An Zhao, Qing-Yu Liu, Zhong-Pu Zhao, Zi-Yu Li, Sheng-Gui He. Coupling of a High-Temperature Ion Trap Reactor with an Electron-Spray-Ionization Source for Reactions of Mass-Selected Organometallic Ions†[J]. Chinese Journal of Chemical Physics, 2026, 39(1): 29-40. doi: 10.1063/1674-0068/cjcp2507097
Citation: An Zhao, Qing-Yu Liu, Zhong-Pu Zhao, Zi-Yu Li, Sheng-Gui He. Coupling of a High-Temperature Ion Trap Reactor with an Electron-Spray-Ionization Source for Reactions of Mass-Selected Organometallic Ions†[J]. Chinese Journal of Chemical Physics, 2026, 39(1): 29-40. doi: 10.1063/1674-0068/cjcp2507097

Coupling of a High-Temperature Ion Trap Reactor with an Electron-Spray-Ionization Source for Reactions of Mass-Selected Organometallic Ions†

通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(9)

Article Metrics

Article views(89) PDF downloads(0) Cited by(0)

Access History

Coupling of a High-Temperature Ion Trap Reactor with an Electron-Spray-Ionization Source for Reactions of Mass-Selected Organometallic Ions†

Abstract: Organometallics play a vital role in catalytic and synthetic processes. Understanding the individual elementary steps of the reactions of organo metallic compounds is crucial for the development and rational design of new organometallic reagents and catalysts. Study of gas-phase reactions is one of the key approaches to probing the individual elementary steps under isolated and reproducible conditions. A series of investigations have been reported on the gas-phase reactions between organometallic ions and neutral molecules under room temperature conditions. However, studies about the reactions between organometallic ions and neutral molecules under heating conditions are very limited. In this work, an apparatus with an electrospray ionization source and an ion funnel trap, which can be coupled with a high-temperature linear ion trap reactor, was designed and built. The apparatus can be used to investigate the reactions between organometallic ions and neutral molecules under heating conditions. By using the apparatus, the adsorption reactions of Rh(PPh3)2++CO→Rh(PPh3)2CO+ and CuPPh3++CO2→CuPPh3CO2+ under variable temperature conditions have been conducted. The experiments showed that the reaction rate constant of Rh(PPh3)2++CO increases first and then decreases with increasing temperature. In contrast, the rate constant of CuPPh3++CO2 decreases monotonically as the temperature increases. Density functional theory calculations indicate that the adsorption reaction of Rh(PPh3)2++CO→Rh(PPh3)2CO+ is subject to a small barrier, while CuPPh3++CO2 →CuPPh3CO2+ is barrierless, which is consistent with the experimentally observed temperature-dependent rate constants. The newly built apparatus can thus provide new kinetic information to address reaction mechanisms for organometallic ions.

    • Organometallics own flexibility of coordination numbers and oxidation states involved with related metal centers [1]. Organometallic compounds are thus widely used in important industrial catalytic processes, such as hydrolysis reactions, hydro-carboxylation, hydro-cyanation, olefin transamination, and polymerization of alkanes [2]. Furthermore, in chemical synthesis, organometallics are commonly employed as nucleophiles for forming C−C bonds and driving a variety of other organic reactions [2]. Understanding the individual elementary steps of the reactions of organometallic compounds is crucial for the development and rational design of new organometallic reagents and catalysts. However, the complexities of condensed-phase systems, such as the solvent effect and the occurrence of consecutive reactions hinder the analysis of the microscopic reactivity [1]. The study of gas-phase reactions is one of the key approaches to probing the individual elementary steps under isolated and controllable conditions [3−6].

