Second harmonic generation (SHG) is the first nonlinear optical effect, discovered in 1961 [1]. Over the years, it has found widespread utility in various scientific domains, including physics, chemistry, materialogy, and biological imaging [2−7]. SHG is restricted to materials with non-central inversion symmetry [8]. It possesses remarkable sensitivity to symmetry, making polarization-dependent SHG preferred technique for characterizing symmetry [9−15]. The popularity of polarization-dependent SHG has surged in research focused on advanced materials, including two-dimensional materials [16], ferroelectric materials [17], and perovskite [18]. Precise polarization control plays a crucial role and significantly influences the accuracy and reliability of the result when the experiment is performed [19, 20].
Polarization-dependent SHG can be categorized into two geometries: transmission and reflection [8]. The polarization-dependent SHG signal is usually obtained by simultaneously changing the polarization azimuth of the fundamental and harmonic beams. In transmissive geometry, as FIG. 1(a) shows, precise polarization control mainly focuses on the fundamental and harmonic beams. The reflective geometry is divided into non-collinear [21] and collinear [12], as shown in FIG. 1 (b) and (c), although the noncollinear geometry is more general and the polarization of its signal is not affected by the beam splitting device [22−25], but the precise control of the incidence angle in this geometry poses challenges, particularly in micro-area detection, primarily due to the limitations imposed by using an objective lens. For collinear geometry, beam-splitting devices are necessary, such as beam splitters (BS) and dichroic mirrors (DM), which inevitably distort the polarization state of both fundamental and harmonic beams. For non-transparent micrometer scale samples, collinear reflective geometry becomes essential, and it is imperative to eliminate the impact of beam-splitting devices to ensure accuracy and precision.
For collinear reflective polarization-dependent SHG, the Fresnel coefficient of the beam-splitting device needs to be considered [26, 27]. In this work, through theoretical analysis and experiment, we analyze the effect of the BS on the SHG signal in various cases and propose a data processing method that can remove the effect of the BS to obtain the SHG signal as well as normalized second-order nonlinear susceptibility tensor.
-
For the sample, we establish an X-Y coordinate system. The expression for the second-order polarization is given by [8]:
where
$\vec{P}^{\left(2\right)}$ is the second-order nonlinear polarization,${\varepsilon }_{0}$ is the vacuum permittivity,${\mathop{\chi}\limits^{\leftrightarrow}}^{(2)}$ is the second-order nonlinear susceptibility tensor, and$\vec{E}$ is the electric field of the fundamental beam. When the laser is normally incident on the sample surface, the electric field directions lie within the sample plane, which is described by the P-S coordinate system. So, all tensor elements with a z-direction do not participate in the SHG process. Due to the presence of intrinsic permutation symmetry, the number of effective and independent second-order nonlinear susceptibility tensor elements is at most six, namely,$\chi_{xxx}^{\left(2\right)}$ ,$\chi_{xyx}^{\left(2\right)}$ (or$\chi _{xxy}^{\left(2\right)}$ ),$\chi_{xyy}^{\left(2\right)}$ ,$\chi _{yxx}^{\left(2\right)}$ ,$\chi _{yxy}^{\left(2\right)}$ (or$\chi_{yyx}^{\left(2\right)}$ ), and$\chi _{yyy}^{\left(2\right)}$ . The ratio between them reflects the symmetry information of the crystal, and their magnitudes determine the intensity of the SHG signal.Considering the polarization distortion by BS, Eq.(1) can be written as
$F_\omega$ and$F_{2\omega}$ are the Fresnel coefficients of the fundamental beam and the harmonic beam respectively [21]. In the experimental setup described in this article,$F_\omega $ is the transmission Fresnel coefficient, while$F_{2\omega} $ is chosen as the reflection Fresnel coefficient.By fixing the orientation of the polarizer as P and S, respectively, the SHG signal intensity as a function of the fundamental beam polarization azimuth can be obtained as:
${\mathop{\chi}\limits^{\leftrightarrow}}^{(2)}$ elements expressed in the crystal coordinates (x, y) are related to those in the lab coordinates (p, s) by [26]with
$i,j,k\in \left\{x,y\right\},l,m,n\in \left\{\rm{p},\rm{s}\right\}$ .In Eq.(3), with the knowledge of the Fresnel coefficients, the normalized second-order nonlinear susceptibility tensor elements becomes achievable through the measurement of the fundamental and harmonic beams.
