2026 Volume 35 Issue 7
Article Contents

Encai Zhong, Yan Zheng, Hongli Liu, Mingming Liu, Jinxu Hong, Yijun Luo, Ke Deng, Jie Zhang, Zehuang Lu. Noise evaluation and optimization for a cryogenic sapphire oscillator with 10−15 level stability[J]. Chinese Physics B, 2026, 35(7): 070601. doi: 10.1088/1674-1056/ae53ba
Citation: Encai Zhong, Yan Zheng, Hongli Liu, Mingming Liu, Jinxu Hong, Yijun Luo, Ke Deng, Jie Zhang, Zehuang Lu. Noise evaluation and optimization for a cryogenic sapphire oscillator with 10−15 level stability[J]. Chinese Physics B, 2026, 35(7): 070601. doi: 10.1088/1674-1056/ae53ba

Noise evaluation and optimization for a cryogenic sapphire oscillator with 10−15 level stability

  • Received Date: 03/02/2026
    Accepted Date: 11/03/2026
    Available Online: 01/07/2026
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(9)

Article Metrics

Article views(209) PDF downloads(2) Cited by(0)

Access History

Noise evaluation and optimization for a cryogenic sapphire oscillator with 10−15 level stability

Abstract: Cryogenic sapphire oscillators (CSOs) simultaneously achieve ultra-low phase noise and outstanding frequency stability, making them widely used in precision measurement applications. In this study, performance evaluation and noise analysis are conducted on a home-built CSO. When compared with an optically generated microwave (OGM) reference, its synthesized 100 Hz signal demonstrates a relative frequency instability (Allan deviation) of 2.1 × 10−15 at an averaging time of 70 s. After evaluation of the dominant noise sources, it is shown that the short-term frequency stability is limited by the 1.4 Hz vibrational noise from the pulse cryocooler, resulting in a limitation of 1.2 × 10−15τ−1. At an averaging time of 100 s, the contribution of temperature fluctuation noise becomes prominent, limiting the frequency stability to 10−16. In the long term, the stability of the voltage reference in the microwave power servo becomes the limiting factor. This research provides guidelines for the development of improved CSOs and their further applications in precision measurement physics.

1.   Introduction
  • A defining characteristic of ultra-stable microwave signal sources is their exceptionally high frequency stability, which is typically quantified in the time domain using the Allan deviation of relative frequency fluctuations. Currently, the primary methods for generating such ultra-stable microwave signals are the optical frequency division scheme based on frequency-stabilized lasers and cryogenic sapphire oscillators (CSOs).[1–5] Microwave signals produced by these two approaches achieve fractional frequency instability at the level of 10−15 to 10−16 for one-second averaging times.[6,7] For a 10 GHz carrier, this corresponds to an absolute frequency fluctuation of a few to tens of microhertz. Benefiting from such high frequency resolution, a key application of ultra-stable microwave sources is serving as the local oscillator for cesium fountain clocks.[8] They are used to precisely measure clock transition frequencies, thereby pushing the frequency stability of microwave clocks to the mid-10−14 level and beyond.[9,10]

    Distinct from the ultra-stable microwave signals derived from lasers stabilized on Fabry–Perot cavities, the resonator of a CSO exhibits a frequency-temperature turning point (FTTP),[11] which grants it superior medium- to long-term frequency stability. Moreover, as an all-microwave system, it offers significant advantages in terms of development cost and operational simplicity. As a result, CSOs can serve as frequency references in very long baseline interferometry (VLBI),[12] where their performance rivals that of active hydrogen masers, maintaining data coherence over extended averaging times in deep-space observations. Furthermore, compared to conventional laboratory signal sources, the exceptionally pure signal generated by a CSO is also suitable for quantum computing operations, as it is less likely to disrupt system coherence during qubit manipulation.[13] The cylindrical design of the sapphire resonator enables electromagnetic waves to propagate in a whispering-gallery mode within the sapphire crystal. This property allows it to function as a sensor for testing Lorentz symmetry violation and as a detector for dark matter.[14,15]

    To measure the frequency stability of a CSO, it is necessary to employ an additional reference source with better frequency stability,[7] or to use a three-cornered hat method with multiple sources.[16] Alternatively, the time-delay decorrelation method may be considered.[17,18] However, this approach is less convenient for signals in the microwave band, as the difficulty lies in constructing a time-delay line at microwave frequencies. In comparison, using another independent CSO or an active hydrogen maser as the reference oscillator to evaluate the performance of the device under test (DUT) remains a reliable solution. With the maturity of optical frequency division technology in recent years,[19,20] optically generated microwaves can inherit the exceptional frequency stability of ultra-stable lasers (USLs). Consequently, they can also serve as a frequency reference for measuring the frequency stability of CSOs.