      Electrospray ionization (ESI) is a soft ionization method that transfers species dissolved in solution into the gas phase. By using ESI, the photoelectron spectra of ligand-protected clusters and the reactivity of organometallic ions have been studied in the gas phase. Tsukuda and coworkers recorded the photoelectron spectra of the ligand-protected homo- and heteronuclear noble metal clusters such as Ag/Au/Pd cluster anions, and these studies provided very important structures for these noble metal systems [7−15]. Schwarz et al. conducted extensive research on the gas-phase reactivities of organometallics toward hydrocarbons. They used a harsh ESI method to obtain the reactivities of metal hydroxides (HMO+, M(OH)+, M = Ti and V) [16], metal hydride (PdH+) and metal deuteride (PdD+) [17] toward methane. They also did research on the reactions between [(C13H25N5)FeN]2+ cations [18] (obtained by a mild ESI method) and alkenes. Furthermore, they carried out investigations on the ligand effects on the reactivities of Ni complexes [19, 20] (NiL+; L = halogen, CHO, OCH3, CN, etc.) and [(CH3CN)ZnO]·+ cations [21] toward methane at ambient temperature. In parallel, they used ESI to produce nanosize vanadium oxide clusters and characterized the structure features by using collision-induced dissociation (CID) [22]. Roithová et al. did research on the reactivities of MgO+ and Mg2O2+ generated by ESI toward methane, propane and butane [23], and were also able to generate copper (III)-oxo complexes in the gas phase and study the reactivities toward H2O, C2H4, and C2H6 [24]. O’Hair et al. studied the reactions between Ni, Pd, and Pt complexes and neutral small molecules such as C2H6 [25], H2O [26], and CH3COOH [27] by CID. McIndoe, Maseras, and coworkers investigated the gas phase reactions of Pd(0) complexes with phenyl halides (PhX; X=F, Cl, Br, I) and observed that the complexes can undergo oxidative addition with PhI and PhBr, and form ion-molecule complexes with PhCl and PhF [28]. Phenyloxenium ion (POI) is an important intermediate in organic reactions, and Pan and coworkers investigated the reactions between gas-phase POI and organic and inorganic small molecules such as benzene, ethanol, and H2O molecules at room temperature [29]. Recently, Sun and coworkers have studied the ligand effects on methane activation by ruthenium species [30]. These investigations have obtained very important reactivity data for gas phase organometallic ions with organic and inorganic molecules.

      The above reported reactions were studied typically under room temperature conditions. The studies of the reactions between organometallic ions and neutral molecules under heating conditions are usually unavailable. Increasing the temperature of reaction systems is commonly used in the condensed-phase studies of organometallic compounds [31−37]. In industry, tin(II) 2-ethylhex-anoate Sn(Ocr)2 catalyzes the ring-opening polymerization of L-lactide to produce biodegradable poly L-lactide at 150−200 °C [32]. In the field of studying CO2 transformation, Prakash and co-workers developed a new method to produce methanol from CO2 hydrogenation using Ru pincer complexes with pentaethylene-hexamine (PEHA) at 155 °C and 7.5 MPa H2/CO2 (3:1) for 40 h in tetrahydrofuran (THF) [35]. Leitner and co-workers achieved the direct CO2 hydrogenation to methanol by using Ru catalysts with a tri-denate phosphine ligand in the presence of triflimide (HNTf2) as an additive in THF at 140 °C and 8 MPa (H2/CO2 = 3:1) [36]. Beller and co-workers reported that direct CO2 hydrogenation was catalyzed by [Co(acac)3] with two equivalents of triphos ligand in the presence of HNTf2 at 100 °C and 9 MPa (H2/CO2 = 7/2) for 24 h in THF/EtOH [34]. Thus, it is of great importance to explore high-temperature reactions between organometallic ions and neutral molecules in order to narrow the gaps between the gas-phase and related condensed-phase reactions.

