Scanning tunneling microscopy (STM) is a powerful tool for characterizing surface electronic states and obtaining precise atomic structures at localized areas. It has significantly advanced the study of correlated and topological materials, as well as superconductors [1−8]. This technique has led to cutting-edge researches in condensed matter physics. With the powerful ability to detect various modulations of the electronic state induced by thermal and magnetic variations, it can observe temperature-triggered commensurate charge density waves in transition metal dichalcogenides (TMDs) [9, 10], field-modulated order parameters in superconductors [5], and magnetic field-provoked Landau quantization in topological materials [7, 11], promoting a thorough understanding of surface properties. Traditional STM systems based on superconducting magnets typically operate at temperatures around 1 K or 4.2 K [12−14]. This is due to the limitation of cooling systems driven by helium-4, which forms superfluid below 2.17 K, complicating the cooling process. This limitation hinders the investigation of many exotic quantum phenomena, especially in few-layer TMD superconductors and correlated-twisted materials [4−6], which require sub-Kelvin temperatures to investigate their intrinsic properties without too much thermal effects. Many STM systems equipped with helium-3 cryostats can reach sub-Kelvin temperatures and some of them can achieve a base temperature of around 0.3 K with the long operation period around 80 h [15], while others can only be above 0.3 K with the operation time around 30 h at this temperature [16−18]. Our He-3 cryostat, which theoretically operates for over 100 h, can maintain the temperature at 0.3 K for 72-h experiments (including 30 h of 1 T mapping and over 10 h of high-field looping up to 12 T) and has the base temperature of 0.28 K without experiment, promising to be an advanced platform for the long-term experiment with stable magnetic field.
Achieving high-field stability in STM imaging is challenging due to electronic noise from high magnetic fields and mechanical vibrations, which can limit accurate probing. While many systems are equipped with high-field magnets, little systems have demonstrated exceptional performance at sub-Kelvin temperatures up to 12 T. To address these challenges, we developed a system based on a superior sapphire scanning head, which has a high eigenfrequency and reduces electronic and mechanical noise from the motor. With the outstanding external passive vibration-isolation system and the robust scanning head, our system’s mechanical noise (presented as current spectral density, CSD) is as low as 2×10−13 A/
$\sqrt{\rm{Hz}}$ at low frequency range ( < 45 Hz) when the input is at a cut-off frequency of 250 Hz at zero field. When scanning at 0.3 K and 12 T, the drifting rate of the images of graphite can be as low as vx=−8.79×10−7 nm/s and vy=−3.26×10−6 nm/s. Additionally, the images of graphite up to 12 T are also clear without the distortion and with little noises at the surface. As for the time-dependent performance, our high-capacity He-3 cryostat (with 15 normal temperature and pressure liter (NTPL) He-3 gas) can operate with the field around 40 h and has a total operation time more than 72 h at 0.3 K. To prove the performance from the experiment, we have captured high-quality dI/dV mapping, which requires over 12 h of continuous data collection per figure, of the bulk 2H-NbSe2 at 1 T and 0.3 K. Ultimately, with the excellent thermal conductance design between the scanning head and the He-3 pot, the base temperature of the system can be maintained at 0.28 K without the experiment. Overall, our system’s performance is outstanding, capable of probing surface electronic states under harsh conditions and low temperature at long period of time. -
The system assembly can be divided into three isolated parts as shown in FIG. 1(a): (i) the ultra-high vacuum (UHV) preparation chamber; (ii) the cryostat and the scanning head; (iii) the vibration-isolation stage. The UHV preparation chamber consists of a 304-stainless-steel chamber (processed by the CHI-VAC Research & Development Co., Ltd.), an iron pump, a load-lock chamber (processed by the CHI-VAC Research & Development Co., Ltd.), a 2-meter 304-stainless-steel manipulator (processed by the CHI-VAC Research & Development Co., Ltd.), and two in-plane 304-stainless-steel manipulators (purchased from the Fermi Instruments (Shanghai) Co., Ltd.). Several observation windows around the chamber allow