    As outlined above, the significant application value of CSOs has made their development, evaluation, and characterization key focuses in both academic and industrial research. This paper introduces the principles and experimental setup of the CSO, including the design of the microwave resonator, oscillation loop, and frequency-locking techniques. In the Results section, measurements of the CSO’s frequency stability and phase noise are presented. To evaluate the frequency stability of the CSO, an optically generated microwave (OGM) based on optical frequency division is applied, which typically achieves stability on the order of 10−16 and phase noise of −105 dBc/Hz at 1 Hz in the X-band.[4,6,20–22] This setup meets the requirements for measuring the CSO’s phase noise down to −95 dBc/Hz at 1 Hz.

    Additionally, theoretical analysis and experimental verification of the CSO’s noise sources are conducted, covering frequency-locking noise, microwave power fluctuation noise, temperature fluctuation noise, and 1.4 Hz vibrational noise from the pulse-tube cryocooler. By addressing these noise sources in future work, the frequency stability of the CSO system could be improved to the 10−16 level, with phase noise below −105 dBc/Hz at 1 Hz, comparable to that of the most advanced systems developed at the University of Western Australia (UWA) and the Franche-Comté Electronique Mécanique Thermique et Optique-Sciences et Technologies (FEMTO-ST).[1,2]

2.   Experimental principles and setup
  • The core of the CSO is a sapphire-loaded copper cavity (SLC), which serves as the resonator and provides the foundation for its frequency stability. It selects the microwave signal resonant with a high-order whispering-gallery mode (WGM) of the sapphire cylinder in the X-band. Together with a low-noise amplifier (LNA) and a phase shifter, it forms an oscillation loop. By satisfying the Barkhausen criterion, it can generate a stable oscillating signal. The sapphire resonator acts as a frequency discriminator for the frequency-locked CSO. The microwave signal fed into the resonator, upon reflection, carries information that identifies the deviation between the resonant frequency and the oscillation frequency. The frequency-locking technique based on this concept was proposed by Pound and has since evolved into the Pound–Drever–Hall (PDH) technique for frequency stabilization of lasers.[23,24]

    The application of the Pound frequency discrimination (PFD) technique in the CSO is illustrated in Fig. 1. Its principle can be described as follows: the carrier at frequency fc, represented as ei2πfct, undergoes frequency modulation with modulation depth φm and frequency fm. Assuming a small modulation depth, this modulated signal can be approximated as a three-frequency wave, consisting of J0(φm)ei2πfct and J1(φm)ei2π(fc±fm)t, where J0 and J1 are the expansion coefficients of the Bessel functions. Since the resonator linewidth Δf0 is much smaller than fm, the double sidebands at frequencies fc ± fm are fully reflected. Only the carrier at fc interacts with the cavity and is partially reflected, expressed as S11(fc)J0ei2πfct, where S11 is the reflection coefficient of the resonator. The reflected three-frequency wave is separated by a circulator and detected to obtain the raw error signal as

    Here, γPD (mV/mW) and Pdet represent the voltage conversion sensitivity and the detected power of the power detector (PD), respectively. S11 is characterized by the coupling coefficient β1, quality factor Q, and resonant frequency f0 of the sapphire cavity. In Eq. (1), the raw error signal can be extracted via demodulation techniques. With weak modulation (φm ∼ 0.1), the approximation J0J1 ≈ φm/2 can be applied, and the frequency-discriminator error signal is expressed as

    where Dν is the frequency-discrimination coefficient. In Fig. 1, the error signal can characterize the deviation between the oscillation frequency and the resonance center as Im S11. This signal is fed back by the servo controller to the voltage-controlled phase shifter (VCP) in the oscillation loop, thereby pulling the oscillation frequency to the resonance center.