      To study the kinetics of ion fragmentation under variable temperature conditions in the gas phase, heating devices have been designed and are usually coupled with commercially available mass spectrometers. McLuckey et al. used bath gases with temperatures from 410 K to 493 K to achieve the thermal dissociation of protonated leucine enkephalin (C28H38N5O7+) and its fragment ion b4+ (C22H25N4O5+) in a quadrupole ion trap [38]. McMahon, Williams, and Riveros et al. employed a heated ion cyclotron resonance (ICR) cell [39] coupled with tungsten wires [38, 40, 41] as blackbody infrared radiative sources to study the dissociation of weakly bound cluster ions such as (H2O)nCl−(n = 2 and 3) [39], X2+(H2O)n, (X = Mg, Ca, and n = 8−10) [42], and biomolecules [43]. Gronert et al. developed a temperature variable ion trap mass spectrometer by applying polyimide-coated heating elements to the end-cap electrodes of a commercial ion trap system, and the temperature of ions could be up to 373 K [44]. For these modified commercially available instruments, there is a limitation in that the heating temperature is not very high (typically < 500 K). In 2022, our research group developed a high-temperature linear ion trap reactor (HIT) that can run stably at temperatures up to 873 K. The HIT was coupled with a pulsed laser ablation ion source and a home-made mass spectrometer for studying the high-temperature gas-phase ion-molecule reactions [45]. Directly combining the ESI source, that generates ions continuously with the HIT reactor, that is operated in pulsed mode, is not useful because the time periods of reaction and heating cannot be well defined (determined). An ion accumulation device that collects and stores ions from the ESI source and then ejects the ions into the HIT should be used to switch the continuous ion beam to the pulsed one. In this study, the ion funnel trap (IFT), which has been reported to be efficient in collecting and storing ions was adopted to couple the ESI source and the HIT [46−48]. These ion handling assemblies (ESI source, IFT, HIT, and others) were then coupled with a home-made time-of-flight (TOF) mass spectrometer. The apparatus can be used to investigate the reactions between organometallic ions and neutral molecules under heating conditions. The temperature variable adsorption rate constants of Rh(PPh3)2+ + CO→ Rh(PPh3)2CO+ and CuPPh3+ + CO2→ CuPPh3CO2+ were measured. The experimental results indicate that an energy barrier exists in the adsorption reaction of Rh(PPh3)2+ + CO, but not in the adsorption reaction of CuPPh3+ + CO2, which is supported by theoretical calculations.

    II.   METHODS
    • The experimental instrument is schematically depicted in FIG. 1. The apparatus is consisted of six sub-systems: (1) an ESI source, (2) a tandem ion funnel system, (3) an IFT, (4) a quadrupole mass filter (QMF), (5) a HIT, and (6) a reflectron TOF mass spectrometer (Re-TOF-MS). It is worth noting that sub-systems (4), (5), and (6) are identical to those used in our previous studies [45, 49, 50]. The ions generated by the ESI source are transmitted through the tandem ion funnel system and subsequently introduced into the IFT for accumulation. After accumulation, the ions are selected by the QMF and entered into the HIT for reaction with small molecules (CO2, CO, and so on) under variable temperatures. Finally, reactant and product ions are transferred into the Re-TOF-MS for detection. The details about the instrument are given below.

    • The ESI source was equipped with a fused silica capillary (150 μm ID, 1.6 mm OD, 100 mm length) connected to a gas-tight syringe (1 mL) through a poly-ether ether ketone (PEEK) capillary (100 μm ID, 1.6 mm OD, 500 mm length) [51]. A syringe pump at a proper flow rate (a few μL/min) was used to feed the solutions with proper concentrations (a few μmol/L). High voltages ranging from 3 kV to 5 kV were applied to the solution to form steady charged microdroplets that were then introduced into the vacuum system through a heated inlet stainless tube (1.6 mm ID, 2.1 mm OD, 120 mm length) oriented orthogonally to the instrument’s main axis. The configuration assists in decoupling ion transfer from gas flow dynamics [52]. The inlet was operated at a variable temperature range of 373 K to 573 K to facilitate ion desolvation [52].

    • The tandem ion funnel system was utilized for ion transport across pressure gradients. It comprises two radio frequency (RF) ion funnels made of printed circuit boards (PCBs), connected in series along the same axis and operated at different pressures [52−54]. A high-pressure ion funnel (HPF) was mounted in a vacuum chamber differentially pumped to about 450 Pa by two pumps in parallel (RV14, 14 L/s, KYKY; total pumping speed: 28 L/s). The HPF was constructed by 120 ring electrodes and 120 insulating sheets that were alternately stacked. The repeller electrode I (FIG. 1), made of PCB, was used to deflect the ions and was located at the top of the HPF and 1 mm away from the first electrode of the HPF. All of the electrodes were assembled onto four stainless steel rods for proper alignment. The repeller region (between electrodes II and III) that received ions from the heated inlet and guided ions into the funnel region was composed of 35 electrodes with the same inner diameter (ID) of 50.8 mm. The funnel region (between electrodes IV and V) was consisted of 85 electrodes of which the IDs linearly decreased from 50.8 nm to 2.5 mm [52−54]. The last plate of the HPF acted as a pressure differential aperture for the conductance limit between the first and second vacuum chambers [52−54]. A low-pressure ion funnel (LPF) was assembled in the second vacuum chamber and was differentially pumped to about 120 Pa by another two pumps in parallel (RV14, 14 L/s, KYKY, total pumping speed 28 L/s). The LPF was constructed by 100 ring electrodes and 100 insulating sheets that were alternately stacked. The barrel electrode VI (FIG. 1), made of stainless steel was located at the top of LPF, 1 mm away from the first electrode VII (FIG. 1) of the LPF and the last electrode (V) of the HPF. The first 28 ring electrodes of the LPF had the same ID of 25.4 mm, while the IDs of the following 72 ring electrodes of the LPF decreased linearly from 25.4 mm to 2.5 mm [52−54]. The last electrode VIII (FIG. 1) of the LPF also acted as a conductance limit between the second and the third vacuum chambers.

