全蒸发技术在铀同位素丰度测量上的应用

上一篇

下一篇

魏兴俭, 徐新冕, 张海路, 华永明. 2001: 全蒸发技术在铀同位素丰度测量上的应用, 质谱学报, 22(1): 7-14. doi: 10.3969/j.issn.1004-2997.2001.01.002
引用本文: 魏兴俭, 徐新冕, 张海路, 华永明. 2001: 全蒸发技术在铀同位素丰度测量上的应用, 质谱学报, 22(1): 7-14. doi: 10.3969/j.issn.1004-2997.2001.01.002
Wei Xingjian, Xu Xinmian, Zhang Hailu, Hua Yongming. 2001: The Application of Total Sample Evaporation Technique in Mass Spectrometric Measurement of Uranium Isotopes, Journal of Chinese Mass Spectrometry Society, 22(1): 7-14. doi: 10.3969/j.issn.1004-2997.2001.01.002
Citation: Wei Xingjian, Xu Xinmian, Zhang Hailu, Hua Yongming. 2001: The Application of Total Sample Evaporation Technique in Mass Spectrometric Measurement of Uranium Isotopes, Journal of Chinese Mass Spectrometry Society, 22(1): 7-14. doi: 10.3969/j.issn.1004-2997.2001.01.002

全蒸发技术在铀同位素丰度测量上的应用

The Application of Total Sample Evaporation Technique in Mass Spectrometric Measurement of Uranium Isotopes

  • 摘要: 采用全蒸发技术、离子流强度积分的方式,用多接收器热表面电离质谱计测量铀同位素丰度。全蒸发技术可克服质量分馏效应带来的测量误差,而直接实现准确测量。开发了用于全蒸发技术的测量软件,对影响测量结果的因素,诸如分析用样量、信号损失与试样残留等进行了研究。最后得出在铀同位素丰度比为0.1~10的范围内,95%的置信水平下,该方法的相对扩展不确定度为0.10%。
  • 加载中
  • 加载中
计量
  • 文章访问数:  353
  • HTML全文浏览数:  96
  • PDF下载数:  14
  • 施引文献:  0
出版历程
  • 刊出日期:  2001-02-28

全蒸发技术在铀同位素丰度测量上的应用

  • 中国工程物理研究院 ,

摘要: 采用全蒸发技术、离子流强度积分的方式,用多接收器热表面电离质谱计测量铀同位素丰度。全蒸发技术可克服质量分馏效应带来的测量误差,而直接实现准确测量。开发了用于全蒸发技术的测量软件,对影响测量结果的因素,诸如分析用样量、信号损失与试样残留等进行了研究。最后得出在铀同位素丰度比为0.1~10的范围内,95%的置信水平下,该方法的相对扩展不确定度为0.10%。

English Abstract

参考文献 (0)

目录

/

返回文章