Xiao-Yu Xing, Jing-Qiang Han, Wen-Tao Lu, Zhi-Qiang Zhang, Yang Wang, Shu Liu, Na Li, Yin-Jun Wang, Zi-Yan Wu, Xin-Yu Liu, Sheng-Kai Xia, Ke-Qi Shi, Bo Yang, Chen-Lu Bu, Guan-Xiang Du. Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy center[J]. Chinese Physics B, 2026, 35(9): 094215. doi: 10.1088/1674-1056/ae6f97
Citation: Xiao-Yu Xing, Jing-Qiang Han, Wen-Tao Lu, Zhi-Qiang Zhang, Yang Wang, Shu Liu, Na Li, Yin-Jun Wang, Zi-Yan Wu, Xin-Yu Liu, Sheng-Kai Xia, Ke-Qi Shi, Bo Yang, Chen-Lu Bu, Guan-Xiang Du. Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy center[J]. Chinese Physics B, 2026, 35(9): 094215. doi: 10.1088/1674-1056/ae6f97

Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy center

  • Received Date: 12/03/2026
    Accepted Date: 17/04/2026
    Available Online: 01/09/2026
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(5)  /  Tables(1)

Article Metrics

Article views(602) PDF downloads(0) Cited by(0)

Access History

Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy center

Abstract: We design a flexible microwave transmission line based on a splitting-combining structure and integrate it with a fiber-diamond probe to form a unified quantum probe, providing a resonant driving microwave magnetic field. This device utilizes a polyimide (PI) flexible substrate, which is conformable. A through-hole is designed at the center of the ring to accommodate fiber insertion, enabling the sensing unit to approach the sample under test. Experimental results demonstrate that the device achieves a 394 MHz bandwidth, with a microwave efficiency of 4.64G/W, balancing broadband with a strong near-field microwave magnetic field. Finally, we characterize the spatial distributions of the static magnetic-field magnitude and the angle between the static magnetic field and the selected NV axis on the surface of the kagome magnet TmMn6Sn6, demonstrating the application potential of the proposed probe for magnetic-material characterization.

1.   Introduction
  • Nitrogen-vacancy (NV) center in diamond is a common intrinsic lattice defect.[1–3] This defect demonstrates stable optical properties and long coherence times at room temperature, allowing for all-optical spin state initialization and readout.[4,5] As a result, it has become a leading platform for solid-state quantum sensing in the field of quantum precision measurement.

    Microwave devices are crucial for achieving precise control of spin states, and their structure and performance strongly affect the microwave-driving capability. Currently, two main types of devices are commonly used: one is resonant structures, such as split-ring resonators[6] and resonant devices;[7] the other is transmission structures, such as microstrip lines[8] and coplanar waveguides.[9] The advantage of resonant structures lies in their ability to concentrate energy effectively during resonance, generating strong near-field microwave magnetic fields at specific frequencies. However, due to their narrow bandwidth, they struggle to adapt to broadband requirements. Transmission lines, on the other hand, offer broader bandwidth but fail to generate an efficient near-field microwave magnetic field. For example, Bayat et al.[10] proposed double split-ring resonators that generate a microwave magnetic field of 5.6 Gauss, but with a bandwidth of only 40 MHz. Babashah et al.[11] proposed a coplanar waveguide (CPW) coil with a bandwidth of 900 MHz, but it generates only a 1.6 Gauss microwave magnetic field. Xia et al.[12] introduced a flexible ring resonator with a microwave magnetic field of 3.5 Gauss and a bandwidth of 140 MHz. Therefore, designing a compact device that balances bandwidth and microwave magnetic field becomes particularly important.

    In this paper, we designed a flexible transmission line incorporating a through-hole within a ring-shaped microstrip structure. By integrating a tapered fiber–diamond probe, an integrated quantum probe was formed. The transmission line adopts a splitting–combining configuration and can be conformally mounted on a supporting substrate. Its operating frequency is tuned to approximately 2.87 GHz while maintaining a balance between microwave magnetic-field efficiency and bandwidth. The microwave magnetic field distribution was scanned at a distance of approximately 20 μm above this microwave device, and the axial attenuation characteristics were further analyzed. In addition, scanning measurements were performed on the micro-magnetic material, demonstrating the potential of the proposed probe for static magnetic-field characterization.

