Jian Tang, Zecong Qin, Yunfei Li, Long Tian, Bin Wang, Qingbo Zheng, Yujie Zheng, Junsong Xie, Han Gao, Yuan Yao, Jun Li, Tianhao Wang, Junpei Zhang, Xin Tong. A dual-Helmholtz-coil magnetic field system for polarized 3He[J]. Chinese Physics B, 2026, 35(7): 070704. doi: 10.1088/1674-1056/ae4580
Citation: Jian Tang, Zecong Qin, Yunfei Li, Long Tian, Bin Wang, Qingbo Zheng, Yujie Zheng, Junsong Xie, Han Gao, Yuan Yao, Jun Li, Tianhao Wang, Junpei Zhang, Xin Tong. A dual-Helmholtz-coil magnetic field system for polarized 3He[J]. Chinese Physics B, 2026, 35(7): 070704. doi: 10.1088/1674-1056/ae4580

A dual-Helmholtz-coil magnetic field system for polarized 3He

  • Received Date: 20/11/2025
    Accepted Date: 23/01/2026
    Available Online: 01/07/2026
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(10)  /  Tables(1)

Article Metrics

Article views(155) PDF downloads(0) Cited by(0)

Access History

A dual-Helmholtz-coil magnetic field system for polarized 3He

Abstract: This study focuses on developing a dual-Helmholtz-coil magnetic field system for polarized 3He to generate a uniform magnetic field. A theoretical analysis of the mechanism by which dual Helmholtz coils produce a uniform magnetic field was first conducted. Based on these findings, the magnetic device was designed with the finite element analysis software COMSOL, employing an optimization algorithm to efficiently set current densities and minimize the transverse gradient in the central region. The simulated transverse gradient achieved was 9.54 × 10−5 cm−1 in the 20 cm × 20 cm × 8 cm central region, and the experimental measurement yielded a value of 6.31 × 10−4 cm−1. Additionally, lifetime testing of two polarized 3He cells revealed lifetimes of 151.0 ± 1.8 h and 206.2 ± 4.4 h, demonstrating the excellent magnetic field uniformity of the device. The current work integrates theoretical, simulated, and experimental results. It advances polarized 3He magnetic technology, offers an upgrade solution for off situ polarized 3He pumping stations to extend the volume of the uniform magnetic field without enlarging the coil size, and provides a method for using polarized 3He on neutron beamlines requiring large acceptance angles.

1.   Introduction
  • Polarized 3He has a wide range of applications in particle physics and nuclear physics,[1,2] condensed matter physics,[3,4] nuclear magnetic resonance imaging,[5] magnetometry,[6,7] and polarized nuclear fusion fields.[8] Moreover, polarized 3He can be used as a neutron spin filter (NSF) in polarized neutron experiments.[9–11] Currently, two commonly used methods for polarizing 3He are spin-exchange optical pumping (SEOP)[12–15] and metastability-exchange optical pumping.[16] In these methods, the magnetic system primarily defines the polarization direction of 3He and maintains its polarized state. Helmholtz coil magnetic field systems are employed to generate highly uniform magnetic fields, mainly in off situ polarized 3He pumping systems[15] and wide-angle polarized 3He devices.[17,18] Institutions such as ORNL,[19] NIST[20] and CSNS[21] have adopted single-pair Helmholtz coil designs for their off situ polarized 3He pumping systems. Facilities including ISIS,[17] JCNS[18] and ILL[22,23] utilize a multi-pair Helmholtz coil configuration in the design of wide-angle polarized 3He devices. Wide-angle polarized 3He devices are essential for separating coherent, nuclear spin incoherent and magnetic scattering,[24] as well as for measuring magnetic excitations[25] and related phenomena. Typical neutron scattering experiments require a 10 cm diameter homogeneous field region. However, wide-angle cells often have an external diameter exceeding 20 cm,[26] a scale at which single-pair coils cannot meet the field-uniformity requirements. Therefore, advancing domestic wide-angle polarized 3He NSF technology necessitates a magnetic field device capable of providing a large-volume uniform field, both for off situ pumping systems and for future wide-angle polarized 3He devices.

    For a typical polarized 3He pumping station, when the pumping stops, the 3He polarization P⁢3He0(t=0) is gradually relaxed over time t in an exponential decaying form as described by

    The relaxation time T1 (also termed the lifetime) of polarized 3He is influenced by three main factors:[27,28] wall collisions, magnetic dipole–dipole interactions, and magnetic field inhomogeneity, as shown below:

    Here, p is the polarized 3He gas pressure in bars. Once the polarized 3He cell has been sealed, the first two factors remain fixed. Therefore, magnetic field uniformity becomes crucial for extending the relaxation time. Consequently, reducing the standardized transverse gradient (|∇Bx|2+|∇By|2)/B02 is key to extending the relaxation time of polarized 3He, and it should usually be reduced to the order of 10−4 cm−1.