  • The sapphire cylinder with a diameter of 50 mm and a thickness of 30 mm is placed inside a copper cavity, with a pair of magnetic probes used to excite the WGH15,0,0 mode. In Fig. 1, the measurement of S21 by a vector network analyzer shows that the insertion loss is 15 dB. Despite this weak coupling causing a low coupling coefficient (β1 ≈ 0.45), it enables a high loaded Q-factor, reaching 5.1 × 108.

    As analyzed in the phase noise model of a phase-locked oscillator,[7] the sapphire resonator serves as the frequency stabilizer of the oscillator. However, the frequency characteristics of the resonator are influenced by environmental factors, including temperature fluctuations, microwave power fluctuations, vibration, and mechanical deformation. Among these factors, the effects of the first two are noticeable, as demonstrated by the measurement illustrated in Fig. 2.

    By scanning the surface temperature of the sapphire resonator or the microwave power fed into the cavity, corresponding shifts in its resonant frequency can be measured. A low-phase-noise signal generator (E8257D), referenced to an active hydrogen maser, provides a 10.7875 GHz signal as the frequency reference for measuring the microwave frequency of the CSO. Thus, the frequency shift is observed as a variation in the 986 kHz beat-note frequency.

    The temperature-induced frequency shift is shown in Fig. 2(b). The temperature of the sapphire resonator is scanned from 3.6 K to 4.6 K, with a frequency-temperature turnover point TFTTP observed at 4.330 K. By stabilizing the resonator at TFTTP with a cryogenic temperature servo, the temperature fluctuation noise can be effectively suppressed. In Fig. 2(c), the microwave power fed into the resonator is varied by adjusting the bias voltage of a voltage-controlled attenuator (VCA), yielding a microwave power-induced frequency shift sensitivity of −0.35 Hz/mW.

    To stabilize the microwave power, a directional coupler extracts 1/10 of the microwave power, which is converted into a voltage signal via a cryogenic PD (DT8012). This voltage is locked to a voltage reference in a low-noise controller (SIM960), thereby compensating for power fluctuation noise by adjusting the bias voltage of the VCA.

    In the frequency-locking process, a 100 kHz modulated RF signal output from a lock-in amplifier (LIA) is applied to a VCP with a bandwidth of MHz, enabling the generation of a coherent three-frequency wave. After being reflected by the resonator, the signal is detected by a cryogenic PD, yielding a 100 kHz error signal synchronized with the modulation RF. After demodulation and amplification by the LIA, the error signal described by Eq. (2), V(Δf) = ADνΔf, is obtained. Here, A is the gain provided by the LIA. By feeding this error signal back to the bias voltage of the VCP, the oscillation frequency is locked to the resonance center.

    In Fig. 2(d), once the frequency lock is achieved, the fluctuation noise of the error signal is suppressed by almost one order of magnitude at Fourier frequencies below 1 Hz.

3.   Frequency stability and phase noise
  • In this study, two CSOs were developed by the National Gravitation Laboratory (NGL) at Huazhong University of Science and Technology, designated as CSO-1 and CSO-2.[7] The frequency stabilities of CSO-1 measured with CSO-2, an H-maser, and an OGM are shown in Fig. 3(a). When compared with the OGM,[21] the Allan deviation of the relative frequency fluctuations reaches 2.1 × 10−15 at an averaging time of 70 s. Here, the frequency stability of the OGM is derived from a 30 cm Fabry–Pérot cavity with a stability of 1.9 × 10−16.[25] A commercial frequency transmitter (Synchronization Technology Ltd, STW-FT series) is used to transmit the synthesized 100 MHz signal from CSO-1 to the distant laser lab. A commercial phase noise analyzer (PNA, Microchip 53100A) measures the 100 MHz signal derived from the OGM (9 GHz) via a 1/90 frequency divider, comparing it with the 100 MHz signal from CSO-1. For τ < 5 s, the frequency stability is limited by the noise floor of the frequency transmitter, exhibiting a trend of 1 × 10−14τ−1.

