冷凝回流三酸体系消解-ICP-MS法测定高纯石墨中的痕量杂质元素

上一篇

下一篇

贾静, 张英新, 童坚, 李文军. 2009: 冷凝回流三酸体系消解-ICP-MS法测定高纯石墨中的痕量杂质元素, 质谱学报, 30(4): 229-233.
引用本文: 贾静, 张英新, 童坚, 李文军. 2009: 冷凝回流三酸体系消解-ICP-MS法测定高纯石墨中的痕量杂质元素, 质谱学报, 30(4): 229-233.
JIA Jing, ZHANG Ying-xin, TONG Jian, LI Wen-jun. 2009: Determination of Impurity Elements in High Purity Graphite by ICP-MS after Mixture Acid Digestion in Reflux Condenser System, Journal of Chinese Mass Spectrometry Society, 30(4): 229-233.
Citation: JIA Jing, ZHANG Ying-xin, TONG Jian, LI Wen-jun. 2009: Determination of Impurity Elements in High Purity Graphite by ICP-MS after Mixture Acid Digestion in Reflux Condenser System, Journal of Chinese Mass Spectrometry Society, 30(4): 229-233.

冷凝回流三酸体系消解-ICP-MS法测定高纯石墨中的痕量杂质元素

Determination of Impurity Elements in High Purity Graphite by ICP-MS after Mixture Acid Digestion in Reflux Condenser System

  • 摘要: 建立冷凝回流混合酸消解,ICP-MS法测定高纯石墨中35种痕量元素的方法.采用1 mL HNO3, 1 mL HClO4和3 mL H2SO4于220 ℃消解0.2 g 样品,以Sc、Rh、Cs、Re作内标,ICP-MS法测定各元素的方法检出限为0.000 6~0.1 μg·L -1 ,加标回收率达82%~110%,相对标准偏差(RSD, n =10)为0.58%~ 5.8% .该方法适用于高纯石墨中痕量杂质含量的测定.
  • 加载中
  • 加载中
计量
  • 文章访问数:  800
  • HTML全文浏览数:  328
  • PDF下载数:  62
  • 施引文献:  0
出版历程
  • 刊出日期:  2009-08-28

冷凝回流三酸体系消解-ICP-MS法测定高纯石墨中的痕量杂质元素

  • 北京科技大学,应用科学学院化学系,北京,100083
  • 北京有色金属研究总院,分析测试技术研究所,北京,100088

摘要: 建立冷凝回流混合酸消解,ICP-MS法测定高纯石墨中35种痕量元素的方法.采用1 mL HNO3, 1 mL HClO4和3 mL H2SO4于220 ℃消解0.2 g 样品,以Sc、Rh、Cs、Re作内标,ICP-MS法测定各元素的方法检出限为0.000 6~0.1 μg·L -1 ,加标回收率达82%~110%,相对标准偏差(RSD, n =10)为0.58%~ 5.8% .该方法适用于高纯石墨中痕量杂质含量的测定.