      The thickness of the electrodes and the insulating sheets making up the two ion funnels was 0.6 mm, and the trace width of each electrode was 2.0 mm [55]. The resonant frequencies were 720 kHz and 1.01 MHz, and peak-to-peak voltages applied to adjacent electrodes for radial confinement of the ions were 260 V and 100 V for the HPF and LPF, respectively. Separate direct current (DC) voltages were applied to the entrance and exit plates of each ion funnel, and DC potential gradients applied for axial ion transport were generated by using a series of 0.487 MΩ resistors connected between neighboring electrodes, and 47 pF capacitors were used for decoupling the DC and RF voltages. The divergent ions from the LPF were focused by two electrostatic lenses and introduced to the IFT. The details of the average electrical potential for ion transformation are shown in FIG. 1.

    • The RF IFT used for the accumulation of ions from the LPF was mounted in the third vacuum chamber pumped to about 10 Pa by a turbomolecular pump (Hipace 300, Pfeiffer) with a pumping speed of 100 L/s. The details of IFT are shown in FIG. 2(a). The IFT was composed of two regions and consisted of 59 ring PCB electrodes and 59 PCB insulator sheets that were alternately stacked. Each of the electrodes and insulator sheets was 0.6 mm thick [56]. The ID of the first electrode (IX, FIG. 1) of IFT was 4 mm, and it was applied with an adjustable DC voltage without RF, while the IDs of the 2nd (X) to the 36th electrodes were uniformly 15 mm, which formed the repeller region. The remaining 23 electrodes, which formed the funnel region, had IDs that decreased linearly from 15 mm to 4 mm. The resonance frequency was 1.22 MHz, and the peak-to-peak voltage applied to adjacent electrodes for the radial confinement of the ions was 80 V for the IFT. The resistors and capacitors on IFT applied for achieving DC potential gradients and decoupling the DC and RF voltages were the same as those in the tandem ion funnel system. The details of the average electrical potential for ion accumulation are also shown in FIG. 1.

      The 10 insulator sheets in the middle of the IFT were designed to be semi-closed so that the buffer gas could enter the IFT (FIG. 2(a)). The IFT operated in two modes: continuous mode and pulsed accumulation mode. Under the continuous model, the axial effective potential of each electrode in the IFT decreased with increasing distance (FIG. 2(b)), which means that ions were driven by the electric field to exit from the IFT. Under the pulsed accumulation model, the last electrode (XI) of IFT was applied a pulsed DC potential of which the magnitute was the same as that applied to the electrode IX. The axial effective potential of each electrode in the IFT decreased with increasing distance and then rose again at the last electrode (FIG. 2(c)). When the ions entered the IFT, their kinetic energies decreased due to multiple collisions with the helium buffer gas that was delivered by a pulsed valve mouted outside of the vacuum chamber. The ions were then trapped and accumulated in the IFT. The time to accumulate the ions was controlled by adjusting the width of the pulse DC potential on the last electrode of the IFT.