    In addition, the frequency dividers also impose a limit on the short-term stability, with a single divider exhibiting a trend of 8 × 10−15τ−1. This explains the observed short-term stability behavior of 2 × 10−14τ−1. To accurately characterize the short-term stability, a second oscillator (CSO-2) is used for this purpose, yielding an Allan deviation of 3.8 × 10−15 at 1 s.[7] However, for τ > 10 s, the stability degrades rapidly due to the drift of CSO-2, which can be verified from the CSO-1 vs. H-maser comparison result.[3]

    For Fourier frequencies above 1 Hz, the phase noise of CSO-1 is measured as shown in Fig. 3(b). The close-in phase noise of the 10.7885 GHz microwave signal reaches −95 dBc/Hz at 1 Hz and −125 dBc/Hz at 1 kHz. It can potentially be improved by employing a lower-phase-noise amplifier.[26] Above 10 Hz, ripples are present in the phase noise spectrum, arising partly from high-frequency vibrations of the cryocooler and partly from ripple noise coupled through the bias supplies of the LNA and VCP. In the frequency range above 10 kHz, the phase noise exhibits a white noise floor, reaching −135 dBc/Hz from 10 kHz to 1 MHz. This is due to the weak microwave power returned to the LNA. In principle, the high-frequency phase noise can be reduced by lowering the insertion loss of the resonator. Additionally, low-noise filtering techniques can also be implemented for this purpose.[3]

4.   Noise analysis
  • A free-running CSO already exhibits frequency stability at the 10−14 level. Only under optimal locking conditions can the frequency stability of its output signal approach that of the resonator itself, potentially reaching the regime of 10−15 to 10−16. According to the noise model of a phase-locked CSO,[7] the primary noise contributions originate from two categories: frequency-locking noise and resonator frequency noise. Frequency-locking noise is typically introduced during the demodulation process of the LIA, including the intrinsic electronic noise of the demodulator and residual amplitude modulation (RAM) noise from the modulator.[27] The sources of resonator frequency noise are diverse. Typical contributors include temperature fluctuation noise, microwave power fluctuation noise, and vibrational noise from the cryocooler. These noises are analyzed in detail below.

  • In Eq. (2), the front-end detection voltage noise nPFD introduced by the PFD technique aliases with the error voltage Verr, thereby manifesting as frequency-locking noise. Since the PFD technique employs high-frequency modulation and demodulation, the predominant component of the introduced voltage noise originates from the high-frequency white noise of the detector. Consequently, the contribution of frequency-locking noise to the relative frequency stability can be expressed as

    where β1, QL, VPD, φm, and hPFD denote the coupling coefficient at the SLC input port, the loaded Q-factor, the voltage detected by the PD in the PFD technique, the phase modulation depth, and the white noise component of the PFD, respectively.

    The hPFD mainly originates from the inherent noise of the LIA, including the output noise nout and demodulation noise nDM. The relationship between them can be expressed as nlim = (nDM ⊕ nout)/A, hPFD ≡ nlim, where nlim and A are the noise limit and the sensitivity gain factor of the LIA, respectively. By connecting a 50 Ω termination to the input of the LIA (SR830) and collecting the voltage noise from its X output channel, the output voltage noise nout of the LIA can be characterized. Under different gain settings, the output voltage noise of the LIA remains nearly identical, exhibiting a 1/f-type voltage noise of 1.3×10−6(Hz/f)1/2V/Hz at low frequencies.

    To measure the demodulation noise nDM, a weak modulation signal (100 kHz, 4 mV) generated by the LIA itself is fed back into its input port for demodulation. When demodulating in-phase, the modulation signal is detected and amplified at the X channel. When demodulating in quadrature, the voltage noise from the Y channel represents the demodulation noise, as presented in Fig. 4(a). As the gain factor increases, the demodulation noise nDM decreases. At a gain factor of 10000, the limiting voltage hPFD as low as 4 nV/Hz can be detected. Based on Eq. (3), the LIA’s detection limit for the frequency stability is shown in Fig. 4(b), reaching 4×10−16/τ at 0.01∼1 s with a gain factor of 10000.

    To reduce the frequency-locking noise, one can increase the frequency-discrimination coefficient Dν or enhance the gain factor A at the detection front-end. With a phase modulation depth φm = 0.1 and a loaded Q-factor QL = 5.1 × 108, the trend of Dν as a function of the power fed into the SLC is shown in Fig. 4(c). As the power increases from −10 dBm to −2 dBm, Dν correspondingly increases from 0.4 mV/Hz to 4.2 mV/Hz.

    In Fig. 4(c), prior to optimization, CSO-1 operated in a high-power mode with a Dν of 11.8 mV/Hz and a gain factor of 500. The results indicated that the contributed short-term stability of hPFD is 3×10−15/τ. After optimization, operating in a low-power mode, the Dν is only 2 mV/Hz, but the gain factor is increased to 5000. The results show that the contribution of hPFD is reduced to 4×10−16/τ, reaching 2 × 10−16 at averaging times from 10 s to 1000 s.