    A common choice for generating a magnetic field in an off situ pumping station is to use a pair of Helmholtz coils.[19–21] However, to provide a standardized gradient on the order of 10−4 cm−1, the size of the Helmholtz coils is usually so large that it increases the footprint of the off situ pumping station. To address this, this study investigates a nested dual-Helmholtz-coil configuration. Theoretical derivations demonstrate that the uniformity of dual Helmholtz coils is superior to that of a single pair of Helmholtz coils. Meanwhile, the finite element analysis (FEA) simulation software COMSOL[29] and its optimization module were used to determine electric current parameters to minimize the gradient in the central region (length × width × height = 20 cm × 20 cm × 8 cm). The nested dual Helmholtz coils were designed according to these simulation results. During the commissioning, a three-dimensional stepper motor combined with a gaussmeter was employed to map the magnetic field in the central region.

    The novelty of the presented approach lies in its compact, nested dual-coil geometry combined with automated current-density optimization. In contrast to conventional designs that employ spatially separated main and compensation coils, the proposed system integrates two concentric Helmholtz pairs with opposing currents. This configuration promotes more effective cancellation of higher-order field gradients within a limited footprint. In simulation, a transverse gradient of 9.54 × 10−5 cm−1 is achieved in the target volume, which represents an order of magnitude improvement over a single Helmholtz pair (2.05 × 10−3 cm−1). When evaluated under a fixed gradient threshold, the designed system provides approximately 2.5 times larger usable uniform field volume than conventional multi-coil designs, offering a distinct advantage for applications that require extended uniform fields without increasing the overall system scale. The performance of the dual-Helmholtz-coil design has also been experimentally validated. The lifetimes of two polarized 3He cells are simultaneously measured by the free induction decay (FID) method.[30,31] Compared with a single pair of Helmholtz coils with the same diameter, this dual-Helmholtz-coil design significantly enlarges the central uniform region, making it suitable for upgrading the magnetic system on off situ pumping stations using the SEOP method. Moreover, the magnetic field design serves as a wide-angle polarized 3He device, featuring a large neutron scattering angle.

2.   The uniformity theory of dual Helmholtz coils
  • Guo[32] derived the magnetic field at the spatial point P(r, θ, ϕ) for a pair of coils arranged on a sphere of radius R with current I. Equation (4) presents the expression of the magnetic field in cylindrical coordinates at point P(ρ, θ, z), and Fig. 1 illustrates the coil layout within this coordinate system:

    To obtain the most uniform magnetic field in the central region, the coefficients of the expansion must be minimized. The n = 0 term provides a constant field, independent of ρ. The condition for nullifying the n = 1 term is cosα=1/5, which leads to Bρ = 0 and defines the classic Helmholtz coil configuration, where the coil radius R sin α equals the separation distance d. The enhanced uniformity of the dual-Helmholtz-coil design is achieved through the systematic cancellation of higher-order magnetic field gradients. While a single Helmholtz pair minimizes the first-order (dipole) variation by satisfying d = R, it retains significant second-order (quadrupole) gradients. The limitation of a single Helmholtz pair design is overcome via a gradient superposition principle: two concentric coil pairs with opposing currents are configured so that their quadrupole contributions destructively interfere. To cancel the n = 2 (quadrupole) term, a dual-coil system is employed. The total magnetic field is the superposition of the fields from two Helmholtz pairs with radii R1 sin α and R2 sin α, carrying currents I1 and I2, respectively. The explicit forms of Bz1,Bz2,Bρ1,Bρ2 follow

    with the corresponding radii and currents. To derive the cancellation condition for the n = 2 term, the expansions of Bρ1 and Bρ2 are written explicitly. For the dual-coil system, the total Bρ is the sum of two such series with parameters (I1, R1) and (I2, R2). The condition for canceling the n = 2 term is that the sum of the n = 2 contributions from both coils vanishes, i.e., Bρ1 + Bρ2 = 0. The n = 2 term of each coil is proportional to I/R5 and shares the same angular dependence. Explicitly, Bρ1∝I1R15⋅sinαP51(cosα)P41(cosθ),Bρ2∝I2R25⋅sinαP51(cosα)P41(cosθ). Setting the sum to zero and canceling the common non-zero factors (the angular part and sinα) yields the condition: I1R15+I2R25=0.

