Achieving single-molecule sensitivity has long been regarded as the ultimate goal of analytical chemistry, with the single-molecule detection technique serving as a critical tool in achieving ultrasensitive sensing of target molecules in minute quantities. The single-molecule technique has shown great potential applications in various domains, especially in biological and medical fields [1−3]. For instance, it enables the observation of dynamic gene regulation and DNA damage [4−7], research into the interactions of signaling molecules and receptors [8], the study of the structure and localization of the constituent molecules in biological complexes [9−11], and the quantification of rare biomarkers in patients’ samples [12, 13]. Recently, the single-molecule immunoassay technique has emerged as a highly promising tool for diagnosing diseases with low abundance blood biomarkers [14−16], particularly for neurodegenerative diseases [17, 18] like Alzheimer’s disease [19, 20]. Moreover, single-molecule detection has been proposed as a method for drug discovery [21−23] and for identifying novel biomarkers for early diagnosis [24−26]. In this context, there is an urgent demand for a new single-molecule detection technique that is cost-effective, highly efficient and offers high throughput.
Among various single-molecule techniques, single-molecule fluorescence stands out as the most valuable tool for biomedical applications, relying on total internal reflection fluorescence (TIRF) microscopy [27]. TIRF microscopy is an elegant optical technique that capitalizes the generation of an evanescent wave through the total internal reflection of an excitation laser within the interface of a transparent cover glass and a liquid. Due to the exponential decay of evanescent waves with distance from the surface of the cover glass, only fluorescent molecules within a few hundred nanometers are effectively excited. This characteristic leads to minimal background interference and high sensitivity for single-molecule fluorescence detection. However, TIRF microscopy face challenges in terms of being expensive and technically sophisticated for many researchers. An alternative method known as a zero-mode waveguide (ZMW) also utilizes the evanescent field for single-molecule fluorescence excitation, offering a potential solution by simplifying the complicated optical alignment [28−30]. Nevertheless, fabricating a ZMW chip necessitates access to a micro/nano processing clean room, which is not readily available in most laboratories.
Silicon photonics, a complementary metal-oxide-semiconductor (CMOS) compatible process, presents a highly scalable and cost-effective avenue for mass production [31]. Recently, it has demonstrated promising potential in sensing chemical and biological molecules [32−35]. Photonic integrated circuits (PICs), derived from silicon photonics, possess the ability to confine, guide, and manipulate light within a chip. One fundamental element of PICs is the higher refractive-index optical waveguides, which exhibit strong evanescent field confinement around their surface [36]. This property can be exploited for fluorescence excitation, resulting in a highly suppressed fluorescent background [36]. Based on this sensing mechanism, PIC chips have recently been utilized for super-resolution nanoscopic imaging of bio-samples [37, 38].
In this study, we developed a silicon photonics-based PIC chip designed for high throughput single-molecule fluorescence detection. Through precisely positioning of nanopores 50 nm above the waveguides, we efficiently excite immobilized single fluorophores within the bottom of nanopores using the evanescent field generated by the waveguide. Through thoughtful design and layout, we successfully implemented a large-area sensing array covering 2 cm×2 cm for extensive single-molecule fluorescence detection. This PIC chip, characterized by its scalability, high-throughput, and cost-effectiveness, has the potential to revolutionize widespread applications in drug discovery, novel biomarker identification, and early diagnosis using single-molecule fluorescence detection.
-
To enable large area and high throughput single-molecule fluorescence detection, we designed photonic integrated circuits within a silicon photonics chip, as illustrated in FIG. 1(a). The excitation laser, originating from free space, was efficiently coupled into the PIC chip via a focus grating coupler. Subsequently, the laser light propagated within a single-mode waveguide. This single waveguide was further divided into 1024 waveguide arrays through 10 levels of multimode interferometers (MMIs). Consequently, the laser light was split into 1024 beams within the PIC chips. Along each waveguide, 1024 nanopores were precisely positioned ~50 nm above the waveguide, with a distribution period of 6.5 μm. The fluorophores immobilized within the bottom of each nanopore could be excited by the evanescent field generated by the single-mode waveguides (FIG. 1(b)). The single-molecule fluorescence could be naturally obtained using the same sensing mechanism as the TIRF microscopy. With a standardized layout in PIC, a sensing array of more than 1 million within a 2 cm×2 cm chip could be readily achieved.
The nanopore-based PIC chips were fabricated using the standard CMOS process. Initially, a silicon oxide layer was deposited onto a silicon substrate, followed by the deposition of silicon nitride onto the oxide layer. Photolithography and dry etching processes were employed to prepare the photonic integrated circuits. Subsequently, another oxide layer was deposited and used as an optical cladding. The array of nanopores was then fabricated within the cladding oxide layer using deep ultraviolet (DUV) lithography. The nanopore structure is depicted in FIG. 1(b), and the transmission electron microscopy (TEM) image was performed using Talos F200i. The diameter and height of the nanopore are approximately 250 nm. The distance between the top surface of the waveguide and the bottom of the nanopore is less than 50 nm. The cross-section dimensions of each waveguide are 330 nm in length and 172 nm in width, as depicted in the TEM image shown in FIG. 1(b).