  • In the PFD technique, a VCP is required to perform phase modulation (PM). However, due to the non-uniform loss of the VCP, a concomitant amplitude modulation (AM) at the same frequency, denoted as αm, is introduced. The AM component is demodulated by the LIA and converted into an offset voltage, thereby affecting the frequency detection. This effect is known as residual amplitude modulation (RAM) noise. According to Eq. (1), the detected error signal with the RAM term is given by Verr = Dν(fRAM + δf), where δf = fc – f0 represents the deviation between the carrier and resonance frequencies. The frequency offset around the locking point, referred to as the RAM frequency, is expressed as

    Here fRAM carries the same weight as the identified frequency deviation δf, thus limiting the long-term stability of frequency locking in the PFD scheme. Furthermore, fRAM acts as a conduit, thereby coupling the microwave power noise nP into the front end of the frequency discrimination scheme, expressed as DνfRAMnP.

    An effective method to reduce RAM noise is shown in Fig. 5(b). The insertion-loss turning point of the VCP is used to suppress amplitude modulation. With the modulation voltage Um applied to the VCP, the amplitude modulation can be expressed as αm=λ˙(Ub)/λ(Ub)⋅Um, thus δαm=Umλ¨/λ⋅δUb, λ˙(Ub)=0, and the E(αm) = 0. Here, Ub is the bias voltage, λ˙ and λ¨ represent the first and second derivatives of the attenuation coefficient with respect to bias voltage.

    In Fig. 5(b), with a modulation voltage Um of 0.1 V and the fitted λ, the contribution of the bias voltage noise δUb to δαm is merely 0.24%. From the expression for fRAM, the transfer coefficient from δαm is 154 Hz/unit. Thus, the conversion coefficient from the bias voltage fluctuation δUb to fRAM is 0.37 Hz/V. The typical δUb is characterized by a 1/f behavior in the close-in frequency range, achieving a level of 5 μV/Hz at 1 Hz for a 5 V bias, which is ensured by the voltage reference module (SRS SIM960). Thus, the predicted contribution to frequency stability is 1.7 × 10−16.

    The scheme in Fig. 5(a) is designed to measure RAM noise. It requires two LIAs. One is employed for the PFD technique, and the other is used to measure the RAM noise. The two LIAs are synchronized with respect to the 100 Hz modulation signal. After demodulation, the voltage characterizing the amplitude modulation is given by u = 2αmγDPdet, where γD is the detection efficiency of the PD, and Pdet is the detected microwave power. The RAM noise can be expressed as

    Here, σu and GLA are the Allan deviation of the output voltage from the RAM-LIA and its gain factor, respectively.

    The measured RAM noise is shown in Fig. 5(c). When the bias voltage Ub of the VCP is set to 2 V and 4 V, the measured RAM noise is on the order of 10−15. When Ub is set to the turning point (5.1 V), the measured RAM noise is notably suppressed. For averaging times between 1–50 s, the stability follows a trend of 1×10−15/τ, reaching a floor of 1.3 × 10−16 at 200 s. Increasing the sampling interval to 4 s, the measured floor reaches 2.0 × 10−16 at 100 s, which aligns with the predicted result.

    In Fig. 5(a), the RAM-LIA also measures the intrinsic voltage noise of the PD.[27] In principle, only the high-frequency white noise within the PD affects the measurement, expressed as

    Here, kB = 1.38 × 10−23 J/K is the Boltzmann constant, and the operating temperature T is 300 K. In Eq. (6), the calculated noise floor is 1.3×10−15/τ, which is consistent with the results shown in Fig. 5(c). The results agree with the intuitive understanding that RAM noise essentially stems from the coupling of carrier amplitude noise into the PFD noise, manifesting as a limitation over longer periods.