2.   Basic theory and experimental setup
  • NV center is a kind of point defect in diamond with C3v symmetry, in which a carbon atom is replaced by a nitrogen atom, and an adjacent lattice site becomes a vacancy, as shown in Fig. 1(a).[13–15] Its Hamiltonian[16] can be expressed as

    Here, D = 2.87 GHz represents the axial zero-field splitting (ZFS) parameter, E is the non-axial zero-field splitting parameter, mainly caused by the local stress in the diamond. γNV = −2.802495 MHz/Gs is the gyromagnetic ratio of electron spin. S is the spin-1 angular momentum operator, with components Sx, Sy, and Sz.

    Figure 1(b) shows the energy levels of the NV center. The negatively charged NV center (NV−) consists of six electrons, two of which are unpaired, giving rise to the spin-triplet ground state 3A2 and excited state 3E. Spin–spin interaction results in the zero-field splitting between the ms = 0 and ms = ±1. Without an external static magnetic field, the ms = ±1 spin states are degenerate, with a ZFS of 2.87 GHz between them and ms = 0.[17,18] Pumped by a 532 nm green laser, the system in the ground state absorbs energy and transitions to the excited state, then returns to the ground state through either radiative or non-radiative transitions.

    Optically detected magnetic resonance (ODMR) combines optical and microwave resonance techniques, enabling the readout of the NV center’s electron spin states through analyzing fluorescence intensity.[19,20] With an axial static magnetic field, the degenerate state ms = ±1 splits to ms = +1 and ms = −1, following the Zeeman effect.[21] Energy level transitions between ms = 0 and ms = ±1 can be driven by a resonant microwave magnetic field.

    Figure 1(c) illustrates the experimental system, consisting of an optical module, a microwave driving module, a pulse control module, and a data analyzing module.

3.   Simulation design
  • Common rigid PCB materials (such as FR-4) lack conformability and have a certain thickness, failing to closely adhere to the surface of magnetic materials, posing challenges for microwave magnetic field imaging. In this work, we designed a splitting-combining flexible microwave device using a polyimide (PI) substrate, with a relative dielectric constant of 3.4, loss tangent tanδ = 0.005, and a substrate thickness h1 of 0.1 mm. The design process of this device is shown in Fig. 2(a).

    We started from the initial flat design, where the substrate is covered with conductive patches on the upper layer, and the lower copper layer serves as the device’s reference ground. The thickness of copper h2 is 0.018 mm. Through two 50 Ω impedance-characterized microstrip lines, the electromagnetic energy is concentrated in the inner ring area through the splitting-combining structure. To facilitate the extension of the tapered fiber, a through-hole is added. The edge of the hole is 0.1 mm away from the inner ring to prevent the drill from breaking the trace during processing. To ensure the transmission performance after conformal mapping, we chose to place the impedance matching part in the same plane. The radius of the sectoral short-circuited lines is then adjusted to fine-tune the operating frequency of the device to around 2.87 GHz. After rounding the corners, the energy loss during signal transmission is minimized. The key dimensions of the optimized device are shown in Table 1, and the related parameters are illustrated in Figs. 2(a) and 2(b). Figure 2(c) shows the design of a 3D-printed bracket with appropriate slots to hold the flexible device. The bottom of the bracket is equipped with holes for fixing the SMA connector. This design secures the device onto the bracket, forming an integrated probe, allowing the fiber to be placed closer to the material under test, facilitating the subsequent scanning of magnetic materials. As shown in Figs. 2(d) and 2(e), simulated comparisons between the flat and conformal configurations show that bending introduces only limited variations in the S11 and S21 responses of the flexible device within the investigated frequency range of 2.6–3.1 GHz.

    We found that as the radius of the middle ring decreases, the generated microwave magnetic field increases. By adjusting the inner radius r3 of the ring, the operating frequency of the device can be tuned. Figure 2(f) shows the varied operating frequency of the device as r3 changes. We further analyzed the dependence of the device bandwidth on other key structural parameters. The results show that, within the investigated range, the middle-ring size mainly affects the operating frequency, whereas its influence on the −10 dB bandwidth does not exhibit an obvious linear trend. By contrast, the radius of the sectoral short-circuited lines and the ring width have a more direct impact on the impedance-matching and coupling conditions, and therefore affect the effective bandwidth more significantly. The corresponding parametric results are shown in Figs. 2(g) and 2(h).

