Fourier transform spectrometer (FTS) is a mature detection system for broadband spectroscopy [1]. It has the characteristics of simultaneously detecting multiple gas components, non-invasive measurement without destroying the molecular structure, and full-range spectrum corresponding to the detector range in one measurement, making it very suitable for quantitative analysis of complex components [2]. Michelson-based FTS obtains interferograms by recording the intensity of light as a function of optical range difference (OPD) with a scanning interferometer, which is then Fourier transformed to obtain molecular spectra. The advantages of FTS include its compatibility with broadband and mid-infrared light as well as high spectral resolution that can be achieved with a relatively compact setup [3]. This is particularly advantageous for measurements in the mid-infrared molecular fingerprint region, where most molecules have strong vibrational-rotational transitions, but the Doppler profiles of the infrared lines are narrow. However, traditional FTS typically uses incoherent thermal light sources as probe light, which makes it difficult to achieve long interaction length in gas-phase sample detection, resulting in low sensitivity and requiring a long-term averaging to improve sensitivity [4]. Moreover, the omnidirectional and low spectral brightness can limit the accuracy of the measurements at infrared lines wavelengths [5, 6]. These factors limit the use of FTS in modern applications, especially for rapid and simultaneous monitoring of atmospheric trace gas concentrations and chemical reaction kinetics, noninvasive breath analysis, and other applications that require high sensitivity real-time detection [7, 8].
The invention of optical frequency comb (ORC) [9] provides a feasible broadband light source for the highly sensitive FTS. The OFC is a phase-stabilized femtosecond pulsed laser, whose formation relies on a nonlinear optical process known as mode locking, which forces equal frequency spacing and establishes phase synchronization between modes to form a short optical pulse [10]. This pulse propagates back and forth between the end mirrors of the laser resonator at its group velocity. At each round trip, a portion of the pulsed light is transmitted through one of the partially transparent mirrors. It is well known that the Fourier transform between the time and frequency domains means that the spectrum of a periodic pulse train produced in this way is actually a discrete frequency comb [11]. The OFC was originally developed to solve the problem of measuring and counting optical frequencies according to the Système Internationale d'Unités (SI) cesium definition. Due to its temporal and spatial coherence, it has the advantages of wide spectral coverage, high spectral brightness and good coherence, and its application range has far exceeded the frequency measurement [12, 13]. At present, OFCs have become an important light source for broadband molecular spectroscopy [8, 14−16]. FTS based on the OFC can use multi-pass cell or F-P cavity as absorption cell, which has high sensitivity, accuracy and consistency in concentration measurements. The broad spectral bandwidth allows for the detection of multiple species, as well as more reliable concentration retrieval, and its applications range from atmospheric trace gas monitoring to industrial process control or leak detection [17, 18]. Besides, the combination of its short acquisition time and broad spectral bandwidth can drive the development of time-resolved spectroscopy, which can be used for the study of reaction kinetics, especially gas-phase reaction kinetics [19−22].
Several demonstrations of OFC-based FTS for the detection of trace gases have been reported in the past. Schliesser et al. first experimentally demonstrated the feasibility of OFC-based FTS [17]. In 2007, Sorokina et al. directly incidented a femtosecond laser beam with a center wavelength of 2.4 μm through a gas cell into a commercial Fourier transform infrared (FTIR) spectrometer, which had a resolution of 0.12 cm−1. The measured bandwidth was 380 cm−1 and the minimum detectable absorption (MDA, determined by the noise equivalent absorption after optical path normalization, spectral acquisition time and the number of resolved spectral elements) was 2.4×10−7 cm−1·Hz−1/2 per spectral element [23]. After that, Mandon et al. utilized this technique near 1.5 μm, in conjunction with higher spectral resolution FTS (~0.006 cm−1) and a longer absorption path (~40 m) to achieve a MDA of 3.9 × 10−8 cm−1·Hz−1/2 per spectral element [24]. Adler et al. carried out rapid detection of trace gases in the mid-infrared molecular fingerprint region using a broadband (2100–3700 cm−1) optical parametric oscillator (OPO) frequency comb in a Herriott-type multipass cell. The instrument had an absorption sensitivity of 3.8 × 10−8 cm−1·Hz−1/2 per detection element at a resolution of 0.014 cm−1 [25]. In recent years, Foltynowicz’s group has measured the mid-infrared molecular broadband spectra of CO2, CH3I, N2O, and CH4 molecules using an OFC-based high spectral resolution FTIR spectrometer, which yield the precise positions of the spectral lines and the interaction parameters for different molecular spectral bands [26−30].
