The quadrupole ion trap mass spectrometer is the most classical ion trap mass spectrometers, which generally obtain the mass spectra by ramping the radio frequency (RF) voltage attached to the ring electrode [1−5]. This method of ion ejection is called the mass instability scan or boundary ejection, by which ions are ejected due to qz reaching 0.908 [6]. The mass instability scan yields a linear relationship between the RF voltage and m/z of the ion ejected [7−9]. However, the mass instability scan requires a higher RF voltage output for the larger mass range detection [10]. An alternative method is the resonance excitation scan method, which effectively extends the mass range of the ion ejection by applying a supplementary alternating current (a.c.) voltage to the end caps, leading to the resonant ejection of ions at the qz values below those corresponding to the normal stability limit [8, 10−12]. In 2004, Ding et al. constructed a digital ion trap mass spectrometer driven with the square waveform, which implemented the frequency scan while scanning the frequency of the a.c. excitation voltage, achieving mass-to-charge ratios up to 17000 Th and mass resolution up to 3500 [13]. Recently, the frequency scans of RF and a.c. excitation voltage were used for the ion trap mass spectrometers due to their larger available mass range and compact instrumental condition [14].
The secular frequency scan generates the ion ejection by only scanning the frequency of the a.c. excitation voltage applied to the end caps [15]. Snyder et al. showed that the inverse Mathieu qz scan offered a significant increase in mass range while maintaining access to linearity, paving the way for a mass spectrometer entirely based on a.c. waveforms for ion isolation, ion activation, and ion ejection [7]. The application of the secular frequency scan to the linear ion trap mass spectrometer was also verified [8, 9, 16−18]. Additionally, the secular frequency scan was demonstrated as a method of mass-selective excitation to enable precursor scans in a single ion trap in a miniature mass spectrometer [19]. In 2020, Lee et al. compared the five different operational modes, including the forward and reverse secular frequency scans for the quadrupole ion mass spectrometer [20]. The secular frequency was able to yield the ion ejection up to m/z = 4 × 105 Da, though the resulted mass resolution was worse than that from the ion ejection mode with the synchronous scans of the frequencies of the ring-electrode and end-cap voltages in the square waveform. In this paper, we report the observation of the secular frequency oscillation of the rectilinear ion trap mass spectrometer with ions ejected by the secular frequency scanning. With the benzene vapor as the sample, the full width at half maximum (FWHM) of the mass peak at mass 78 is 0.012 kHz, which corresponds to 8600 m/Δm in the mass resolution.
-
FIG. 1 shows the schematic diagram of the newly built VUV photoionization ion trap mass spectrometer. It is mainly composed of a two-chamber vacuum system, a VUV lamp, a pulsed sampling device, a VUV photoionizer, an ion trap electrode assembly, a pulsed helium device, an ion trap controller unit, and an ion detection and intensity record unit as shown in FIG. 1. The two-chamber vacuum system consists of two cubic aluminum alloy chambers, two turbo molecular pumps (350 L/s Leybold 350i), one foreline pump (11 L/s Ecodry 40 plus), and a couple of vacuum gauges. The two vacuum chambers are each pumped by a turbo molecular pump. The two turbo molecular pumps share a foreline pump. The two vacuum chambers are connected by a differential plate featuring a centrally positioned aperture with a diameter of 2 mm. In the ion injection period, the vacua of the ion source chamber and the detection chamber were ~1 and 1 × 10−2 Pa, respectively. The pulsed sampling device is made up of a four-feet long Φ=1/8 stainless tube, a needle valve (NV 1), two solenoid valves (SV 1 and SV 2), and two pulsed DC-controlled 24 V DC power supplies. One solenoid valve (SV 1) controls the time of sampling, while another (SV 2) controls the pumping out of the residual gas in the sample line. The minimum operation time of the solenoid valve is about 0.3 s. The laboratory-built VUV lamp is a krypton lamp outputting ~5 × 1014 photon/s at 123.9 nm [21]. A plano-convex MgF2 lens (f = 50 mm) is used to focus the VUV light into the photoionizer. The structures of the VUV lamp and the photoionizer have been reported in detail by Guo et al. [21]. The photo-ionizer region is a tapered hole to match the focusing region of the MgF2 lens. The ion trap electrode assembly is composed of two x-electrodes with a 1 mm (H) × 10 mm (L) slit in the center, two y-electrodes, and two axial z-electrodes. The ions are injected into the ion trap through the slit of one x-electrode and ejected through the slit of another opposite x-electrode. The ion trap electrodes above are insulated with the PEEK spacers. The central part of the axial z-electrode has a small hole for filling the cooling gas. Half the distance of the x-electrodes (r0) is equal to 1.125 times the half distance of the y-electrodes (z0). The ion trap with r0 = 9 mm was used in this study. The pulsed helium device has a structure similar to a pulsed sampling device. One solenoid valve (SV 3) controlled by the pulsed DC signal opens for a certain time to let helium flow into the ion trap and then shuts. Another solenoid valve (SV 4) subsequently opens to pump out the residual helium in the Φ=1/8 Teflon tube connected to the ion trap.
