sources, such as pulsed discharge, electron impact ionization, laser ablation, ion production processes required additional energy to be injected into the supersonic jet expansion, which decreases the internal cooling of the target species. In the present design, most energetic processes take place in the side plasma jet, while the mater jet serves as a nearly free-running environment for additional ion/electron-molecule reaction and efficient supersonic cooling of produced anions. Although we have only tested the production of copper and iron containing anions here, other metals could also be applicable, which will extend the application of this anion source useful in future studies on photoelectron spectroscopy and photochemistry of such reactive ions.

    Figure (6)  Reference (39)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return