    Thus, when the currents in the two coil pairs satisfy this relation, the quadrupole (n = 2) gradient is eliminated at the center. This allows the dual-Helmholtz-coil design to enhance the magnetic field uniformity while maintaining the same outer diameter as a single pair.

3.   The design and simulation of dual Helmholtz coils
  • Based on the theoretical framework established in Section 2, the design and simulation stage was initiated. The primary objective was to translate the theoretical dual-coil concept into a practical magnetic field device and to optimize its performance using FEA.

    A finite-element model of the dual-Helmholtz-coil system was established in COMSOL Multiphysics®[29] to ensure reproducible simulations. The coil geometry (Fig. 2) was constructed according to the parameters in Table 1, incorporating the actual rectangular cross-section of the windings (2 mm × 4 mm). The magnetostatic Magnetic Fields (mf) interface was applied within a spherical air domain of radius 3.0 m, sufficiently large to avoid boundary effects. The external boundary condition was set to Magnetic Insulation to simulate free-space conditions.

    A user-controlled free-tetrahedral mesh with an element size of 8 mm was applied in the coils and the central region of interest. Mesh independence was verified by recalculating the average transverse gradient in the central target region (20 cm × 20 cm × 8 cm) using a systematically refined mesh. The variation in the computed gradient was less than 0.3%, confirming that the chosen mesh yields grid-independent results.

    First, initial parameters were set according to the theoretical analysis. The current density J1 of coil B1 was fixed at 1.25 A/mm2. The initial current density J2 of coil B2 was determined from the theoretical condition Bρ1 + Bρ2 = 0, resulting in J2 = 0.71 A/mm2. Using this initial parameter set, models of the B1 and B2 coils were built in COMSOL and simulated. This initial simulation gave a magnetic field of 9.6 G and an average volumetric gradient of 2.59 × 10−4 cm−1 in the central target region (20 cm × 20 cm × 8 cm). For reference, the theoretical (analytical) values for the dual-coil system at J2 = 0.71 A/mm2 are a central field of 9.6 G and a mean gradient of 1.34 × 10−4 cm−1 (from Eq. (5)) (Figs. 3(a) and 3(b)). The corresponding single-coil (B1 only) theoretical values at the same J1 = 1.25 A/mm2 are 22.1–22.8 G and 1.48 × 10−3 cm−1, respectively (Figs. 3(c) and 3(d)). However, this initial simulated mean gradient, 2.59 × 10−4 cm−1, was about 1.9 times larger than the purely theoretical prediction of 1.34 × 10−4 cm−1. The discrepancy stems from the idealization of the coils as line currents in the theoretical model, whereas the actual coils possess a finite rectangular cross-section. This finite geometry alters the current distribution and the effective winding distance. Two primary effects arise: (1) the current is distributed over a finite volume, which modifies the magnetic field profile, particularly near the coil edges; and (2) the effective radial position of the windings relative to the central region is shifted. Collectively, these effects distort the field geometry, rendering the simple line-current model inadequate for precisely predicting the gradient within our target volume.

    To compensate for the effects of the finite geometry, an automated optimization was performed. The optimization aimed to minimize the volume-averaged normalized transverse gradient in the central region, defined as the objective function:

    where V is the volume of the central region. The optimization was performed using COMSOL’s built-in COBYLA (Constrained Optimization BY Linear Approximation) algorithm. The current density for coil B1 was fixed at 1.25 A/mm2; the current density J2 of coil B2 (initial value 0.71 A/mm2) served as the optimization variable, constrained within ±3 A/mm2 of its initial value. The convergence criterion was set to a relative change in fobj of less than 1 × 10−6. The algorithm converged to an optimal current density J2 = 0.68 A/mm2, representing a 4.2% adjustment from the theoretical starting point. This adjustment effectively rebalances the magnetic contributions of the two coils, counteracting the distortion caused by their finite size. A final simulation with the optimized parameters (J1 = 1.25 A/mm2, J2 = 0.68 A/mm2) yielded significantly improved results: a central field of 10.1 G and a much-reduced mean gradient of 9.54 × 10−5 cm−1 (Figs. 4(a) and 4(b)). For comparison, a single pair of B1 Helmholtz coils operated at a current density of 1.25 A/mm2 was also simulated (Figs. 4(c) and 4(d)). The single-coil system produced a field of 22.1–22.8 G and a mean transverse gradient of 2.05 × 10−3 cm−1 in the same central region. Clearly, the dual-Helmholtz-coil configuration provides a substantially more uniform magnetic environment than a single coil pair.