Although frequency comb Fourier transform spectrometers have been used for spectroscopic studies, due to the complexity and high cost of mid-infrared OFC systems, the reports on mid-infrared OFC-based FTIR spectrometer are still scarce, and most of the spectroscopic work has been carried out in the near-infrared (NIR) region where the technology is relatively mature [8]. For molecule sciences, however, the mid-infrared band is of much more attraction region because it contains the strong, characteristic, and fundamental vibrational resonances of many compounds of interest [18]. Therefore, in this work, we construct a FTIR spectrometer based on mid-infrared OFC, and several characteristics of the setup are evaluated. With this setup, we measure the mid-infrared absorbance spectrum of individual molecules such as CH4, H2O, and N2O. A full bandwidth fit of the absorbance of CH4 is performed, and the fitting residual is 4.3×10−7 cm−1, corresponding to the MDA of 4.4×10−8 cm−1·Hz−1/2 per spectral element. The consistency of the concentration measurements is obtained to be 1% by multiple measurements of the H2O absorbance spectra. Finally, the center positions of the (0002)–(1000) band of the N2O molecule are determined, and the upper state constants of the (0002)–(1000) transition lines are derived, and the results are in good agreement with those reported by Toth, which is the source of the HITRAN data for these bands.
-
FIG. 1 shows the schematic of the experimental setup, which consists of three parts: an optical frequency comb, an absorption cell, and a Fourier transform infrared spectrometer.
The mid-infrared optical frequency comb, purchased from Menlo System, model FC1500-250-WG and Mid-IR comb, generates mid-infrared light pulse by difference frequency generation (DFG). The seed source of mid-infrared optical frequency comb is an Erbium-doped femtosecond fiber oscillator with a central wavelength of 1.5 μm and a repetition frequency of 250 MHz. The light pulse output from the Erbium-doped fiber oscillator is divided into two paths by a beam splitter, one path is converted into 1 μm by Raman-induced solitons self-frequency shift of highly nonlinear fiber, and this pulse light is amplified as the pump light; the other one path comes directly from the seed source to provide the signal light. Both fundamental lights have to be synchronized in time and space, and then combined at a dichroic mirror before being collinearly focused on the PPLN crystal, resulting in a free of carrier-envelope mid-infrared idler. As the fundamental signals are not completely converted, a long-pass filter blocks them from separating the DFG signal from the fundamentals. Benefiting from the wide tuning range of the soliton, the mid-infrared optical frequency comb has a wide tuning range as well. In our experiments, the mid-infrared optical frequency comb has a spectral coverage range of 2800–3500 cm−1, with a total output power of up to 180 mW, and the linewidth of the comb mode is about 2 MHz. In addition, to obtain a stable frequency comb, the repetition rate is locked to the rubidium clock reference source in the laboratory.
The outgoing mid-infrared pulse is collimated into the Herriott-type multipass cell [31], which consists of two circular spherical mirrors. The mid-infrared light is reflected back and forth 49 times in the absorption cell, and the effective optical path is about 60 m. The absorber cell has two connections, one is the inlet port and the other is the extraction port. The sample gas enters the cell through a flow meter, and there is a multi-function ball valve at the inlet, which can be used to change the gas species by switching the direction of the multifunctional valve. The pumping port is connected to a pressure gauge, butterfly valve and vacuum pump, which are used to suction the absorption cell and control the pressure inside the absorber. In this experiment, we controll the pressure in the absorber to be about 630 mbar, and use zero air purge for several hours before each measurement to eliminate the residual gas interference attached to the inner wall of the multipass cell.
After passing through the cell, the mid-infrared beam is focused by an achromatic lens and enters the FTIR spectrometer (Vertex 80v, Bruker Ltd.). The FTIR is a Michelson-type air-bearing interferometer with built-in KBr splitter, which can measure the spectrum in the range of 8000–350 cm−1 with an optimal resolution of 0.06 cm−1. The detector is a mercury cadmium telluride (MCT) detector for the mid-infrared band, which requires liquid nitrogen cooling before use. The moving mirror of the interferometer and spectral wavenumber calibration are controlled by a 633 nm frequency-stabilized He-Ne laser. The interior of the spectrometer is evacuated (~2 mbar) to prevent water vapor and carbon dioxide from interfering with the IR spectra. In this experiment, Norton-Beer apodization function is chosen to suppress the sidelobe amplitude and obtain high-sensitivity spectrum, and Forman convolution method is chosen to reduce the phase error caused by the sampling drift of interferogram data points [32].