    As a two-port device, the flexible transmission line can be characterized by scattering parameters to describe its reflection and transmission properties, where S11 denotes the input reflection coefficient and S21 denotes the forward transmission coefficient.[23] To provide an intuitive description of how different structural sections contribute to the overall transmission behavior, we introduced an approximate cascade model based on the linear magnitude of S21, so as to evaluate the consistency between the reconstructed cascade model and the overall structure, as shown in Fig. 2(i). This approximation was used only under the assumptions that multiple reflections at the interfaces are negligible and that impedance matching between adjacent sections is reasonably good.

    Let |S21v(f)|, |S21p(f)|, and |S21all(f)| denote the linear magnitudes of the transmission coefficient for the vertical section, the planar section, and the overall structure, respectively, at frequency f. Since the combined structure can be approximately regarded as consisting of two vertical sections and one planar section connected in series, the reconstructed transmission magnitude |S21comp(f)| can be written as

    To quantify the difference between the approximate cascade model and the overall structure, the relative error at frequency f is defined as

    Then in the 2.60–3.10 GHz, the calculated RMSRE is 0.04 between the approximate cascade model and the overall structure. The result indicates that conformal mapping introduces only a minor impact on the device transmission performance.

4.   Experimental test of the conformal device
  • Figure 3(a) shows the processed flexible device, featuring microwave power input on one side and a 50 Ω end load on the other. The device’s S11 was measured using a network analyzer. As shown in Fig. 3(b), from 2.6 GHz to 3.1 GHz, the simulation data align well with the measured data, with an actual bandwidth of 394 MHz when the input return loss is −10 dB. The operating frequency is approximately 2.87 GHz, with an S11 less than −30 dB. The measured value is −29.1 dB, indicating low return loss. This demonstrates the device’s excellent impedance matching capability.

    In addition, to provide a more complete experimental characterization of the device under conformal operation, we further measured and compared both S11 and S21 for the flat and conformal configurations. As shown in Figs. 3(c) and 3(d), only limited variations are observed in both the reflection and transmission responses under the investigated bending condition, indicating that conformal deformation does not significantly degrade the impedance matching or forward transmission capability of the device. Furthermore, we suppose that the discrepancy between simulation and measurement mainly originates from the difference between the practical RF connectors used in the experiment and the ideal wave-port excitation adopted in the simulation.

    Under an applied static magnetic field, different projections of the field onto the four NV axes lead to eight resonance peaks in the ODMR spectrum, corresponding to the four NV orientations.[24–26] In the data analysis, each resonance peak was fitted individually using a single-Lorentzian function.[27] As shown in Fig. 4(a), the ODMR spectrum under a static magnetic field was obtained.

    After the ODMR measurement, a typical Rabi oscillation experiment was performed at a position approximately 20 μm above this microwave device to determine the effective resonant microwave magnetic-field amplitude for the selected NV transition at 2.65 GHz, as shown in Fig. 4(b). This effective driving field, denoted as Bmw, can be calculated from the Rabi oscillation frequency according to

    For the selected transition, the ground-state three-level system of the NV center (ms = −1, 0, +1) is projected onto the effective two-level subspace (ms = 0, +1) or (ms = 0, −1). Therefore, the field Bmw extracted from the Rabi oscillation corresponds to the effective resonant driving field for the selected NV transition, rather than the total local microwave magnetic-field amplitude. For a locally linearly polarized microwave field, the effective circular driving component is smaller than the total field amplitude by a factor of 1/2. Therefore, the actual local microwave magnetic-field amplitude satisfies B=2Bmw

    Since the frequency of the fitted curve in Fig. 4(b) is 2.93 MHz, the corresponding actual microwave magnetic field amplitude is 1.48 Gauss (B). We use the microwave efficiency Λ to quantify the device’s ability to convert excitation power into a microwave magnetic field, as expressed in formula[28]

    Here, P represents the actual microwave power delivered to the device input after correcting for the insertion loss of the microwave delivery path, including the cable and connectors.

    With an experimentally measured excitation power of 20.08 dBm, the microwave efficiency at this point is calculated to be 4.64G/W. For ensemble NV centers, a sufficiently high microwave efficiency is generally required to achieve effective microwave driving. The microwave efficiency obtained in this work is adequate for the subsequent experimental measurements.

    A complete Rabi oscillation curve consumes time. When the microwave pulse duration is extremely short, the Rabi oscillation curve approximates a linear function.[29] Therefore, as shown in Fig. 4(b), the corresponding fluorescence intensity can be measured and the microwave magnetic field distribution of the device can be rapidly scanned.