To provide a more intuitive understanding of the impact of polarization distortion from BS on results, we have selected the
$D_{3\rm{h}}$ point group as a model system, and monolayer MoS2 belongs to this point group. The relation of the nonzero second-order nonlinear susceptibility tensor elements is$\chi_{yyy}^{(2)}=-\chi_{xyx}^{(2)}\left({\rm{or}}\; \chi_{xxy}^{(2)}\right)= -\chi_{yxx}^{(2)}$ . In FIG. 2, we present a representation of the simulated distortion of the six-fold SHG signal resulting from the effects of polarization distortion. As we can see the blue figure at the center, when$F_{\omega ,\rm{p}}$ =$F_{\omega ,\rm{s}}$ ,$F_{2\omega ,\rm{p}}$ =$F_{2\omega,\rm{s}}$ , the SHG signal exhibits a perfect six-fold symmetry. The other cases exhibit polarization distortion, leading to distortion, as depicted by the pink pattern. -
For the experiment, we built the polarization-dependent SHG setup in FIG. 3, we utilize an 808 nm laser from an 80 MHz Ti:Sapphire oscillator (MaiTai HP, Spectra-Physics) as the fundamental beam. To achieve the desired polarization, the laser beam first passes through a linear polarizer (GL10, Thorlabs) and a half-wave plate (WPH10M-808, Thorlabs). After being directed through a BS, the laser beam is focused onto the sample using an objective lens (×40/NA 0.60, Olympus). A monolayer MoS2 on a quartz substrate is chosen as the sample. At the same time, the objective lens also collects the SHG signal from the back of the sample. The SHG signal is reflected by the beam-splitting devices, and passed through linear polarizers (GL10, Thorlabs), and short-pass filters (FF01-451/106-25, Semrock and FF01-405/150-25, Semrock), then captured by a PMT (H7422-40, HAMAMATSU). The SHG signal is obtained by simultaneously rotating the half-wave plate and the polarizer in front of the detector. During data processing and correction, our tests were performed by rotating the half-wave plate and fixing the orientation of the polarizer in front of the detector in the S and P directions, respectively. By fitting the test results with Eq.(3), we can obtain the second-order nonlinear susceptibility tensor elements, and then obtain the polarization-dependent SHG signal without the polarization distortion influence of the beam splitter.
As an illustration, we devise four distinct experimental schemes, taking the BS as an example. In the first case, which is the ideal situation with the Fresnel coefficients of
$F_{\omega,\rm{s}}=F_{\omega,\rm{p}},\quad F_{2\omega ,\rm{s}}=F_{2\omega,\rm{p}}$ , no existing BS satisfies this condition, necessitating a transmissive polarization-dependent SHG test. In the second, third, and fourth cases, the Fresnel coefficients align as$F_{\omega,{\rm{s}}}\ne F_{\omega,{\rm{p}}},\quad F_{2\omega,{\rm{s}}}=F_{2\omega,{\rm{p}}}$ ;$F_{\omega,{\rm{s}}}=F_{\omega,{\rm{p}}},\quad F_{2\omega,{\rm{s}}}\ne F_{2\omega,{\rm{p}}};\quad F_{\omega,{\rm{s}}}\ne F_{\omega,{\rm{p}}},\quad F_{2\omega,{\rm{s}}}\ne F_{2\omega,{\rm{p}}}$ , making them reflective setups. For these cases, we utilized BS013, BS014, and BSW10R which are purchased from Thorlabs as the respective BS, with specific parameters detailed in Table I. -
In the four aforementioned cases, we conduct separate tests on the raw and the corrected SHG signal using the same sample, as illustrated in FIG. 4. In FIG. 4(a), the SHG signal from case 1 (transmission) demonstrates adherence to the expected theoretical six-fold symmetry and is consistent with the results in FIG. 2(a). When applicable, the transmissive geometry is optimal because there are no beam splitters and therefore is not affected by polarization distortion. Subsequently, in FIG. 4(b–d), we observe the remarkable efficacy of our data correction method, yielding excellent results for signal correction. We can see in FIG. 4(d) that the standard sixfold symmetric signal is compressed in the vertical direction, which is consistent with the results in FIG. 2(d). FIG. 4 (b) and (c), on the other hand, both exhibit being compressed in the horizontal direction, which is also consistent with the results in FIG. 2 (b) and (c), respectively.