-
The single-molecule fluorescence was measured using homemade microscopy, as schematically depicted in FIG. 1(c). A continuous-wave single-mode laser operated at 561 nm severed as the excitation source, transmitted into the PIC chips via grating assistant fiber-to-chip coupling [39]. The input optical power was approximately 150 mW. The chip was securely held by a vacuum chuck on a two-dimensional translational stage, while the fiber was affixed to a three-dimensional translational stage to facilitate laser coupling for chips. A homemade optical cage system was constructed for fluorescent signal collection, comprising a 60× water-immersion objective with a numerical aperture (N.A.) of 1.0 (NIR Apo, Nikon), a dichroic mirror (Chroma), a bandpass filter (Chroma), a C-mount 0.7× Relay Lens (Nikon), and a scientific CMOS camera (ORCA®-Fusion BT, Hamamatsu). Additionally, an LED was employed to illuminate the chips for focal plane adjustment. The inset image shown in FIG. 1(c) illustrates the illuminated chip.
The PIC chips underwent treatment with air plasma for 2 min to ensure that the oxide surface was saturated with the hydroxyl group for silanization [40−42]. A customized polydimethylsiloxane (PDMS) frame was then affixed to the PIC chip, and a mixture of isopropanol and deionized (DI) water in a 1:1 ratio was applied at the corner of the frame to ensure complete wetting of the nanopores [43]. Saline-PEG-biotin and saline-PEG (Nanocs, Inc.) were dispersed in ethanol at a concentration of 0.6 mg/mL, respectively. Subsequently, a series of mixed solutions with different volume ratios of saline-PEG-biotin to saline-PEG were prepared at ratios of 1%, 0.1%, 0.01%, 0.001%, and 0.0001%. The PIC chips were then immersed in the mixed solution overnight for surface silanization. Following this, the PIC chips were rinsed with ethanol and DI water to remove any excess mixed solution, and subsequently incubated with 1.6 μmol/L streptavidin for 10 min. After washing with DI water, all chips were incubated with atto565-biotin (Sigma-Aldrich) for 10 min at concentrations of 0.04, 0.07, 0.4, 0.8, 4.4, 8, 44, and 80 nmol/L, respectively. Finally, after three rinses with DI water washes, the atto565-biotin fluorophores were immobilized at the bottom of the nanopores (FIG. 1(b)).
By adjusting the ratio of saline-PEG-biotin to saline-PEG, the surficial binding sites of biotin can be precisely controlled. Following conjugation with streptavidin, the atto565-biotin can be readily immobilized within the nanopores. Theoretically, the fluorophores located within approximately 100 nm at the bottom of the nanopore will be excited, as the evanescent field decays exponentially along the vertical direction of the nanopore. FIG. 1(b) depicts the schematic of fluorophore excitation within a nanopore. All images were captured with a dwell time of 100 ms.
-
Numerical simulations were performed using commercial finite-difference time-domain (FDTD) software (FDTD Solutions, Lumerical Inc. Canada). The simulation setup involved modeling a nanopore with a diameter of 250 nm and a height of 250 nm, while the waveguide was positioned beneath the nanopore with a distance of 50 nm. The region both inside and above the nanopore was filled with water (refractive index 1.33). The waveguide material used was Si4N3 with a refractive index of 2.0, while the surrounding material was set as SiO2 with a refractive index of 1.46 [44, 45]. The source in TE mode propagated in the waveguide with a wavelength of 561 nm. A 3D non-uniform meshing with a grid size of 1 nm was employed in the total field domain, with perfectly matched layer (PML) absorption boundary conditions. To evaluate the electric field distribution map at 561 nm, a 2D frequency domain field profile monitor placed in the midsection of the nanopore was used.
-
FIG. 2(a) shows the fluorescent image of a nanopore-based PIC chip, demonstrating the selective excitation of fluorophores located within nanopores. This phenomenon arises from the exponential decay of the evanescent field originating from the waveguide surface, with an effective excitation depth of approximately 100 nm. In single-molecule experiments, real-time image sequences undergo analysis by fitting the fluorescence intensity distribution using a superposition of Gaussian functions. Subsequently, the occurrences of photobleaching steps are quantified, facilitating the characterization of single molecular dynamics over time [46]. As illustrated in FIG. 2(b), the presence of a single fluorophore within the nanopore, results in a distinct photobleaching or a blinking step, indicating the discrete behavior of a single molecule. Conversely, when multiple fluorophores are localized within the nanopore, a stepwise photobleaching curve of individual fluorescent spots emerges, as depicted in FIG. 2(d). The fluorescence intensity distribution analyses presented in FIG. 2 (c) and (e) correspond to the traces in FIG. 2 (b) and (d), respectively. These distributions clearly delineate the presence of either a single fluorophore or multiple fluorophores within the nanopore. Thus, the nanopore structure enables the precise extraction the single-molecule events, facilitating high throughput and ultra-sensitive detection capability.