easy positioning of the sample stage and the manipulator’s connecting head. The chamber is supported by a four-leg adjustable aluminum frame (processed by the CHI-VAC Research & Development Co., Ltd.), which allows precise height adjustments to align the cryostat and the chamber with the gate valve. The ion pump (purchased from the Gamma Vacuum Co., Ltd.), connected to the chamber at the end flange and supported by the aluminum frame, provides UHV to maintain the cleanliness of sample. Connected orthogonally to the chamber, the manipulators, facilitate accurate sample transfer from the load-lock chamber to the bottom of the cryostat, which directly fixes the scanning head. The cryostat (purchased from Janis Research Company Co., Ltd.), connected to the chamber, is inserted into the 12 T magnet (purchased from Cryogenic Co., Ltd.) (with a 78 mm bore), which is supported by the vibration-isolation stage at the top edge of the Dewar. The He-3 pot cools down the system by pumping out He-4 from the magnet. The magnet has a low liquid helium consumption (maximum volume 75 liter (L), 4 L per day without operation, and an average of 9 L per day during the experiments), operating in persistent mode when the magnetic field is applied.
Overall, the assembly is clearly illustrated in FIG. 1. In the following section, we will introduce the most critical part: the cryostat and the scanning head.
-
The cryostat model, with a base temperature of 0.28 K, is provided by the Janis Research Company (model name: He-3-TLSUHV-STM INSERT, United States). Significant parts marked in FIG. 1(b) include He-3 tank, sample transfer port, multi-pins connector, SMA connector, baffle, charcoal sorption pump, 1 K stage, shutter, He-3 pot, single crystal copper thermal conductor rods, and self-built scanning head (ceramics processed by Fangcunzhizao Technology (Dongguan) Co., Ltd.). The cryostat operates on a loop-locked circuit principle, where heating or cooling the charcoal sorption pump releases or sorbs He-3 gases, maintaining the sub-Kelvin temperature for the He-3 pot. The scanning head is directly connected to the He-3 pot via three single crystal copper rods, ensuring the entire scanning head is chilled. With an improved He-3 tank volume of 15 NTPL, this cryostat can operate at the base temperature for over 100 h without loading, and has been tested to work over 72 h (with a magnetic field over 40 h), demonstrating its durability for low-temperature experiments.
The multi-pins connector features 6 coaxial lines and 10 twisted-pairing lines, directly inputting signals to the scanning head and outputting tunneling current to the preamplifier, minimizing signal interference. The assembly of the scanning head, as depicted in FIG. 2(b), includes several components such as the guiding part, sapphire plate, copper sample stage and holder, sapphire body, scanner Piezotube (purchased from EBL Products Co., Ltd. EBL#3 Piezotube), sapphire tip holder, polished zirconia slide rail with a polished arched BeCu spring, polished sapphire rod with hooked bottom, polished zirconia shaft with copper hook at the top, motor Piezotube (EBL Products Co., Ltd. EBL#3 Piezotube), sapphire motor body, capacity displacement sensor, and copper heat sink. The outer and the inner diameter of the scanner piezotube are 3 mm and 2.15 mm while the outer and the inner diameter of the scanner piezotube are 10 mm and 9 mm. To ensure the stability at the high magnetic field and the effective power of the motor, the length of the scanner is 3 mm and the motor is 40 mm.
Assembly of the scanning head involves tightening six titanic screws to combine components in a sandwich-like manner. Thermal epoxy (purchased from Epoxy Technology Co., Ltd.) bonds the zirconia slide rail directly to the head of the sapphire body. The sapphire rod, attached to the zirconia slide rail with a polished arch BeCu spring, connects to the bottom of the scanner via thermal epoxy. The head of scanner connects to the sapphire tip holder with thermal epoxy, while the hook of the sapphire rod quasi-connects with the zirconia shaft’s hook. A polished zirconia tube, connected to the motor with thermal epoxy, holds the zirconia shaft with a polished arch BeCu spring. The motor is assembled to the sapphire body’s end with a sapphire ring, bonded with thermal epoxy. The capacity displacement sensor, measuring displacement of the zirconia shaft, connects to the sapphire frame’s end with thermal epoxy. The heat sink’s end connects to the upper part with three tantalic screws.