The measurement of molecular spectra using a FTS requires two steps [1]. First, zero air is measured as the background gas to measure the interferogram for Fourier transform, which serves as the reference spectrum
$ R\left(\nu \right) $ . Then the sample is injected into the absorption cell, keeping other parameters constant, and the sample spectrum$ S\left(\nu \right) $ is measured. The measurement time of each spectrum is about 35 s. The sample spectrum is similar to the background spectrum, showing intensity variations only at the wavenumbers with absorption. The transmission spectrum of the molecule can be obtained by dividing the sample spectrum by the reference spectrum. However, for the sake of comparison, we convert the transmission spectrum into absorbance spectrum according to Lambert-Beer’s law, and normalize the optical path to get the final spectral absorbance$\mathrm{\alpha }\left(\mathrm{\nu }\right)=-{L}^{-1}\mathrm{l}\mathrm{n}\left(S\right(\nu )/R(\nu \left)\right)$ , where$ L $ is the optical path length inside the absorption cell. Due to the performance of the FTIR spectrometer changes during long-term operation, the light energy measured by the detector will change, which can cause drift, tilt and concave phenomena of the absorbance spectrum, and affect the quantitative analysis, so the baseline correction of the absorbance must be made [28]. This can be done by masking the absorption lines, then fitting the remaining spectral structure using a sum of fifth-order polynomials and trigonometric functions. Trigonometric function is used to eliminate interference fringes that are not normalized by the background, and the measured spectrum is subtracted from the fitting baseline to get the final spectrum. -
High-resolution measurements of broadband spectra of molecules are an important way to distinguish and quantitatively analyze individual species, which is important in many applications [33]. FIG. 2 shows the absorbance spectra of several molecules, including water vapour (H2O), nitrous oxide (N2O), methane (CH4), ethane (C2H6), nitrogen dioxide (NO2), and acetylene (C2H2), measured in the interval from 2800–3500 cm−1. These gases are diluted by zero air, and water molecules are brought into absorption cell by zero air via a set of humidification bottles containing a small amount of distilled water. In the experiment, the resolution of the spectrometer is set to 0.08 cm−1 (2.4 GHz), and each spectrum is averaged 16 times, with an acquisition time of about 35 s. The black line in the figure shows the experimental measurements and the red line shows the results of the simulation using HITRAN with the same experimental parameters [34], from which it can be seen the spectral structure of the measured individual molecules is consistent with the database simulation results. Because of the different line strength of the molecules in this region, the concentration of different molecules varies in order to avoid absorption saturation. For CH4, C2H6, and C2H2, the line strengths are at the level of about 10−19 cm−1/(molecule·cm–2), so lower concentrations (a few ppm) are used. For the N2O molecule, the line intensity is at about 10−24 cm−1/(molecule·cm–2) level in this band, thus using a higher molecular concentration. Here we have chosen to measure the infrared absorption spectra of only some of the gases extant in the lab. These measurements show that regardless of the structure of the absorption spectrum, the instrument is able to quickly cover a wide spectral region and extract concentrations at high resolution. This setup could also be used to measure other species, especially some compounds containing carbon-hydrogen bonds, which have strong absorption lines in this region [35].