To investigate the optimal conditions for single-molecule fluorescence detection using the silicon photonics platform, we conducted an investigation involving the variation of the surficial biotin ratio for fluorophore immobilization. The biotin ratio was adjusted by mixing different proportions of saline-PEG-biotin and saline-PEG solutions, specifically testing ratios of 1%, 0.1%, 0.01%, 0.001%, and 0.0001%. Subsequent to incubation with streptavidin, the atto565-biotin fluorophores were sparsely anchored in nanopores. Quantifying the single-molecule events of fluorophores blinking, we determined that the optimized biotin ratio was approximately 0.01% (refer to FIG. 3(a)). Deviations from this ratio, either higher or lower, led to a decreased likelihood of single-molecule events in nanopores. Furthermore, we investigated how the concentration of atto565-biotin influenced single-molecule fluorescent events in nanopores while maintaining the biotin ratio at 0.01%. In this experiment, the modified biotin ratio of 0.01% was consistently applied. As the concentration of atto565-biotin increased, there was a corresponding rise in the occurrence of a single-molecule fluorescent event, peaking at a concentration of 8 nmol/L (see FIG. 3(b)). However, further increases in the concentration of atto565-biotin led to a reduction in single-molecule events due to the increased possibility of multiple fluorophores being present in nanopores. Additionally, we analyzed the probability of single-molecule events in the nanopore-based PIC chip, finding it to be approximately 57.9%, significantly surpassing the expected value of 40% derived from a Poisson distribution [47]. This elevated probability suggests the potential anchoring of fluorophores in the side wall of nanopores, thereby potentially increasing the likelihood of single-molecule events.
To verify the spatial distribution of the electric field within nanopores, we conducted an electromagnetic simulation using the FDTD method. FIG. 4(a) illustrates the spatial distribution of the electric field at 561 nm. The laser propagated through the waveguide is highly confined, and the resulting leaked electric field generates an evanescent field at the bottom of the nanopores, facilitating the excitation of the fluorophores. Along the vertical axis above the waveguide, the electric field intensities at the central axis of the nanopore exhibit exponential decay, as depicted in FIG. 4(b), signifying effective excitation of fluorophores within approximately 100 nm. Furthermore, the simulation results indicate that the electric field intensities within the nanopore are not influenced by its size. FIG. 4 (b) and (c) demonstrate minimal changes in electric field intensities with varying diameter and depth of the nanopore, respectively. This observation suggests that the electric field distribution within a nanopore is solely determined by the waveguide. Previous studies have indicated that the distance between the bottom of the nanopore and the fluorophore is approximately 3 nm [48]. Therefore, we simulated and extracted the electric field intensity at 3 nm above the bottom of the nanopore, as depicted in FIG. S1 (a) and (b) (Supplementary materials, SM). The results reveal that the maximum excitation field is situated at the nanopore’s center. Consequently, the fluorophore located in this region exhibits higher fluorescent intensity owning to the increased excitation power.
-
In this study, we have introduced a novel single-molecule fluorescence detection technique implemented on an integrated silicon photonics chip. Our investigation focused on exploring the optimal condition required to achieve a high probability of single-molecule events, and determining a biotin ratio of approximately 0.01% and a concentration of 8 nmol/L as the most effective parameters. Through electromagnetic simulations, we successfully illustrated the confinement of the evanescent field, which enables fluorescent excitation with minimal background noise, akin to TIRF excitation. This on-chip single-molecule fluorescence technique not only offers a cost-effective approach but also facilitates high-throughput detection, making it a promising strategy for ultra-sensitive diagnostics. The potential applications of this technique extend to high-throughput single-molecule diagnosis of diseases, highlighting its significant impact and broad utility in biomedical research and clinical settings.
Supplementary materials: The electric field intensity at 3 nm above the bottom of the nanopore with varied depth and radii of the nanopore are shown in FIG. S1.
Single-Molecule Detection by Silicon-based Photonic Integrated Circuits Chip
- Received Date: 12/04/2024
- Available Online: 27/04/2025
-
Key words:
- Single-molecule fluorescence /
- Silicon photonics /
- Photonic integrated circuit /
- Total internal reflection fluorescence microscopy
Abstract: The single-molecule detection technique plays a pivotal role in elucidating the fundamental mechanisms of various scientific processes at the molecular level, and holds essential importance in multiple fields including physics, biology, and chemistry. Recently, single-molecule detection has garnered increasing attention owing to its practical utility in medical diagnosis, primarily due to its exceptional sensitivity and the minimal sample volume required for analysis. However, the conventional single-molecule technique, represented by total internal reflection microscopy, faces challenges such as sophisticated operation procedures and limited detection throughput, thereby impeding its broader application. To address these limitations, we have demonstrated single-molecule detection using an integrated silicon photonic chip, offering a cost-effective and user-friendly alternative. By employing basic optics, we efficiently introduce the excitation source for single-molecule fluorescence by harnessing the strong evanescent field of high refractive-index waveguides. Subsequently, fluorescence signals are collected using basic optics comprising a water-immersion objective, relay optics, and a digital camera. Our results highlight a low-cost, high-throughput single-molecule technique achieved through the integrated silicon photonic chip. This innovative approach is promised to facilitate the widespread adoption of single-molecule fluorescence in medical diagnosis.