  • To precisely stabilize the resonator temperature at FTTP, the measurement scheme shown in Fig. 2(a) is employed to determine the resonant frequency. The relationship between the resonant frequency and temperature can be obtained via curve fitting. At FTTP, the conversion between temperature fluctuations and the relative change in resonant frequency can be expressed as

    Here, f0−1⋅∂2f0/∂T2 is the second derivative of frequency with respect to temperature, with a value of 1 × 10−9 Hz/K2, and ΔTFTTP is the uncertainty of the measured FTTP.[7]

    According to the scheme in Fig. 2(a), the frequency reference is supplied by a hydrogen maser. The uncertainty of the measured frequency can be reduced to 100 μHz when each measurement is averaged over 5 minutes. The accuracy of temperature measurement is ensured by the temperature sensor (CernoxTM), which has a temperature resolution of 89 μK at 4.2 K. By performing peak fitting on the measurement curve in Fig. 2(b), the FTTP can be determined with a fitting uncertainty of 3 mK. Thus, ΔTFTTP can be reduced to 4 mK.

    The temperature of the resonator is measured by an independent temperature sensor. In Fig. 6(a), over a period of one day, the peak-to-peak temperature fluctuation remains within 3 mK. According to Eq. (7), the temperature fluctuation noise can be converted into the contribution to frequency stability, as presented in Fig. 6(b). For averaging times from 1 to 1000 s, the contribution of the temperature fluctuation is on the order of 10−16.

  • In the power servo, part of the microwave power is coupled and converted into a voltage signal by the PD. The fluctuation of the voltage signal can be considered to be caused by the relative intensity noise (RIN) of the microwave signal. The signal is fed back to the VCA after passing through the PI controller, realizing compensation of the RIN. Such a design relies on a crucial assumption: the coupled power and the main power are correlated.

    The principle of the power servo is illustrated in Fig. 7(a). A signal source (E8257D) generates the microwave signal under test at 10.8 GHz. The VCA acts as the actuator by adjusting the attenuation according to the applied bias voltage. The power splitter divides the test signal into two paths: one serves as the sensing branch for the power servo loop, and the other acts as the monitor branch for the main power. Both branches are connected to identical PDs. The Data Acquisition (DAQ) unit simultaneously samples the voltages from both paths. The acquired voltage signals are calculated to characterize the correlation between the sensing branch and the main power branch.

    The voltage samples obtained by the two PDs are Vs and Vd, respectively, and their intrinsic detection noises nPD of the two PDs are uncorrelated. Their cross-spectra satisfy the condition |Ssd(f)|2 < Ss(f)⋅Sd(f). The coherence coefficient between the two sets of voltage samples can be defined as

    The extinction of the coherence coefficient γsd2 is inevitable due to the inherent noise of the detector. The γsd2 can directly characterize the effective bandwidth of the power servo.

    Two independent PDs (DT8012) are employed to analyze the power coherence between the inner and outer loops of the power servo. In Fig. 7(c), the coherence coefficients γsd2 under different detection powers are measured. When the detected power is −14 dBm, the coherence bandwidth (γsd2>0.6) of the two PDs is only 1 Hz. On the other hand, when the detected power is −18 dBm, the coherence bandwidth is significantly enhanced, reaching 10 Hz.

    The RIN of the microwave signal coupled into the sensing branch is suppressed when the power servo is enabled. The in-loop power (Ps) noise reduction is characterized by the out-of-loop detected power (Pd). It achieves a gain of αηPDGPI, where α is the attenuation and GPI is the gain of the PI controller.

    Figure 7(b) shows the RIN of the out-of-loop microwave power. When the detected power is set to −14 dBm, the poor correlation between the inner and outer loops results in the RIN being even higher than that in the open-loop case. At a power level of −18 dBm, the RIN is effectively suppressed, indicating that the power servo is functioning properly.

    The front-end noise in the power servo arises from the reference voltage Vref and the intrinsic noise of the PD (nPD). With the measured microwave power-induced frequency sensitivity (−0.21 Hz/dB) and the conversion efficiency of the PD (21 mV/dB at −18 dBm), an estimation of the fundamental limit that RIN noise places on frequency stability is shown in Fig. 7(d). In the short term, the limit of the RIN on the stability is 5.5 × 10−16, which is constrained by flicker frequency noise converted from the 1/f-type noise of the reference voltage. For averaging times exceeding 1000 s, the RIN begins to degrade, which is due to random walk and drift of the reference voltage.

