Anion photoelectron spectroscopy (PES) is a very powerful technique for the determination of electron affinities and the study of exotic species such as metal and semiconductor clusters [1−4]. Negative ions can be mass-selected prior to spectroscopic investigation and selection rules for photodetachment versus excitation of a neutral molecule are drastically relaxed. With the development of ion imaging [5] and velocity-map imaging (VMI) [6] techniques, the photoelectron (PE) imaging method was considered as an attractive PES alternative that combines the high collection efficiency with the ability to detect low kinetic electrons and measure angular distributions [7−9]. Some other laboratories have performed the slow electron velocity map imaging (SEVI) to obtain the high-resolution PES [10−12]. Several years ago, a PES apparatus with the laser-vaporization supersonic cluster source based on the VMI technique, was built in our laboratory [13]. It has been used to study various negative species produced in the supersonic beam, providing electron affinities and information about electronic and vibrational states of the neutral clusters [14−16]. However, there are some limitations of that apparatus.
On one hand, since the ion source is in the laser-vaporization mode, the anionic species can be studied only from solid samples, which means the target sample should be macroscopically prepared and the high power of ablative laser will easily break the basic structure of molecules. Thus, such a configuration of the ion source is not suitable to study the biomolecule and lab-synthesized nano-cluster, which are dissolved in solution phase easily and more suitable for studies with the electrospray ionization (ESI) source. ESI is a technique coupled with the mass spectrometer to produce ions using an electrospray in which a high voltage is applied to a liquid to create an aerosol [17]. The birth of this technique made it possible to study the solution phase species in the gas phase. All biologically related processes associated with the chemical reactions take place in an aqueous solution, and it is a considerable challenge for obtaining a molecular-level understanding of these phenomena in the solution phase. The combined ESI and PES is a powerful experimental technique to investigate solution phase species and their chemical process in the gas phase under molecular-level. Up to now, this combination technique was used to carry out researches on various negative ion species, such as the electronic structure and chemical bonding characteristics studies on multiply charged anions [18−21], dipole-bound state (DBS) of the anions [22−24] and biorelated systems [25, 26]. The time-resolved dynamics process of various size anions was also investigated by ESI-PES apparatus combined with the ultrafast laser [27−29].
A Wiley-McLaren type time of flight mass spectrometer (TOFMS) is used in our previous PES experiment to select the photodetachment species [12−14]. The mass resolution of this type TOFMS is usually in the region of several hundreds. For larger clusters which contain heavy atoms, such resolution makes it difficult to distinguish the target cluster ions accurately. And this will make contamination in the PE spectrum. For example, the PE imaging results of AgH− and Ag2H− also contained the signal from Ag− and Ag2−, respectively [30]. In this work, a dual-reflection mass analyzer is used, which can increase the flight time of ions in a limited length flight tube and make energy compensation, thus effectively improving the resolution of mass spectrometry.
Here, we report the design and configurations of a novel PES instrument, which consists of an ESI source, a cryogenically cooled ion trap, a dual-reflection TOFMS, and a high-resolution VMI spectrometer. The detailed description on the features of this new apparatus will be demonstrated in the following sections.
-
The structural schematic diagram of the whole experimental apparatus is shown in FIG. 1. According to the overall composition and function of the instrument, the apparatus can be divided into three major sections: (I) ESI ion source (including a set of radio-frequency (RF)-only ion guide and an ion trap); (II) a dual-reflection TOF mass analyzer; (III) a VMI photoelectron spectrometer. Charged liquid droplets from a syringe needle (1) enter the desolvation capillary (2). Molecular ions exiting the desolvation capillary are guided into the 3D Paul trap (9) through an array of RF-only ion guide system which consists of two quadrupole (4, 6) and one octopole rods (8). Ions pulsed out of the ion trap are guided and focused by a plate lens into the extraction zone (10) of the TOFMS, where ions are extracted perpendicularly. The extracted ions in the TOFMS flight tube are focused twice by dual-reflection focus lens (13, 14), and mass selected (17), before entering the detachment region of VMI photoelectron spectrometer. A tunable wavelength laser beam perpendicular to orientation of ions flight is crossed with the ion beam in the detachment point. The emitted photoelectrons are all collected by an imaging detector (22, 23) through the VMI lens system (20). Details are described in the following.