The sapphire motor body, which is very limited and compact to endure the vibration, is only 15 mm from the measurement of the diameter (as shown in FIG. 2(c)). Besides, the sapphire top body can ensure the robustness of the scanner against the external mechanical vibration while the separation of the shaft and rod can resist the perturbation, which will disturb the quality of the images at the high magnetic field and low temperature, caused by the piezo motor. These conclusions have been proved by the experiment at the water cool magnet and hybrid magnet (the field can be up to 30.1 T [19−21]). The resistance of the external mechanical vibration is estimated by the calculation of the eigen frequency, which reveals the high frequencies of 34 kHz, 60 kHz, 91 kHz, and 100 kHz for front-back, left-right, twisted, and axial directions respectively. Sample transportation relies on three orthogonal manipulators and the perpendicular manipulator can directly connect the scanning head. Thermal connection involves connecting the guiding part’s head to three single crystal copper rods, connected to the heat sink with screws, ensuring effective cooling for the scanning head. The cooling process involves: the heater inside the charcoal absorption pump will come up to 43 K to release the He-3 gas from the pump and the He-3 gas will go to the tube chilled by the 1-K stage. Being liquified at around 1.8 K, the He-3, which will be stored at the He-3 pot, will be continually liquified for 2 h until the He-3 pot is full. During this process, the 1-K stage, which is connected to the He-3 pot with a metal connector, will keep the He-3 pot to be around 1 K. After fulfilling the He-3 pot, the heating of the charcoal absorption pump will be stopped and the cooling coil will cool down the charcoal absorption pump by switching on the small needle valve at the top and keep the charcoal pump to be 4.2 K to turn on the He-3 mode, which will keep the He-3 pot to be 0.28 K by vaporizing process. Besides, the connection between the 1-K stage and the He-3 pot will be cut off to reduce the heat consumption (process shown in FIG. 2(a)). After construction, the system operates in He-3 mode, keeping a minimum temperature of 0.28 K without experiment and maintaining around 0.3 K for over 72 h with operations (with at least 40 h for looped field up to 12 T and 1 T mapping experiments).
-
The vibration-isolation system consists of an aluminum triangular platform positioned on three passive isolators (purchased from Minus K Technology Co., Ltd.). These isolators, with a theoretical natural vertical vibration frequency of at least 0.5 Hz, effectively reduce external mechanical perturbations. We examined the frequency with a 250 Hz cut-off frequency from the controller and collected data from 0 to 1000 Hz. Our results show that aside from the obvious 50 Hz signal and its harmonics, which have CSD up to 1.6×10−12
$\rm{A/\sqrt{Hz}}$ , other frequencies have low CSD of less than 4×10−13$\rm{A/\sqrt{Hz}}$ when the tip is engaged in the tunneling region (FIG. 3(a)). Focusing on the low-frequency region ( < 45 Hz), we find that the dominant low-frequency CSD are below 1×10−13$\rm{A/\sqrt{Hz}}$ (FIG. 3(b)) and the maximum CSD is around 2×10−13$\rm{A/\sqrt{Hz}}$ . This performance at the low frequency is exceptional compared to other STM systems globally [18, 22, 23].The good performance of the vibration isolation system can also be expressed by the good images of the graphite and the low drifting rate at 12 T and 0.3 K. The scanning at 12 T continues a relatively long time at 3.3 nm×3.3 nm for 1536.0 s and the calculated drifting rate can be vx=−8.79×10−7 nm/s and vy=−3.26×10−6 nm/s. The uneven drifting rate for different axis might stem from the different scanning speed. Although the electronic noises mentioned above still exist at the scanning, they will only induce the hidden periodic stripes at the images but will not affect the electronic states from the surface and the shape of atoms. The influence of the noises to the dI/dV spectra is also not very obvious because the gap of the NbSe2 is the normal size (around 1.25 meV) and the variation of the size of the gap versus the magnetic field is also plausible. The influence of the dI/dV spectra might be the unexpected jump in few lines but the problem can be solved by applying more measurements until the results become repeatable. To further improve the noise spectra of the tunneling junction, we can equip more suitable filter for the output of tip, hang the scanning head with some springs, utilize the shorter scanning piezo tube, operate the system at the low-noise time period and outfit heavier parts on the triangular platform.