    • The HIT was used to study the reactions of mass-selected ions under variable temperature conditions. The design of the HIT was originally described in Refs.[45, 50] and the one used in this study is schematically shown in FIG. 2(d). The HIT contains two ion focusing lens (FIG. 2(d) 1, 9), gas inlet quartz tubes (FIG. 2(d) 2, 3), hexapole rods (FIG. 2(d), 4, RF frequency used was 680 kHz), K-type thermocouples (FIG. 2(d) 5, 6), gas cell (FIG. 2(d) 7), and resistance wire (FIG. 2(d) 8) that was twisted on the gas cell (FIG. 2(d) 7). The bath and reactant gases were injected into the gas cell through fused silica tubes (FIG. 2(d) 2, 3). The temperature of the wall of the gas cell could be measured by a K-type thermocouple (FIG. 2(d) 5) fixed on the wall. Another K-type thermocouple (FIG. 2(d) 6) was stuck on the resistance wire of which the temperature was adjusted by a temperature controller. The temperature reported in this study was measured by the thermocouple fixed on the wall of the gas cell. We assume that the ions in HIT after a few milliseconds of storage had the same temperature as the bath gas and the wall of the gas cell, as there were enough (for thermal equilibrium) numbers of collisions between ions and bath gas and between bath gas atoms (He) and the gas cell [57]. The Re-TOF-MS was consisted of an ion extraction and acceleration assembly, an ion reflecting mirror, and a micro channel plate (MCP detector). The details about the HIT [45, 50] and Re-TOF-MS [49] have been reported in our previous works.

    • The time and pulsing sequences for accumulation [47, 48], reaction [58], and detection of cationic ions are schematically depicted in FIG. 3. For anionic ions the time and pulsing sequences are similar except that the polarity of the electric potential switches. Before the continuous ion beam from the LPF arrived in IFT, a pulse (FIG. 3(a)) of low-pressure (around 5 Pa) cooling gas Helium (99.999%) was delivered into the IFT through a pulsed valve. After the ions entered the IFT, the electric potential of the electrode XII increased (FIG. 3(b) left). Following the electrostatic repulsion from the electrode XII and multiple collisions with helium buffer gas, the ions underwent kinetic energy dissipation, thereby enabling their confinement and subsequent accumulation within the IFT (see FIG. 2(c)). After a period time of accumulation, the electric potential of the electrode XII decreased (FIG. 3(b) right). The ions were then ejected under the potential gradient of the IFT (FIG. 2(b)), and mass-selected by a quadrupole mass filter (FIG. 1) before entering the HIT. Slightly before the ions enter into the HIT, a pulse of helium gas was injected into the HIT through a second pulsed valve (FIG. 3(c)), so that the ions could be cooled in terms of translation and vibration [58], and then the electric potentials of electrodes XIV (FIG. 3(f)), XV (FIG. 3(e)) increased and were used to confine the ions in the axis direction. The electric potential of electrodes XIII (FIG. 3(d)) increased, so that the continuous ion flow from the LPF could be prevented and a relatively accurate heating and reaction time could be determined. After confining the cluster ions in the ion trap for about 1 ms, the reactant gas molecules were delivered through a third pulsed valve (FIG. 3(f)) to interact with the ions in the HIT. After interacting for a period of time (tR), the reactant and product ions were ejected from the trap and entered into the Re-TOF-MS for mass (m/z) and abundance measurements (FIG. 3(h)).

    • All the compounds (reserpine, Rh(PPh3)3Cl, and Cu(PPh3)3Cl) used in this study were purchased from Shanghai Macklin Biochemical Co., Ltd. and Beijing InnoChem Science & Technology Co., Ltd. without further purification. The C33H41N2O9+ cations were generated with ESI of 0.1 mmol/L reserpine in CH3OH:H2O=50:50 and 1% acetic acid [59]. The Rh(PPh3)2+ cations were generated with ESI of 0.1 mmol/L Rh(PPh3)3Cl in acetontrile. The Cu(PPh3)2+ cations were generated with ESI of 0.3 mmol/L Cu(PPh3)3Cl in acetontrile. The CuPPh3+ cations could be generated from the collision between Cu(PPh3)2+ and Ar in the IFT. In these experiments, the samples were introduced through the silica capillary at the flow rates ranging from 3 μL/min to 10 μL/min. Typical ionization conditions were as follows: spray voltage of 4 kV, distance between the silica capillary needle and stainless inlet capillary of 2−3 mm, and temperature of the stainless inlet capillary of 423 K.