-
As shown in FIG. 2, electrospray ion source is a soft ionization technique used to generate positive and negative ions. The mechanisms of the ion formation have been discussed previously [16]. The sample solution is injected into the ambient atmosphere through a fused quartz needle tube with a diameter of micrometers. The fused quartz needle is connected by a stainless steel through-joint, which can be applied with a DC high voltage to form negatively charged droplets, and the liquid sample stream is pushed by the syringe pump (LEGATO 100). Under the pressure difference and the electric field, the negatively charged droplets are introduced into the heated desolvation capillary assembly, which consists of a copper block with a capillary tube and two heaters, and the temperature of the copper block can be raised to 100 °C.
After being desolvated by the heating module, the negative ions should be guided into the ion trap, which is set in the high vacuum chamber. To complete the transition from the atmospheric pressure to the high vacuum, four vacuum differential chambers are isolated by three skimmers with the diameter of 1.5 mm. These four chambers are pumped by the mechanical pump, the roots pump, the turbo pump, and the vacuum are maintained at about 500 Pa, 20 Pa, 2×10−3 Pa, and 1×10−4 Pa, respectively. For the short molecular mean free path of ions in low vacuum, the ions will lose a lot by colliding with the background gas. In this case, the electrostatic lens group is not suitable for focusing and transporting ions. Quadrupole can pull radial ions back to their paraxial position by loading a RF electric field, and the focusing effect of the electric field at the quadrupoles is enhanced for the ion energy decreasing due to collisions between ions and gases, resulting in a more compact ion beam. This focusing effect can make the transmission efficiency of the quadrupole reach up to a high level. The quadrupole with a RF-only voltage is choosed as a powerful technique to focus and transmit the ions in the lower vacuum regime.
These two quadrupoles with the same radium of 6.35 mm are used to guide the negative ions in the second and third vacuum chamber. The lengths of the first and second stainless-steel quadrupole rods are 194.5 and 275.0 mm, respectively. The field radius of the quadrupole is designed to be 5.53 mm. In order to get high efficiency transition of the ions, the RF voltages of the first quadrupole in the second vacuum chamber and the second quadrupole in the third vacuum chamber are a little different with values about 400 V around 0.9 MHz and 110 V around 1 MHz, respectively. It should be noted that the octupole is used for guiding the ions into the low-temperature trap in the fourth vacuum. For keeping high efficiency transition in the short distance, the 141.6 mm long octupole consists of eight slim stainless-steel rods with the diameter of only 3.18 mm, and the field radius of the octupole is designed to be 4.75 mm. The RF-only power supply for the octupole is about 40 V around 0.8 MHz. All of the quadrupoles and octupole are shielded by stainless steel shells to prevent leakage of RF electric fields. Using these quadrupoles and octupole to guide the ions, we can achieve a total transmission of more than 10% from the exit of the first skimmer to the entrance of the ion trap.
In order to get the high energy resolution of VMI spectrometer, it is necessary to lower the temperature of negative ions. In this work the 3D Paul trap (purchased from R. M. Jordan Company in Canada) is used because it can confine ions similar to a point source in both axial and radial directions, which is suitable for coupling with subsequent low repetition rate TOFMS. The ion trap is in good thermal contact with the cold head, which is cooled by a closed-cycle helium refrigerator (Sumitomo RDK-408D2; 1 W cooling power at 4 K). The temperature of the ion trap is measured with a calibrated silicon diode (Lakeshore DT670 ) and can be adjusted between 5 K and 300 K through a feedback heating system (LakeShore Model 335). The configuration of this ion trap has been descripted in previous studies [10, 31].