Despite our scanning head being robustly connected to the He-3 cryostat, the system’s performance remains excellent without an inner vibration-isolation system, demonstrating the effectiveness of the passive isolators and the robust scanning head structure.
-
We examine the STM system’s performance at low temperatures and high magnetic fields. The system uses an RHK R9 SPM controller with an RHK Preamplifier IVP-200, which provides a 5 kHz bandwidth and gain of 108 (100 mV/nA). Graphite is exfoliated and fixed on the sample stage using EPO-TEK H20 E silver conductive epoxy (purchased from the Epoxy Technology Co., Ltd.) at 88 °C for 2.5 h. Fresh surfaces are prepared using Scotch tape (purchased from the 3M Co., Ltd.) and immediately placed in the load-lock chamber for high vacuum pumping. Then the sample is sent to the body with the operations of three manipulators with the gate valves of the sample chamber opening.
Using constant current mode, we image graphite at 0, 5, 10, and 12 T at 0.3 K, with a bias of 200 mV and tunneling current of 2 nA. The images (especially the long-term images at 12 T) remain nearly unchanged and ordered up to 12 T, demonstrating our system’s stability (as shown in FIG. 4(a)–(d) and FIG. 5(a)–(o)). The uneven strips and the irregular atoms on the images are caused by instability of the tip, the electronic noises and the influence of the magnetic field. The shape of the atoms and the surface are good after the long-term scanning at the stable electronic environment shown in the FIG. 5(a)–(o). The dI/dV spectra, which is applied by the modulation frequency of 0.731 kHz and bias of 5 mV, of graphite at 0.3 K and 3 T revealed Landau Levels (LLs), with energy levels proportional to the integrals of the level, fitting well with previous data (FIG. 6(a, b)) [24, 25].
Next, we examine the superconductivity and the surface topography of the bulk 2H-NbSe2. After transferring the sample to the scanning head, we obtain the atomic topography, identifying defects and wrinkles, which indicate the excellent large-area scanning performance of our system (as shown in FIG. 7(a)). When zooming in a flat area, we characterize the surface topography and apply the dI/dV spectra measurement at 0.3 K and zero field, obtaining stable, repeatable spectra consistent with BCS theory (with Δ=1.25 meV), using the modulation frequency of 0.731 kHz and bias of 0.5 mV (FIG. 7(b, c)). The BCS model is constructed as follows [16]:
where the
$\beta =1/{k}_{\rm{B}}T$ , Δ is the order parameter of superconductor, and V is the bias from the sample. Then the effect of the magnetic field to the dI/dV spectra at 0.3 K is introduced in FIG. 7(d). The magnetic field is applied from 0 T to 3 T and the gap of the superconductivity gradually decreases when the field increases. This result can qualitatively show that the measurement of the spectra under the magnetic field is also plausible enough by this system.For long-term stability, we conduct dI/dV mapping from −4 mV to 4 mV at 1 T, observing a vortex structure with a 28 nm diameter at −0.8 mV (FIG. 8(a, b)) [18, 26]. The mapping takes nearly 12 h to ensure the accuracy of the dI/dV spectra in the scanning area. All the results above confirm the system’s excellent performance at sub-Kelvin temperatures, over long periods, and under high magnetic fields.