    ISIS[17] and ILL[22] also employed two sets of coil designs to create magnetic field devices. For a typical Helmholtz coil, a uniform region is rendered at the center of its region, proportional to the size of the coil itself. Therefore, it is impossible to directly compare the magnetic field systems. A dimensionless figure of merit, denoted as A, is defined and employed here as the ratio of two volumes: the volume within which the magnetic field gradient at the device center is less than or equal to a fixed gradient threshold (∼10−4 cm−1) and the total volume occupied by the magnetic field device itself.

    The ISIS magnetic system comprises a main coil pair (outer diameter 695 mm, 300 turns) and a counter-wound compensation pair (outer diameter 416 mm, 69 turns), with axial separations of 334 mm (compensation) and 410 mm (main). The cell-placement region is a cylinder with an internal diameter of 10 cm, an external diameter of 24 cm, and a height of 17 cm. A COMSOL model of this setup was constructed, and the average gradient within the cell position region was optimized, yielding a best value of 7.97 × 10−4 cm−1, consistent with the reported performance of the ISIS device (8 × 10−4 cm−1). Subsequently, the volume corresponding to a gradient ≤ 10 × 10−4 cm−1 was simulated. Figure 5 shows the transverse gradient distribution in the Z = 0 plane, with the red contour marking the 10 × 10−4 cm−1 level. Panel (a) corresponds to the ISIS model, yielding AISIS = 5.39%, while panel (b) shows our dual-Helmholtz-coil model, for which Adual = 13.26%. This result indicates that, for the same overall device volume, the uniform-field region of the dual-Helmholtz-coil design is about 2.5 times larger than that of the ISIS style system.

    The comparison above relies on a fixed gradient threshold (10 × 10−4 cm−1) to define the uniform field boundary. To verify that the concluded advantage is not an artifact of this particular choice, the sensitivity of the volume ratio Adual/AISIS to the threshold value was examined. The ratio remains between 2.3 and 4.0 across a wide range of thresholds from 1.0 × 10−3 cm−1 down to 1.0 × 10−4 cm−1, with a value of 2.5 at the originally selected threshold. This confirms that the dual-Helmholtz-coil configuration consistently provides a substantially larger uniform field volume, and the conclusion is robust against variations in the defining gradient criterion.

4.   Uniformity measurements of dual Helmholtz coils
  • Following the simulation-guided design and optimization described in Section 3, the dual-Helmholtz-coil system was fabricated and experimentally characterized. This section aims to validate the predicted field uniformity through direct magnetic field mapping and, more importantly, by measuring the relaxation lifetime of polarized 3He cells — a direct indicator of field quality. The B1 and B2 coils were wound according to the parameters in Table 1 and fixed with epoxy resin approximately 1.8 mm thick. The corresponding mechanical supports and adjustment mechanisms for the B1 and B2 coils were designed based on Fig. 2(a). The mechanical design allows the central heights of the B1 and B2 coils to be adjusted within ± 3 mm of the theoretical design value. It also enables each coil to rotate ± 3° around the z-axis. After the magnetic field device was designed, fabricated, and assembled, the mechanical positioning structure was used to fine-tune the coil positions based on the actual magnetic field measurements. In addition, the currents in the B2 coils were carefully adjusted based on the test results of polarized 3He. A three-dimensional stepper motor combined with a gaussmeter was used to measure the magnetic field. Measurements were performed in the central core region (20 cm × 20 cm × 8 cm) of the device, with a resolution of 5 mm in each direction. The measured magnetic field in the central core region was 10.1–10.2 G, with a mean volumetric transverse gradient of 6.31 × 10−4 cm−1 (Fig. 6). The difference from the simulated gradient (9.54 × 10−5 cm−1) primarily arises from practical experimental factors. These include coil-positioning uncertainties (contributing up to ∼ 10−4– 10−3 cm−1), DC current ripple, environmental stray fields, and instrumental limitations of the gaussmeter. Controlled perturbation simulations confirm that the combined effect of these quantifiable tolerances is on the order of 10−4 cm−1, consistent with the observed discrepancy. Hence, the deviation is attributable to typical experimental variations and does not undermine the design or simulation validity.