    To determine the optimal position for the optical fiber, a meticulous scan of the planar area with a radius of 250 μm was conducted approximately 20 μm above the microwave transmission line. We used the first selected resonance peak in Fig. 4(a) for analysis. This scan was performed under precise control of the LABVIEW software, employing a step size of 10 μm. Figure 4(c) shows the simulated surface normalized distribution of the microwave magnetic field, while Fig. 4(d) shows the experimentally measured normalized distribution near 2.87 GHz. For quantitative comparison, the similarity between the two distributions was evaluated using the Pearson correlation coefficient. The calculated correlation coefficient is 0.86, indicating good consistency in the overall spatial distribution trends. Although discrepancies in magnitude remain, the correlation analysis shows that the experimental results capture the main spatial features of the simulated distribution.

    Furthermore, we investigated axial attenuation using a displacement stage. With the diamond probe positioned approximately 100 μm from the borehole edge, we performed longitudinal scans at 10 μm intervals, starting approximately 20 μm above the microwave transmission line, as shown in Fig. 4(e). The measured and simulated data points were normalized respectively. To quantitatively evaluate the agreement between the measured data points and the simulated curve, RMSE between the normalized measured data and the normalized simulated values at the same sampling positions was calculated. The resulting RMSE is 0.05, indicating a generally consistent trend between the simulated and measured axial attenuation profiles.

    Moreover, the magnetic field at 100 μm from the edge of the scanning plane exceeded 1 Gauss, sufficient to excite the NV center. Therefore, we coupled the optical fiber to the NV center, which is encapsulated within a device and support structure, as shown in Fig. 2(c).

    In order to more effectively verify its application value in the characterization of magnetic materials, we used the integrated probe shown in Fig. 2(c) to conduct a two-dimensional characterization on the surface of kagome magnet TmMn6Sn6, which is a representative micro-magnetic material.[30] Based on reasonable inference, the selected scanning area was approximately located at 50 μm above the surface of the micro-magnetic material, as shown in Fig. 5(a). The optical image of the sample is shown in Fig. 5(b), using a Mitutoyo long-working-distance objective with far-field correction. The applied bias magnetic field was approximately 200 Gauss, the scanning area size was 500 μm × 500 μm, and the scanning step size was 9.375 μm.

    Under the applied bias magnetic field, the resonance frequencies of the selected ODMR doublet shift with the local static magnetic field. At each scanning point, we first fitted the ODMR spectrum and selected one pair of resonance peaks belonging to the same NV axis, whose center frequencies were denoted as fu and fl (fu > fl). These two fitted center frequencies were then used to reconstruct the local static magnetic-field magnitude B and the angle θ between the local magnetic field and the selected NV axis.

    Specifically, the magnetic field magnitude B is given by[31]

    And the angle θ is obtained from[31]

    In this work, the parameters D and E were determined from the zero-field ODMR spectrum and treated as fixed during the scan. The external magnetic field was aligned approximately along the vertical direction. The selected pair of peaks corresponds to the outermost resonance pair, which indicates that the projection of the magnetic field onto the corresponding NV axis is the largest. Therefore, this NV axis is the closest to the direction of the applied magnetic field.

    The resulting spatial distribution of the static magnetic-field magnitude is shown in Fig. 5(c), reflecting the spatial variation of the magnetic-field distribution on the sample surface. Correspondingly, the spatial distribution of θ is shown in Fig. 5(d). The relatively larger θ values in the lower-right region indicate that the local static magnetic field there has a stronger transverse component with respect to the selected NV axis. These results demonstrate that the fiber–diamond integrated probe can be used for static magnetic-field characterization of micro-magnetic materials.

5.   Conclusion and perspectives
  • This paper designs and implements a flexible microwave transmission structure suitable for diamond NV center quantum sensing. This structure balances efficient microwave magnetic field and broadband, with a bandwidth of 394 MHz and a microwave efficiency of 4.64G/W. Furthermore, this microwave structure is integrated with a fiber-diamond probe.

    A scanning plane with a radius of 250 μm was established approximately 20 μm above the transmission line, and the microwave magnetic field distribution was measured. Simultaneously, the microwave magnetic field remained above 1 G at approximately 100 μm from the aperture edge, verifying the attenuation law of the microwave magnetic field in the axial direction, thus determining the appropriate placement of the fiber. Two-dimensional scanning characterization of the local static magnetic field on the surface of the magnetic sample was also performed.

    Future research could further optimize flexible structures to improve the local microwave magnetic field strength and spatial uniformity, and combine them with smaller diamond sensing units and high-precision scanning systems to enhance magnetic-field imaging capability.