-
A significant advantage of OFC-based FTS over conventional FTIR spectrometer is the high spectral brightness and good spectral coherence of the OFC, which allows the use of a multipass cell or an F-P cavity as the absorption cell to increase the detection sensitivity of the spectra and reduce the spectral acquisition time. Therefore, we evaluate the detection limit of the device. According to the literatures [15, 36], it is known that when multiple absorption signals with a high bandwidth are measured simultaneously using a spectrometer, the detection limit expression contains features characterizing the measurement bandwidth and spectral resolution of the spectrometer and defines the minimum detectable absorption (MDA) of each spectral element, which is equal to
$\sigma \sqrt{T/M},$ where$ \sigma $ is the noise equivalent absorption (NEA) normalized to the optical range in cm−1, T is the spectral acquisition time, and M is the number of resolvable spectral elements, which is obtained by dividing the spectral bandwidth by the spectral resolution.To facilitate the evaluation of the instrumental detection limits, the absorbance spectra of CH4 are selected and fitted using the global fitting least squares technique [37]. The standard deviation obtained by fitting was 4.3×10−7 cm−1, which can be considered as the noise equivalent absorption (NEA) after optical path normalization. FIG. 3 shows the multi-spectrum fitting results. The spectrum span is about 300 cm−1, between 2875 and 3175 cm−1, contains 3750 spectral elements, and the spectral acquisition time is about 35 s, so the minimum detectable absorption is equal to 4.4 × 10−8 cm−1·Hz−1/2 per spectral element. Using the maximum signal-to-noise ratio (~80) of the absorption lines in FIG. 3, a detection limit of 16 ppb is allowed for the molecular concentration of CH4 in 35 s. Table I lists the results of different groups using the multi-pass cell as absorption cell. It can be seen from the table that there are still some differences between our results and those of some groups, which are due to the relatively short acquisition time, and the resolution of our spectrometer currently used is still poor, resulting in fewer resolvable spectral elements. According to the Ref.[24], this result is about 13 times more sensitive than using a traditional incoherent light source, such as white lamp, which means that the acquisition time of the spectrum will be increased by 169 times or more to obtain the same result with traditional FTIR. In the future, with the F-P cavity, the detection limit can be reduced by another three orders of magnitude, which makes the OFC-based FTS suitable for the detection of most trace gases.
-
Another feature of spectroscopic measurements using OFC-based FTS is the good reproducibility of its measurement results, which is important for quantifying the properties of the species analyzed, and in particular for quantifying the concentration information of the species [38]. Since the Fourier transform spectrometer is calibrated using the absorption spectrum of water vapor as a standard, and the spectral line strengths of water molecules near the transition line in this band vary greatly, a certain single line can be used to invert the molecular concentration, so we take H2O as an example. The H2O molecules are brought into the absorption cell using the zero-air bubbling method while keeping the parameters of the spectrometer unchanged, and then the absorption spectra of H2O are measured five times independently. When retrieving the concentration of H2O, we only select the absorbance spectra of part of H2O, as shown in FIG. 4, and perform multispectral line fitting on it, and then select one of the spectral lines with less interference, as shown in FIG. 4(b), calculate the integral area under the spectral profile, and deduce the concentration of H2O according to Lambert-Beer’s law. After five independent measurements, the concentration of H2O is 495 ppm with a standard deviation of 5 ppm, which give a repeatability of about 1% for H2O concentration measurement.
-
An important application of FTIR spectrometers is their ability to measure broadband spectra of simple molecules with an absolutely calibrated frequency scale. Such measurements allow rigorous testing of fundamental theories and ab initio quantum-chemical calculations to determine physical constants precisely, and are one of the important avenues of research in the field of molecular spectroscopy [3, 39, 40]. Here, limited by the resolution of FTIR, we have chosen the (0002)–(1000) band of N2O molecules with a large transitions line interval as the study band. The absorbance spectrum of the (0002)–(1000) band near 3132 cm−1 of the N2O molecule is shown in FIG. 5, which consist of the P-branch and the R-branch, and the nested plot in FIG. 5 shows the full-bandwidth absorbance spectrum of N2O. By fitting the measured spectra line-by-line, 84 resolvable spectral lines can be obtained, of which 39 are in the P-branch and 45 are in the R-branch. According to HITRAN database, the line intensity of N2O in this region is known to be between 2.2 × 10−25 and 2.7 × 10−24 cm−1/(molecule·cm–2).
PGOPHER is used to calibrate the central wave-number of each absorption line, which is a versatile tool for molecular spectroscopy simulation and fitting [41, 42]. Here, we first use the peak marking function in PGOPHER to roughly mark the position of each absorption peak, and then use Voigt profile to fine fit the intensity, line width and central position of each absorption peak, and minimize the fitting residual error of the full spectrum through continuous iteration, thus obtaining the exact central position of each absorption peak. The fitted residual is used as the uncertainty of the central position. Because each absorption peak has good symmetry, it can be considered that the fitting central wavenumber is accurate. The linewidths of the individual spectral lines are broadened by the pressure in the absorption cell of 630 mbar, which allows the spectrometer to distinguish the line widths of each absorption peak. According to Toth [43], the pressure frequency shift coefficient of N2O is about –0.002 cm−1/atm in this band. In the experiment, this corresponds to a pressure-induced frequency shift of about 0.0012 cm−1, which has been taken into account in the results. The final wavenumbers of each transition line are listed in Table II, and their wave number uncertainties are within 0.001 cm−1.