-
The principle of TOFMS has been widely discussed and applied in the negative ion PES and other fields [18, 32]. The initial spatial distribution and energy dispersion of ions are the main factors restricting the resolution of time-of-flight mass spectrometry [33]. Although the initial spatial distribution of ions can be completely focused by the double acceleration field structure of ``Wiley-Mclaren" [34], the energy dispersion of ions cannot be compensated due to the turn-back time. The reflection electric field proposed by Alikanov [35] and Mamyrin [36, 37] can solve this problem very well. The second-order spatial focal point of the double acceleration field is used as a virtual ion source for the reflective electric field uses. Here there is no spatial dispersion for ions with the same mass, and the energy dispersion can be compensated by the reflective electric field. For achieving the effect of improving mass resolution, and considering the photo detachment and PE imaging of the negative ions, the mass spectrometer of this device is designed as adding a dual-reflection field on the basis of a linear double field accelerator, which can also effectively extend the free flight path of ions.
The design of the ion accelerator in this work is based on the double field structure of ``Wiley-Mclaren". As shown in FIG. 3(a), the accelerator is mainly composed of 15 electrode sheets, all of which are machined by the 304 stainless steel with a surface roughness of less than 0.8. Each electrode sheet is 1 mm thick and isolated by ceramic gaskets with a thickness of 4 mm. The first electrode plate (labeled 1 in the FIG. 3(a)) as shown in FIG. 3(b) is a square pole sheet with a side length of 120 mm and a round hole of 2 mm in diameter, which is used to collimate the instrument during installation by the collimating laser. The fifth electrode plate is design as FIG. 3(d), leaving a gap on the square electrode plate to facilitate the ion stream being thrown from the low temperature ion trap into the acceleration zone. The ninth and fifteenth electrode plates are design as FIG. 3(e), similar to the first plates except that the diameter of the center hole is about 20 mm and attached grids with transmittance of 90%. While the other eleven electrode plates are square hollow electrode plates in the same size with inner and outer side lengths of 90 and 120 mm, respectively, as shown in FIG. 3(c). The pulsed high voltage (HV) is applied on the electrode plate 1, at the same time the pulsed sub-high voltage (SHV) is applied on plate 9, and the electrode plate 15 is grounded (GD). Both of the accelerating electric fields are well-proportioned distribution by the 10 MΩ resistors parallel with 1 nF capacitors between each two electrode plates.
The dual ion reflectors are fixed on the same inclined plane to ensure that the included angle between the axis of the two reflectors and the ion path flying out of the accelerator is consistent, which is 3.43°. Both the ion reflectors are made up of 22 evenly distributed electrode plates with a thickness of 1 mm, and each of them is insulated by a ceramic gasket with a thickness of 9 mm. The electrode plates are similar to the accelerator field plates as shown in FIG. 3(c) with the inner and outer side lengths of 80 and 100 mm, respectively. The length of each reflector is three times that of the accelerator. And the spatial focusing point of the second reflector is designed to be the photon detachment point of the VMI system, which is based on the multilens design by Wang’s group for the high-resolution PE imaging [38]. Both of the ion reflectors, similar to the accelerator, are double-field electrostatic fields. The first, tenth and 22nd electrode plates facing the direction of the ions are grounded, DC sub-high voltage and DC high voltage, respectively. In order to distribute the voltage evenly to each piece, we divide the voltage between two pole pieces with 2 MΩ resistor with one-thousandth accuracy. In order to prevent the high-voltage electrostatic field of the reflectors from interfering with the flight trajectory of the ion beam, the reflectors are ground shielded by copper shell.
In particular, the cooled ions, after being thrown out of the low temperature ion trap, will be accelerated to a high kinetic energy, then fly a distance of about 3730 mm, and finally hit the detector. Because of the high kinetic energy of ions flying long time and long distance, it is almost impossible to get the ions fling right all at once time. The signal should be debugged step by step.
-
The negative ions are produced from the electrospray, where a microprocessor-controlled syringe pump is used to deliver the liquid samples which contain the target anions through a 10 μm i.d. fused quartz capillary directly to the ambient atmosphere at room temperature. The voltage between −2000 and −2500 V is applied on a stainless-steel needle which is connected to the capillary. The solution delivery rate is 60 μL/h, typically. The copper block containing the desolvation capillary is heated to 80–100 °C to produce the bare anions. The RF voltage is adjusted to optimize the transmission of a given mass range.