-
We have developed a high-stable STM system equipped with a He-3 cryostat and a 12 T magnet, achieving magnetic fields up to 12 T and temperatures as low as 0.3 K. The UHV chamber maintains sample cleanliness, and precise sample transportation is ensured by an orthogonal manipulator with a guiding part atop the scanning head. Our system can operate in a closed-loop at 0.3 K for over 72 h and theoretically over 100 h without loading. The vibration-isolation system, featuring a triangular aluminum stage with isolators at its vertices, minimizes mechanical noise with a theoretical natural vertical vibration frequency of 0.5 Hz and from the noise spectra the amplitude is less than 2×10−13
$\rm{ A/\sqrt{Hz}}$ at low frequency ( < 45 Hz). Besides, the low drifting rate of the scanning is estimated from the images to be vx=−8.79×10−7 nm/s and vy=−3.26×10−6 nm/s at 12 T. To enhance performance in high magnetic fields, we have implemented a hook structure in our scanning head to separate the scanner and motor and mitigate noise generation, enabling excellent surface characterization of graphite up to 12 T. For the stability of large-area scanning, we have employed the system for large-area scanning of the bulk 2H-NbSe2, accurately characterizing defects and wrinkles on its surface. Spectra functionality is ensured by the RHK R9 controller and its high-gain 108 (100 mV/nA) and wide bandwidth (5 kHz) preamplifier, facilitating precise spectra acquisition, including the LLs of graphite at 3 T, the order parameter which can be well fitted with the simulation for the bulk 2H-NbSe2 and the magnetic effect to the gap of the superconductor. Ultimately, the long-term scanning is proved by the dI/dV mapping of the bulk 2H-NbSe2, which has a vortex structure of 28 nm at −0.8 mV and 1 T and the images of the graphite about 1536 s. In summary, our system is stable and powerful from the space and time dimension at sub-Kelvin temperature and high magnetic field, capable of investigating various quantum materials, including topological superconductors and correlated materials in the future [27−30].
Robust, High-Magnetic-Field-Resistance Design of Scanning Head Equipped with 12 T Superconductor Magnet and sub-Kelvin He-3 Cryostat
- Received Date: 23/05/2024
- Available Online: 27/12/2025
-
Key words:
- sub-Kelvin /
- High magnetic field /
- Low noise /
- High-stability /
- Scanning tunneling microscopy
Abstract: We have constructed the robust and high-field-stability scanning tunneling microscopy (STM) system which is equipped with a 12 T-superconductor magnet and the He-3 cryostat, obtaining the high-quality data of graphite which can be clearly characterized up to the 12-T high magnetic field and bulk 2H-NbSe2 whose vortex structure and surface defects can be clearly visualized by the scanning and dI/dV mapping. The pivotal factor of our superior data stems from the sapphire frame and its unique structure, which ensure an exceptional performance for the noise, which has the current spectral density low to 2×10−13$ \mathrm{A}/\sqrt{{\mathrm{Hz}}} $ at the low frequency with the cut-off frequency of 250 Hz and the engaging status, of vibration, promising to the robust platform for the investigation of the surface of materials. The low drifting rate from the image of graphite is ${v}_{x}= -8.79\times {10}^{-7}\;{\rm{nm/s}}$ and ${v}_{y}= -3.26\times {10}^{-6}\;{\rm{nm/s}}$ at 12 T and 0.3 K. Besides, equipped with the high-capacity He-3 cryostat (with 15 normal temperature and pressure liter He-3 gas), the system, which has the base temperature for 0.28 K, can operate at the low temperature for more than 72 h, ensuring the enduring research at the sub-Kelvin temperature for the dI/dV mapping and large-scale scanning. When the magnetic field is applied up to 12 T, the STM system also obtains the excellent-clarity and undistorted images of graphite, which proves the outstanding performance of the system at the high-field researches.