For further comparison, we give the difference between the measured line position and the HITRAN database, the source of HITRAN is based on the Toth’s result. FIG. 6 shows the comparison results, from which we can seen that the measured results are basically consistent with the results of Toth, but there are some differences (within 0.008 cm−1), which is also the range of the spectral line position accuracy of FTIR we used. In the future, with the use of higher precision spectrometers, the accuracy will be further improved.
The vibrational transition of a linear molecule can be expressed using the following equation [44],
Among them, the
${G}_{\nu}$ and$ {G}_{0} $ are the vibration energy of the upper and lower levels,$ {B}_{\nu } $ and$ {B}_{0} $ are the rotation constants of the upper and lower levels,$ {D}_{\nu } $ ,$ {H}_{\nu } $ and$ {D}_{0} $ ,$ {H}_{0} $ respectively are the four and six centrifugal distortion constants of upper and lower levels,${J}{'}$ and${J}{''}$ are the fluctuation of the angular momentum quantum number. In order to simplify the fitting process, the rotation and centrifugal distortion constants of the ground state are fixed to the values determined by Toth [43], and then the constants of the molecular upper state are determined by the nonlinear least squares global fitting method. Since the fitting is nonlinear, it is necessary to repeat the fitting for several cycles to converge. The fitted upper state constants of the (0002)–(1000) band are presented in Table III. We compared them with Toth’s results, and the results of our measurements are in good agreement with Toth’s. However, the fit standard deviation of our results is large because the uncertainty in the central position of each absorption line is poor, about 0.001 cm−1 (30 MHz). In the future, by switching to higher-resolution spectrometers, we can further improve the fitting accuracy of spectral transition constants and provide support tools for molecular precision spectroscopy. -
In this work, we have built a mid-infrared OFC-based FTS using OFC produced by differential frequency generation. We have analyzed several properties of the spectrometer, which demonstrate that the instrument enables the determination of spectral line position and upper state constant. The MDA of the instrument is 4.4×10−8 cm−1·Hz−1/2 per spectral element obtained by full-bandwidth multi-spectrum line fitting, which is essentially the same as the known results using a multi-pass cell. We have also evaluated the consistency of the concentration retrieval performed by the instrument and obtain a concentration uncertainty of about 1%. Furthermore, the spectral line positions of the (0002)–(1000) band of the N2O molecule are measured, and the upper state constants of the band are determined by PGOPHER, and the results are in good agreement with Toth’s.
OFC-based FTIR spectrometer has many advantages and its application potential is great. However, limited by the experimental conditions, we only show some of the advantages of this technique here. As the spectral coverage, optical intensity, and ease of use of optical frequency combs continue to improve, the corresponding spectrometer is also developing rapidly. In the future, with high-resolution spectrometers [45] and F-P cavities with high finesse and wide spectral coverage [46], we will further improve the sensitivity and resolution of OFC-based FTS and apply it to trace gas detection and free radical reaction kinetics.
Mid-infrared Optical Frequency Comb-based Fourier Transform Spectrometer for Broadband Molecular Spectroscopy
- Received Date: 11/01/2024
- Available Online: 27/08/2024
-
Key words:
- Mid-infrared optical frequency comb /
- Multi-pass cell /
- Fourier transform infrared spectrometer
Abstract: Optical frequency comb-based Fourier transform spectroscopy has the features of broad spectral bandwidth, high sensitivity, and multiplexed trace gas detection, which has valuable application potential in the fields of precision spectroscopy and trace gas detection. Here, we report the development of a mid-infrared Fourier transform spectrometer based on an optical frequency comb combined with a Herriott-type multipass cell. Using this instrument, the broadband absorption spectra of several important molecules, including methane, acetylene, water molecules and nitrous oxide, are measured by near real-time data acquisition in the 2800–3500 cm−1 spectral region. The achieved minimum detectable absorption of the instrument is 4.4×10−8 cm−1·Hz−1/2 per spectral element. Broadband spectra of H2O are fitted using the Voigt profile multispectral fitting technique and the consistency of the concentration inversion is 1%. Our system also enables precise spectroscopic measurements, and it allows the determination of the spectral line positions and upper state constants of N2O in the (0002)–(1000) band, with results in good agreement with those reported by Toth [Appl. Opt. 30, 5289 (1991)].