To satisfy the vacuum requirements of the ion trap operation and ion transmission, the four stages differential chamber is used for the RF-only ion guide system in this new ESI-PES apparatus. The RF-only ion guide is also differential designed and isolated by three skimmers with 1.5 mm diameter hole. To detect the effect of ion transmission, the 1 mmol/L NaI ethanol solution was prepared to deliver into the electrospray ion source. The current values produced by the anions were measured at the position of first, second and third skimmer through the Picoammeter (Keithley Series 6485/E), respectively. The measured current values were −500, −250, and −120 pA at first, second and third skimmer, respectively. The ion beam was transmitted through a 12 mm diam hole on the entrance end cap and finally into the ion trap, in which the current value was measured about −70 pA. That means the number of anions entrance into the trap is about 108 per second for single charge negative ions. According to the change of current value in each stage, the ion transmission rate was estimated to be about 50%.
-
FIG. 4 shows the TOF mass spectra of the Br− and I− at room temperature by spraying the 1 mmol/L mixture of NaBr/NaI in an ethanol solvent. The mass spectrum of p-iodophenol anion (p-C6H4IO−) with mass number 219 is also shown in FIG. 4. The p-C6H4IO− anions were produced using ESI from sample solutions of p-C6H4IOH dissolved in mixed solvents of CH3OH/H2O at pH≈10. The electric parameters were optimized to obtain the best performance of TOFMS. According to the relationship M/ΔM = t/(2Δt), as indicated in FIG. 4, the mass resolution (M/ΔM) can be achieved about 2243 and is good enough to resolve the 13C isotope mass peak of p-C6H4IO− for which the difference of mass number is 1. The delay time relative to the ion trap ejection pulse has to be tuned and optimized for different mass ranges, because the flight time of light ions from ion trap to the center of perpendicular acceleration region is shorter than that of heavy ions. The configuration of dual-reflection lens is to satisfy the kinetic energy focusing of ions in the flight route. Thus, the high and sub-high voltages of both sets of reflection lens are −1500 and −900 V (DC), which is in accordance with configuration of perpendicular acceleration field. There are two key factors affecting the mass resolution. The first is the kinetic energy of the ions exiting the ion trap, that is the transverse kinetic energy in the accelerate field region, the smaller transverse energy the better the resolution, since ions with a smaller transverse energy can maintain more straight trajectories and do not need to be deflected much. The second factor is the size of the ion beam in the extraction zone. The smaller the size in the perpendicular direction to the ion extraction the better the resolution. With our current parameters, the nominal kinetic energy is about 5 eV giving a spread of the beam size less than 3 cm in the extraction zone.
The image of I− is also recorded as shown in FIG. 5 to test the spatial focusing performance of the new mass analyzer. Since the phosphor screen is located at 505 mm behind the ions focal point, the ions will disperse to a certain extent when arriving the imaging detector. The raw image is shown in FIG. 5(a), which is a two-dimensional matrix with size of 1032 × 1032. The effective diameter of the MCP in front of phosphor screen is 77 mm (HAMAMATSU F2226) corresponding to 940 pixels in the image of the CCD camera (BASLER, avA 1000-100 gm). FIG. 5 (b) is obtained by integrated matrix data along the X-axis of raw image. FIG. 5(c) is obtained by integrated matrix data along the Y-axis of raw image. As indicated in FIG. 5 (b) and (c), the full with of maxima is about 300 pixels, which means the size of focusing ions is about 2.4 cm. It illustrates that the spatial focusing performance of the new mass analyzer is favorable.
-
The configuration of our VMI lens system is based on the multilens design by Wang and co-workers [38]. It can perform more careful field shaping and shielding to achieve optimal focusing. As shown in previous studies, such a configuration can achieve almost perfectly circular and concentric images for electrons with different kinetic energies. Images of such quality do not require post-imaging corrections prior to inverse-Abel transformation and is helpful to achieve high-resolution imaging. The performance of VMI system in our new ESI-PES apparatus has been tested using PE images of I− and p-IC6H4O−.