During the data correction process, we successfully derive the values of the normalized second-order nonlinear susceptibility tensor elements, as presented in Table II. To quantitatively evaluate the effectiveness of the correction, we employ the deviation function
$ \delta=\displaystyle\sum_i|\chi_i^{(2)}-\chi_0^{(2)}| $ to assess the error between each tensor element and its theoretical value for every case. The results demonstrate that the proposed data correction method effectively determines the normalized second-order nonlinear susceptibility tensor elements, minimizing the influence of the BS.To verify the generality of our methods, we conducted four tests on BiB3O6 (BIBO) crystals with the angle
$\theta=80^\circ,\quad \varphi=5^\circ$ . And this crystal is a monoclinic structure and belongs to the C2 point group. The results, illustrated in FIG. 5, further confirm the excellent applicability of the data correction techniques presented in this work.Similarly, to quantitatively describe the effect of this method, we calculate the size of the normalized second-order nonlinear polarizability tensor elements in each case and calculate the error from the standard value, as shown in Table III. It is worth noting that case 1 is selected as the standard value in this experiment. This is because the nonlinear optical coefficient of BIBO was exclusively measured at 1079.5 nm [28, 29], with no corresponding data at 808 nm. Through these experiments, it can be shown our data correction methods are widely applicable.
Based on the theoretical analysis and experiment, it is evident that an optimal beam-splitting device should be selected with a minimal difference in the Fresnel coefficients between the S and P directions at the wavelengths of the fundamental and harmonic beams. A prime example is BS013, in this work, showcasing exceptional applicability at both 808 nm and 404 nm. The distortion of the SHG signal is minimized, as shown in FIG. 4(b) and FIG. 5(b).
-
In conclusion, we investigate the impact of beam-splitting devices in collinear reflective polarization-dependent SHG measurements, highlighting the resulting distortion in the SHG signal due to the distortion of the polarization state in both the fundamental and harmonic beams. To eliminate the distortion, a data correction and optimization method is proposed, and normalized second-order nonlinear susceptibility tensor elements can be obtained for accurate quantitative analysis and validation. Experimental tests on MoS2 and BIBO with BS demonstrate the substantial efficacy of our method. In addition, this paper guides on selecting the right BS assembly, as well as ensuring the robustness and accuracy of polarization-dependent SHG measurements in the absence of a perfect BS. It also provides an accurate polarization control scheme for other nonlinear optical collinear systems.
-
This work was supported by the National Natural Science Foundation of China (No.U2230203) and the Fundamental Research Funds for the Central Universities. Thanks to Professor Ruifen Dou from the Department of Physics and Professor Nan Liu from the School of Chemistry, Beijing Normal University, for the sample of MoS2.
Quantitative Measurements of Second-Order Nonlinear Susceptibility with Collinear Reflective Second Harmonic Generation†
- Received Date: 29/02/2024
- Available Online: 27/04/2024
-
Key words:
- Second harmonic generation /
- Polarization distortion /
- Beam-splitting device /
- Second-order nonlinear susceptibility
Abstract: Polarization-dependent second harmonic generation is a widely utilized technique for characterizing symmetry. However, in collinear reflective geometry, the essential beam-splitting device significantly influences both the polarization state of the fundamental and harmonic beams, thereby affecting the accuracy of the obtained second-order nonlinear susceptibility. Here, we propose a data correction method to solve this problem to obtain accurate second-order nonlinear susceptibility. The feasibility and generality of the method are demonstrated through theoretical and experimental validation.