The ion trap controller unit consists of a computer, a multifunction I/O device (NI 6122), a radio-frequency generator (Stahl HF-DR), a multichannel HV power supply (ISEG), an arbitrary waveform generator (Tektronics AFG 31022 upgraded with the memory of 128 M), and a DC pulser. The DC pulser based on the solid-state relay outputs a positive voltage pulse to control the ion injection. The ion detection and intensity record unit consist of an electron multiplier (Detech 397), a current amplifier (HB-877 (C)), and a digital oscilloscope (RTA 4004).
A Labview-based software is programmed to run the ion trap mass spectrometer. The software is divided into three modules, including the ion trap operation, the data collection, and the mass calibration. The ion trap operation module executes the setting of the ion trap’s parameters and controls the RF power supply for on/off/ramping of the RF voltage. Additionally, the ion trap operation module also runs the multifunction I/O device to generate a string of timing DC pulses to activate the pulsed sampling device, the pulsed helium device, the solid-state relay-based DC pulser, and the arbitrary waveform generator. The data collection module is responsible for initializing the oscilloscope, acquiring the data from the oscilloscope, and saving the data into the hard disk of the PC. The mass calibration module retrieves the data from the hard disk, calibrating the mass values with the parameters of the ion trap and the used RF voltage, compressing the data points of mass spectra, and saving the resulting mass spectra.
FIG. 2 shows the operational mode of the ion trap mass spectrometer. The RF voltage is maintained at a preset voltage during the stages of ion injecting, cooling, RG pumping, and a.c. frequency scanning. SV 1 and SV 3 open about 2 s (T1) earlier than the DC pulser for the tube flushing. The DC pulser usually lasts 1 s (T2) and applies a 40 V DC to the photoionizer made of stainless steel to inject ions into the ion trap. The cooling time (T3) is about 200 ms. SV 2 and SV 4 open 300 ms (T4) before the AC scanning and pump out the residual sample gas in the sample tube and the residual helium in the cooling gas tube. The secular frequency scan usually lasts 500 ms (T5).
A 60-L sample chamber (not shown in FIG. 1) is used to prepare the sample gas with the desired concentration. The sample chamber is an open, thin-walled stainless steel cylinder covered with a Teflon bag to maintain a constant pressure during sampling. The sample chamber is heated at 60–80 °C with a silicon rubber heater attached to the outer wall of the stainless steel cylinder. A magnetic-driven fan is mounted at the bottom of the chamber to mix the gas during the sample preparation.
-
FIG. 3(A) shows the signal intensity of benzene ion responding to the scan of the secular frequency. The RF voltage is 431 V. The concentration of benzene sampled is 6.78 ppmv. A reverse linear secular frequency scan is conducted from 210 kHz to 30 kHz at 2.01 V in 0.5 s. The resonance signal peaks of benzene and its 13C benzene appear at 99.8 and 98.6 kHz. FIG. 3(B) shows the zoomed signal peak of benzene ions, which shows the clear oscillation of the benzene ions corresponding to the scanning of the secular frequency. The FWHM of the responding resonance frequency of benzene is 0.012 kHz. The ions confined in the ion trap RF electric field have a specific oscillation frequency ω, also known as the secular frequency. The secular frequency of ions is related to the parameters of the ion trap. Its expression of the secular frequency for rectilinear ion trap is shown in the following equations when the DC voltage applied to the y-electrodes (U) is zero.