    After the actual magnetic field measurement, two polarized 3He cells (Cell20221230 and Cell20230324), produced at the gas filling station at CSNS,[33] were used to test the magnetic field uniformity. Cell20221230 and Cell20230324 both have a diameter and length of 8.0 cm, with the cell wall thickness being approximately 0.4 cm. The pressures of Cell20221230 and Cell20230324 were determined by neutron transmission measurements according to[34–36]

    where T0 is the neutron transmission for the unpolarized 3He, Te is transmission through an empty glass, p is the 3He pressure in bar, λ is the neutron wavelength in angstroms, and l is the path length inside the cell in centimeters. By plotting the neutron transmission as a function of wavelength (Fig. 7) and fitting using Eq. (7), the measurement results show that the pressures of Cell20221230 and Cell20230324 were 1.93 bar and 2.52 bar, respectively, at the room temperature of 25 °C. Subsequently, the lifetimes T1 of the two cells were tested on the pumping station at CSNS,[21] which features a Helmholtz coil system with an outer diameter of 86 cm. Within the central region of the Helmholtz coils (10 cm × 10 cm × 10 cm), the mean transverse gradient simulated was 9.05 × 10−5 cm−1, as shown in Fig. 8, and the measured mean volume transverse gradient was 5.8 × 10−4 cm−1.[21] The lifetimes of Cell20221230 and Cell20230324, measured at the pumping station, were 165.1 ± 1.4 h and 251.9 ± 2.4 h, respectively. The measured lifetimes are shown in Fig. 9. Subsequently, the dual-Helmholtz-coil magnetic field device was used to test Cell 1 and Cell 2. The FID method[30,31] was used separately to pick up signals from Cell20221230 and Cell20230324. After acquiring the FID signal, the T2 value of the FID signal quickly reflects the magnetic field uniformity at the FID coil position,[13] allowing an initial assessment of the magnetic field uniformity in the central region. Finally, the fitted lifetime T1 was used to definitively determine the magnetic field uniformity.

    Subsequently, the two cells were placed in the dual Helmholtz coils for performance demonstration. Cell20221230, acting as the polarizer, was placed inside an aluminum alloy box, while Cell20230324, serving as the analyzer, was mounted on a bracket approximately 4 cm away from the box. The two cells together were centered in a region of 20 cm × 20 cm × 8 cm. The placement of Cell20221230 and Cell20230324 and the magnetic field device is shown in Fig. 10(a). The FID method[30,31] was used simultaneously to measure the signals from Cell20221230 and Cell20230324 and to fit T1. Cell20221230 exhibited a lifetime of 151.0 ± 1.8 h, while Cell20230324 demonstrated a lifetime of 206.2 ± 4.4 h. The specific test results are shown in Fig. 10(b). The lifetimes measured in the dual-Helmholtz-coil system are close to those obtained on the filling station but present a decrease. This can be attributed to two aspects. In addition to the fact that the measured transverse magnetic field gradient is slightly higher than that of the filling station, the simultaneous FID measurements by employing two pick-up coils would have further altered the field uniformity and aggravated the polarization relaxation. On the other hand, the lifetimes measured in the dual-Helmholtz-coil system are sufficiently long, and the uniform field area is extended to one three times larger. Moreover, compared with conventional neutron beam transportation solenoid devices (neutron scattering angle < 10°),[21] this system achieves a much larger neutron acceptance angle of approximately 20°. These advantages make the current design suitable for future application on neutron beamlines.

5.   Conclusion
  • Combining the theoretical, simulated, and experimental results presented in the preceding sections, we have successfully developed and validated a dual-Helmholtz-coil magnetic field system for polarized 3He applications. The key findings and implications are summarized as follows. A dual Helmholtz magnetic field device for polarized 3He was successfully proposed and developed, comprising two pairs of Helmholtz coils. Theoretical derivations show better uniformity in the dual coils than in a single pair with the same outer diameter. The dual-Helmholtz-coil system was designed using the FEA software COMSOL, where an optimization algorithm was employed to minimize the gradient in the central region. This resulted in a simulated magnetic field uniformity of 9.54 × 10−5 cm−1 within a 20 cm × 20 cm × 8 cm volume. The measured gradient of the final device was 6.31 × 10−4 cm−1.

    Finally, two polarized 3He cells (Cell20221230 and Cell20230324) were placed in the device’s central region for lifetime testing by the FID method. The lifetimes of the two cells were 151.0 ± 1.8 h and 206.2 ± 4.4 h, respectively. Combined with Cell20230324, the device enables a neutron scattering angle of approximately 20°. The magnetic field design offers a solution for off situ3He pumping stations to extend the volume of the uniform magnetic field without enlarging the size of the coils, and provides a method for using polarized 3He on neutron beamlines requiring large acceptance angles.

Figure (10)  Table (1) Reference (36)