    • The density functional theory (DFT) calculations using the Gaussian 16 [60] program package were carried out to investigate the reaction pathways of Rh(PPh3)2+ + CO and CuPPh3+ + CO2. The TPSS functional [61] was employed in conjunction with the basis sets of 6-31G(d,p) [62, 63] for H, C, O, and P atoms, TZVP [64] for Cu atom, and SDD [65] for Rh atoms. Different spin multiplicities were also considered. The reaction mechanism calculations involved geometry optimization of reaction intermediates (Is) and transition states (TSs). The TS optimizations were performed by using the Berny algorithm [66]. The initial guess structures of the TS species were obtained through relaxed potential energy surface (PES) scans using single or multiple internal coordinates [67]. Vibrational frequency calculations were carried out to check that the intermediates and TSs have zero and only one imaginary frequency, respectively. The intrinsic reaction coordinate calculations were conducted to make sure that a TS connects two appropriate intermediates [68, 69]. The zero-point vibration-corrected energies (ΔG298K, ΔG423K for Rh(PPh3)2+ + CO, and ΔG343K for CuPPh3+ + CO2) in units of eV are reported in this work. The energy decomposition analysis (EDA) was performed by using sobEDA and sobEDAw [70, 71] at TPSS-D3(BJ) [72, 73] level/(6-311G(d,p) [74, 75] for H, C, O, P, TZVP for Cu, and SDD for Rh).

    III.   RESULTS AND DISCUSSION
    • In order to test the response of the instrument to the ions generated by ESI, a 100 μmol/L reserpine solution was used as a test sample. FIG. 4 shows the mass spectrum of reserpine under 298 K. The ion with a signal intensity of 1100 mV and a mass-to-charge ratio of 609.3 is C33H41N2O9+, while the ions with mass-to-charge ratios of 397.1 and 448.1 are fragments from the C33H41N2O9+. The experiments about the ion collecting performance of the IFT, and the reactions of Rh(PPh3)2+ + CO and CuPPh3+ + CO2 under variable temperatures are described below.

    • To assess the ion collecting performance of the IFT, a comparative experiment between signal intensities and accumulating times was performed in pulsed accumulation mode, taking Cu(PPh3)2+ as an example. The mass spectra of Cu(PPh3)2+ under accumulation times from 5 ms to 85 ms were recorded (FIG. 5(a)). The total signal intensities of Cu(PPh3)2+ (including isotope peaks) increased from 100 mV to 800 mV when the accumulation time increased from 5 ms to 65 ms (FIG. 5(b)). However, when the accumulation time exceeded 65 ms, the slope of the curve decreased (FIG. 5(b)), which means the accumulation efficiency of IFT decreased and the increase in signal intensity became less pronounced. This might be attributed to the condition that the ion storage capacity of the IFT had been reached [48]. When a sufficient number of ions had been accumulated, the potential of the last electrode was reduced, and the ions were ejected into QMF.

    • The TOF mass spectra for the reaction of the mass-selected Rh(PPh3)2+ (including the isotope peaks) with He and CO under variable temperatures (298 − 623 K) are shown in FIG. 6(a)−(f). Upon the mass-selected Rh(PPh3)2+ cations exposed to He at 298 K, there was no appearance of any product peak (FIG. 6(a)). In the presence of CO, the product ions Rh(PPh3)2CO+ and Rh(PPh3)2(CO)2+ were detected (FIG. 6(b)), indicating the occurrence of CO adsorption reactions (Reactions (1) and (2)).

      As the temperature increased, the proportions of Rh(PPh3)2CO+ and Rh(PPh3)2(CO)2+ gradually decreased at the same reaction time of 10.8 ms (FIG. 6(b)−(f)). The rate constants of Reaction (1) and Reaction (2) under variable temperatures were fitted and determined (FIG. 6(g) and FIG. S1 in Supplementary materials, SM). When the temperature was increased from 298 K to 373 K, the rate constants of Reaction (1) and Reaction (2) increased. However, when the temperature exceeded 373 K, the rate constants of Reaction (2) decreased. As the temperature continued to rise above 473 K, the rate constant of Reaction (1) also decreased.

      Increasing the temperature is usually beneficial for reactions with barriers (barrierless adsorption reactions usually have decreased rate constants with increase of the temperature). One can conclude that Reactions (1) and (2) have reaction barriers from the experimental results in FIG. 6(g). FIG. 7 shows the DFT calculated favorable pathway for the reaction of Rh(PPh3)2+ with CO to generate the Rh(PPh3)2CO+ and Rh(PPh3)2(CO)2+. The C6H5 groups in Rh(PPh3)2+ have weak interactions with CO, resulting in the formation of an encounter complex I1 with a binding energy of ΔH0K = 0.02 eV. Then the CO approaches the Rh atom and is repelled by benzene with a Gibbs free energy barrier of 0.38 eV (I1→TS1) at 298 K, and forms an Rh−C bond in Rh(PPh3)2CO+ with a binding energy of ΔH0K = 1.76 eV. The C6H5 groups in Rh(PPh3)2CO+ also have weak interactions with the second CO to form I2. After the formation of I2, the second CO approaches the Rh atom with a small Gibbs free energy barrier of 0.09 eV (I2→TS2) at 298 K to form Rh(PPh3)2(CO)2+ with a CO binding energy of 1.40 eV (ΔH0K). The theoretical potential energy surface shows that both of the adsorption Reactions (1) and (2) have barriers, which is consistent with the experimental results in FIG. 6(g).