  • The sapphire resonator is mounted on the 4 K cold plate. Vibrational noise from the pulsetube cryocooler causes variations in the relative position between the microwave probe and the sapphire cylinder. It affects the excitation of the WGH mode, causing variations in the resonant frequency, coupling coefficient, and Q-factor. A prominent feature is the 1.4 Hz pulsed vibration signal, which manifests in the phase noise spectrum of the CSO or in the in-loop noise spectrum of the PFD technique, as shown in Fig. 2(d).

    In Fig. 8(a), the phase noise spectrum of the CSO is zoomed in near 1.4 Hz, and two frequency components can be observed, which originate from the vibrational noise of two independent cryocoolers. Two distinct vibration signals are observable at 1.41 Hz and 1.44 Hz, originating from CSO-1 and CSO-2, respectively. Their contributions to the phase noise spectrum in terms of power and bandwidth are −70 dBc/Hz with a 0.02 Hz bandwidth and −79 dBc/Hz with a 0.03 Hz bandwidth, respectively.

    The phase noise Sφ introduced by the two vibration signals contributes to the frequency stability, which can be calculated as follows:

    In Fig. 8(b), the calculated contributions of the two vibration signals to the frequency stability are 2.5 × 10−15τ−1 and 1.2 × 10−15τ−1, respectively. These contributions constitute a limiting factor for the short-term frequency stability but decrease rapidly with increasing averaging times.

5.   Discussion
  • Figure 9 summarizes the main noise contributions to CSO-1, determining the current-limited frequency stability. In the short term, the main contributor is the 1.4 Hz vibration noise unique to the pulse-tube cryocooler. Although its contribution in the time domain follows a τ−1 decay trend, it limits the frequency stability at the second scale. It can be improved by an assembly with lower vibration sensitivity or by using a cryocooler with lower vibration noise.[28] In the future, mitigating this noise could break the barrier preventing the 1-second frequency stability from reaching the E-16 level.

    By optimizing the frequency-discrimination coefficient and the front-end gain factor of the LIA, the frequency-locking noise is effectively reduced, reaching 4×10−16/τ at 0.4–2 s, as shown in Fig. 4(d). However, it does not decrease with longer averaging times because the frequency-locking noise also includes the output noise of the LIA and the RAM noise, which manifest as a floor of 2 × 10−16 for τ > 2 s.

    For averaging times between 2–200 s, temperature fluctuation noise and microwave power fluctuation noise become the dominant noise sources in CSO-1. Their combined contribution lies in the range of 6 × 10−16 to 1 × 10−15, which limits the frequency stability from reaching 10−16. Notably, at an averaging time of 140 s, the temperature fluctuation noise manifests as a hump in the Allan deviation plot. This is attributable to the time constant of the temperature feedback loop. For the temperature monitor outside the loop, only temperature fluctuations with averaging times greater than 200 s can be measured for effective compensation. This problem can be mitigated by performing thermal conduction simulations to identify optimal sensing or control locations and to tailor the temperature control scheme for the sapphire cavity.

    For averaging times beyond 200 s, power fluctuation noise rises and becomes the dominant limit to frequency stability. This originates from locking the microwave power to the controller’s voltage reference — a noise source that is inherently difficult to suppress. Because the voltage reference noise shows 1/f characteristics near 1 Hz, it typically appears as random-walk noise and drift at lower frequencies, thereby translating into resonant frequency drift. To mitigate this, one could lock the microwave power to a voltage reference with better long-term stability or operate at the power-induced frequency-shift turnover point. This scheme allows the CSO to maintain frequency stability at the 10−16 level, even over medium-to-long averaging times.[29]

6.   Conclusion
  • In the frequency-locked CSO-1, the PFD locking noise is optimized, reaching 4×10−16/τ at 0.4–2 s. However, its performance will not be maintained for longer averaging times due to the RAM noise; a noise floor of 2 × 10−16 is achieved at 100 s. For averaging times beyond 200 s, the voltage reference in the microwave power servo becomes susceptible to random walk or drift, thus limiting the CSO-1’s frequency instability to levels above 10−16. In the future, it is necessary to optimize the assembly of the sapphire resonator to reduce the effect of 1.4 Hz vibrational noise on PFD locking noise in the short term. Effective monitoring of the sapphire resonator’s temperature fluctuations is required, as it constitutes the primary barrier preventing CSO-1 from reaching a stability of 10−16. A more stable voltage reference is also required, both to suppress the RAM noise and to ensure the long-term stability of the microwave power.

Figure (9)  Reference (29)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return