where V and Ω are the voltage and frequency of the radio frequency applied to the y-electrodes. r0 and z0 are, respectively, half the distance between the pairs of parallel electrodes for the x-electrodes and y-electrodes. The above equations can be obtained from the Mathieu equation [22]. Goeringer et al. proposed the resonance ejection technique, which greatly improved the efficiency of the ion ejection [11]. The resonance ejection technique scanned the secular frequency while increasing the y-electrodes voltage. The ion ejection technique using the only secular frequency scan was usually used in small ion trap mass spectrometers due to the simple structure and energy savings. However, the experimental phenomenon of the fine secular frequency resonance oscillation has never been reported.
FIG. 4(A) shows the mass spectra of benzene obtained with the relation between m/z and the secular frequency shown in Eqs. (1)–(3) and small parameter modifications [23, 24]. FIG. 4(B) shows the zoomed mass spectra of benzene. The FWHM of the benzene mass peak was 0.009 m/Δm, which is equal to 8600 in the mass resolution at m/z 78. The mass resolution of the ordinary ion trap mass spectrometer derived from the quadrupole mass spectrometer ranges from hundreds to thousands of m/Δm. The mass resolution of the secular frequency scan ion trap mass spectrometer developed by Snyder et al. was ~700 [16]. The digital ion trap mass spectrometer developed by Ding et al. achieved a 0.015 Da FWHM of the mass peak by implementing the frequency scan while scanning the frequency of the a.c. excitation voltage [13]. The better mass resolution obtained in this experiment may contribute to the slow scan rate and a proper vacuum condition. It is worth noting that the position of signal peak tailing in FIG. 4 is contrary to FIG. 3. The conversion of the x-axis unit from the secular frequency Ω to the ratio of mass to charge m/z changes the direction of the signal peak tailing due to the reciprocal relationship between Ω and m/z.
FIG. 5 shows the sensitivity curves of benzene in the concentration ranges of 200–800, 1000–4000, and 4000–10000 ppbv. In this experiment, adopting the same experimental conditions would lead to signal saturation due to the limitation of the electron multiplier in our instrument. Therefore, the sensitivity curves are divided into three concentration ranges. The sensitivity curves present good linear relationships in each concentration range with correlation coefficients (R2) ranging from 0.98 to 0.99. The noise level of the mass spectrometer is 0.23 mV. The sensitivity of benzene is 16.5 mV/ppmv. The 3σ LOD of benzene is 42 ppbv. Therefore, these results indicate that the detection method has high accuracy and reliability across different concentration ranges, making it suitable for environmental monitoring and quantitative analysis with further improvements.
-
In this work, we reports an experimental observation of the fine secular frequency resonance oscillation with the rectilinear ion trap mass spectrometer. The FWHM of the resonance frequency of benzene was 0.012 kHz. The experimental phenomenon of the fine secular frequency resonance oscillation of the trapped ions has not been documented yet. The corresponding mass resolution was 8600 m/Δm for m/z=78. The mass spectrometer has a linear response to benzene in the concentration range from 200 ppbv to 10000 ppbv. This study indicates secular frequency scanning has the potential to generate higher mass-resolution mass spectra for ion trap mass spectrometers.
Experimental Observation of Fine Secular Frequency Resonance Oscillation with a Rectilinear Ion Trap Mass Spectrometer†
- Received Date: 22/08/2024
- Available Online: 27/12/2024
-
Key words:
- Fine secular frequency /
- Resonance oscillation /
- Mass resolution /
- Rectilinear ion trap mass spectrometer
Abstract: The mass resolution of the ordinary ion trap mass spectrometer derived from the quadrupole mass spectrometer usually ranges from hundreds to thousands. In this study, the fine secular frequency resonance oscillation was observed with a rectilinear ion trap mass spectrometer. The FWHM of the responding resonance secular frequency of benzene was 0.012 kHz. The corresponding mass resolution was 8600 m/Δm for benzene (m/z=78). The mass spectrometer with secular frequency scanning has a linear response to benzene in the concentration range from 200 ppbv to 10000 ppbv. This study shows a new feature of the secular frequency scanning for ion trap mass spectrometers.