      Additionally, EDA shows that the main interactions of Rh(PPh3)2+-CO and Rh(PPh3)2CO+-CO are the orbital interactions (Table S1 in SM), accounting for 52.6% and 49.7% of the total attractive interaction, respectively. In addition, the electrostatic interactions (36.9% and 37.7%) also make a non-negligible contribution. The exchange and repulsion interactions have adverse effects on combination, offsetting 73.9% and 76.3% of the total attractive interactions for Rh(PPh3)2+-CO and Rh(PPh3)2CO+-CO, respectively.

    • The TOF mass spectra for the reaction of mass-selected 63CuPPh3+ (excluding the isotope peaks) with CO2 under variable temperatures (298 − 363 K) are shown in FIG. 8. The CuPPh3+ is generated from the collisions between Cu(PPh3)2+ and Ar in IFT (FIG. S2 in SM). Upon the interaction of CuPPh3+ with 100 mPa CO2 for about 2.8 ms at 298 K, half of the mass-selected CuPPh3+ ions are depleted, and a new strong peak with a mass larger than that of CuPPh3+ by 44 amu emerges (FIG. 8(b)), indicating the occurrence of CO2 adsorption reaction (Reaction (3)).

      As the temperature increases, the relative intensity of CuPPh3CO2+ decreases gradually (FIG. 8(c)−(f)). At a certain temperature of 343 K, the relative intensity of CuPPh3CO2+ increases as CO2 pressure increases (FIG. 8(h)−(l)). The rate constants of pseudo-first order reactions at different temperatures are fitted and determined (FIG. 9(a) and FIG. S3 in SM). It is found that as the temperature increases, the adsorption rate constants gradually decreases (FIG. 9(a)). The DFT calculated pathway of Reaction (3) (FIG. 9(b)), shows a barrierless adsorption process. The binding energy (ΔH0K) of CO2 on CuPPh3+ is 0.71 eV. Furthermore, the EDA analysis shows that the electrostatic interaction between CO2 and CuPPh3+ is the primary interaction, and accounts for 50.7% of the total attractive interaction. The orbital interaction that accounts for 36.2% of the total attractive interaction can not be neglected either (Table S1 in SM).

    IV.   CONCLUSION
    • In summary, an experimental apparatus with a high-temperature linear ion trap (LIT reactor) coupled with an electron spray ionization (ESI) source for studying reactions between organometallic ions and neutral molecules under variable temperatures was designed and tested. Harnessing an ion funnel trap (IFT) to capture ions from the continuous ion beam produced by ESI, not only enhances the signal intensity per pulse but also enables the accurate determination of the reaction time in the high-temperature LIT. The variable-temperature experimental results show that the rate constant of Rh(PPh3)2+ + CO increases first and then decreases with increasing temperature, while the rate constant of CuPPh3+ + CO2 decreases monotonically. The experimental results indicated that the adsorption reaction of Rh(PPh3)2+ + CO→Rh(PPh3)2CO+ has a small barrier, while CuPPh3+ + CO2→CuPPh3CO2+ is barrierless, which is supported by density functional theory calculations. The apparatus with a temperature variable reactor designed in this work can be used to obtain the reaction mechanisms that may not be obtained by instruments with temperature fixed reactors to investigate the reactions between organometallic ions and neutral molecules under high-temperature conditions.

      Supplementary materials: Rh(PPh3)2+/product ion signals vs. CO pressure (298–623 K) and CuPPh3+/product ion signals vs. CO2 pressure (298–366 K) (with fitted data and rate constants); Cu(PPh3)2+ dissociation mass spectra (He/Ar, 298 K, m/z 250–650), and EDA results for three ion-small molecule systems (energy components).