FIG. 6 shows the PE image taken from photodetachment of I− at photon energy of 0.59 eV, with detachment laser wavelength at 340 nm, above the threshold of the I (2P3/2) state for which the electron affinity (EA) is 3.059 eV [39], and the corresponding transition is I (2P3/2)←I− (1S0). The full width at half maxima (FWHM) corresponding to the electron kinetic energy (KE) is 5.8 meV at 0.59 eV. And the FWHM for the image is only about 1.3 pixels. The KE is proportional to the square of the image radius (R), the relationship between R and KE can be expressed as KE = a×R2, in which a is a constant. The relative resolution (ΔKE/KE) obtained is 0.98%, which is comparable with the report of previous study [38]. The spectrum was taken with repel voltage at −650 V.
FIG. 7 shows the PE image and spectrum of p-C6H4IO anion cooled at 5 K in a cryogenic ion trap with detachment laser wavelength at 450.08 nm. This spectrum is used to showcase the capabilities of the spectrometer effectively for the molecular ion at a low temperature. As shown in FIG. 7, there is no hot band obtained, and the EA of p-C6H4IO radical is measured to be 2.609 eV corresponding to the maximum position of peak X. Peaks A to H show the vibrational features of the neutral sample. The binding energies of peaks A to D are determined to be 2.640 (111), 2.669 (112), 2.681 (101) and 2.698 (113) eV, respectively. These vibrational and the EA values are in good agreement with the previous report [40]. More vibrational information is obtained in this spectrum. The peaks E and H at 2.712 and 2.742 eV are assigned to the combination bands of modes 101111 and 101112, respectively. The peak G at 2.736 eV which is offset by 1024 cm–1 is assigned to the breathing mode of benzene ring.
-
In conclusion, an electrospray PES instrument with a cold ion trap, a dual-reflection TOFMS and a high-resolution VMI technique has been developed to investigate solution phase species. The dual-reflection mass analyzer has high quality mass resolution at 2243, which is good enough to resolve the 13C isotope, and is beneficial to the separation of actinide elements containing clusters such as U and Th. The relative resolution of the VMI spectrometer is measured to be 0.98%, it suggests that the performance of the VMI system is as excellent as others’ high-resolution imaging spectrometers. This facility will provide new research opportunities to investigate many novel solution phase species, such as lanthanide and actinide compounds, metal modified fullerene and biomolecular species, in the gas phase.
-
The authors declare no competing financial interest.
-
This work was supported by the Scientific Instrument Developing Project of the Chinese Academy of Sciences (No.YJKYYQ20190027), Youth Innovation Promotion Association, Chinese Academy of Science (No.2021255), Shandong Energy Research Institute Enterprise Joint Fund (SE1 U202312), ``Strategic Priority Research Program" of the Chinese Academy of Sciences (No.XDA02020000), the National Natural Science Foundation of China (No.22273065).
Development of High-Resolution Photoelectron Imaging with Dual-Reflection Time of Flight Mass Analyzer†
- Received Date: 28/12/2023
- Available Online: 27/04/2024
-
Key words:
- High-resolution photoelectron imaging /
- Dual-reflection time of flight /
- Photoelectron spectroscopy
Abstract: A velocity map imaging photoelectron spectroscopy, coupled with an electrospray ionization source and a dual-reflection time of flight (TOF) mass analyzer has been developed for the investigation of anions in the gas phase. Anions formed in the electrospray source are guided by a radio-frequency quadrupole ion guide into a quadrupole ion trap, where the ions are accumulated. A unique feature of this apparatus involves the coupling of a dual-reflection mass analyzer to the original Wiley-McLaren design. This can effectively improve the mass spectrum resolution, so as to accurately select the target ion. The mass resolution (M/ΔM) above 2000 can be achieved. The velocity map imaging spectrometer resolution is about 5.8 meV full width at half maximum for the photoelectrons with the kinetic energy of 0.59 eV, leading to a relative resolution of ~0.98%. The detailed design, construction, and operation of the new apparatus are presented.